Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K150343

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-04-03

    (51 days)

    Product Code
    Regulation Number
    870.1255
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V8 Transluminal BAV Catheter is indicated for Balloon Aortic Valvuloplasty.

    Device Description

    The V8 Transluminal BAV Catheter System features an hour glass shaped dilatation balloon on the distal end of a catheter. The catheter is inserted through a percutaneous entry site into the common femoral artery via an introducer sheath and advanced retrograde to the aortic valve. The catheter is always delivered over a guidewire. The balloon is then inflated to dilate the stenotic aortic valve leaflets in an effort to increase valve opening dimensions and systemic blood flow by improving leaflet mobility. The hour glass shaped balloon, with the undersized waist segment, is intended to minimize over-dilatation of the valve annulus while allowing the full dilation of the valve leaflet. The bulbous proximal balloon segment is appropriately sized for the patient's aortic root dimensions to maximize valve leaflet opening.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the "V8 Transluminal BAV Catheter" (K150343). The purpose of this submission is not to introduce a new device, but rather to update the labeling of an existing device (V8 Transluminal BAV Catheter, K133607) and establish its substantial equivalence to another predicate (Z-Med/Z-Med II, K122012).

    The key finding is that no new performance data or studies were conducted for this specific submission (K150343) because the device design, materials, and function remain identical to the predicate V8 device (K133607). Therefore, the acceptance criteria and the studies proving the device meets them refer to previous submissions (K133607 and K123111).

    Given this information, I will synthesize the available information regarding the acceptance criteria and prior studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this submission explicitly states "no design changes have been made to the predicate K133607 V8 device" and "performance testing performed during design verification and shelf life testing were not repeated because the design and materials of the device remain the same as the V8 predicate," the acceptance criteria and performance are referred to the previous submissions. The document lists the types of tests performed in the predicate design verification. Specific quantitative acceptance criteria are not explicitly stated in this document but are implied by the successful completion of these tests in the predicate device.

    Acceptance Criteria (Implied by Predicate Verification)Reported Device Performance (from Predicate)
    Meet specified balloon rated burst pressurePerformance met for predicate V8 (K133607)
    Meet specified balloon compliancePerformance met for predicate V8 (K133607)
    Conform to critical dimensionsPerformance met for predicate V8 (K133607)
    Compatible with guidewire and introducerPerformance met for predicate V8 (K133607)
    Achieve specified deflation timesPerformance met for predicate V8 (K133607)
    Maintain integrity over repeat inflationsPerformance met for predicate V8 (K133607)
    Exhibit no leaksPerformance met for predicate V8 (K133607)
    Withstand specified tensile forcesPerformance met for predicate V8 (K133607)
    Resist kinkingPerformance met for predicate V8 (K133607)
    Exhibit adequate torque performancePerformance met for predicate V8 (K133607)
    Biocompatible for patient contactAchieved (test data provided in K123111)
    Sterile for patient-contacting componentsAchieved (sterilization validation in K123111)
    Maintain shelf-lifeAchieved (shelf life testing data in K133607)
    Demonstrate safety and effectiveness for device delivery, inflation, deflation, and removalAchieved (design verification in K133607, additional bench testing summarized in Section 11.5 of this submission)
    Demonstrate performance, including ability to treat aortic stenosis (in vivo)Achieved (based on predicate device in K132728, and identical shape/principle of operation)

    2. Sample Size Used for the Test Set and Data Provenance

    This document does not provide details on sample sizes for test sets for the initial predicate device studies. It only states that the performance testing was not repeated because the design and materials are the same. Therefore, the sample sizes and data provenance would refer to the original K133607 and K123111 submissions. The nature of these tests (design verification, shelf life, biocompatibility) implies bench and potentially in-vitro/animal studies, rather than human clinical trials for this type of device. There is no mention of country of origin or retrospective/prospective for the previous studies within this document.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This request is primarily applicable to studies involving interpretation of data (e.g., medical images). The listed tests are primarily engineering and material performance tests (e.g., burst pressure, tensile strength, biocompatibility, sterility). For such tests, "ground truth" is typically established by objective physical measurements and adherence to recognized standards, rather than expert consensus on interpretive data. Therefore, this question is not directly applicable in the context of the provided document.

    4. Adjudication Method for the Test Set

    As explained above, the tests are primarily objective physical and material performance evaluations. Adjudication methods like 2+1 or 3+1 are used for subjective assessments (e.g., radiographic interpretations). This is not relevant to the types of tests described in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating substantial equivalence of a medical device (a catheter) through bench testing, material characterization, and reference to prior regulatory submissions. MRMC studies are typically performed for diagnostic devices, especially those involving human interpretation of complex data (like imaging algorithms), to assess the impact of a device on reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the loop performance) Was Done

    No, this question refers to the performance of an algorithm without human intervention, which is relevant for AI/ML devices. The V8 Transluminal BAV Catheter is a medical instrument (hardware), not an algorithm or AI system. Therefore, a "standalone algorithm performance" study is not applicable.

    7. The Type of Ground Truth Used

    For the performance data referred to from the predicate submissions (K133607 and K123111), the ground truth for tests like burst pressure, compliance, critical dimensions, tensile strength, kink resistance, and torque would be established through:

    • Objective physical measurements: Using calibrated equipment and standardized test methods.
    • Chemical/Biological analysis: For biocompatibility and sterility, adhering to ISO standards and laboratory protocols.
    • Engineering specifications: The device is designed to meet specific engineering tolerances and performance parameters.

    Ground truths like expert consensus, pathology, or outcomes data would typically be used for diagnostic or treatment effectiveness studies involving patient data, which are not detailed here as the primary evidence for this 510(k) submission.

    8. The Sample Size for the Training Set

    This question is only relevant for artificial intelligence or machine learning models. The V8 Transluminal BAV Catheter is a physical medical device. Therefore, the concept of a "training set" is not applicable.

    9. How the Ground Truth for the Training Set Was Established

    As explained above, there is no training set for this type of physical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1