Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K150284

    Validate with FDA (Live)

    Date Cleared
    2015-05-15

    (99 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System, Model 154000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and shap endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, grnecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatic use. It is intended to be used by trained physicians in an operating room enviroument in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    The subject IS4000 da Vinci EndoWrist Instruments are to be used with the IS4000 da Vinci Surgical System. The EndoWrist Instruments consist of the housing, shaft, wrist, and tip. The housing contains the connection mechanism to the IS4000 da Vinci Surgical System arm. The shaft and wrist allow for different axes of rotation, and the instrument tip is used to interact with the patient tissue. These instruments are reusable and provided non-sterile.

    AI/ML Overview

    This document describes K150284 for the IS4000 da Vinci EndoWrist® Instruments. It is a 510(k) premarket notification, indicating a claim of substantial equivalence to a predicate device. Therefore, a full clinical study with acceptance criteria, sample sizes for test and training sets, expert qualifications, and MRMC studies, as typically seen for novel high-risk devices, is not required nor provided in this summary.

    The summary focuses on design verification and validation to demonstrate that the modified instruments are as safe and effective as the predicate device.

    Here's the information extracted from the provided text, focusing on the available details:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) for modified instruments claiming substantial equivalence, explicit "acceptance criteria" in terms of clinical performance metrics (like sensitivity, specificity, or AUC) are not provided. Instead, the "acceptance criteria" are implied by successful completion of design verification and validation testing, confirming that the new instruments meet design input requirements and perform comparably to the predicate.

    Acceptance Criteria (Implied by Verification/Validation)Reported Device Performance
    Instrument reliability and durabilityVerified through bench testing.
    LeakageVerified through bench testing.
    Roll, pitch, and yaw range of motionVerified through bench testing.
    Jaw close and open positionsVerified through bench testing.
    FrictionVerified through bench testing.
    TorqueVerified through bench testing.
    Instrument insertionVerified through bench testing.
    Adequacy of labelingVerified through bench testing.
    Software compatibility with IS4000 System softwareVerified through software compatibility testing.
    Instrument data verification with IS4000 System softwareVerified through instrument data verification testing.
    Features meet user needs and intended use (as documented in product requirements)Validated in representative simulated clinical settings (porcine in vivo and cadaveric tissue models).
    Safety and efficacy in normal and expected worst-case clinical useAssessed in representative simulated clinical settings (porcine in vivo and cadaveric tissue models).

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not explicitly state "sample sizes" in terms of number of cases or patients for a test set in a traditional clinical study sense. The testing involved "representative simulated clinical settings that utilized porcine (in vivo) and cadaveric tissue models." The number of porcine or cadaveric models used is not specified.

    Data Provence:

    • Country of origin: Not specified.
    • Retrospective or prospective: The design verification and validation testing are prospective, performed specifically for regulatory submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not mention the use of "experts" to establish a ground truth in a clinical study context. The validation involved assessing safety and efficacy in simulated clinical settings, which would likely have involved qualified personnel (e.g., engineers, medical professionals familiar with the product) but not specifically "experts" establishing a diagnostic ground truth.

    4. Adjudication Method for the Test Set

    Not applicable. There was no clinical study with a test set requiring adjudication of diagnostic outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The submission relies on demonstrating substantial equivalence through non-clinical performance data (bench testing, simulated use).

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. This device is a surgical instrument, not an AI algorithm. Its performance is always with a human-in-the-loop (the surgeon).

    7. The Type of Ground Truth Used

    For the design validation, the "ground truth" was established by the expected functional performance of the instruments in simulated clinical settings (e.g., successful grasping, cutting, suturing, demonstrating intended interaction with tissue) and meeting documented product requirements. This is based on functional performance in controlled, simulated environments rather than diagnostic pathology or long-term clinical outcomes data.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical surgical instrument, not an AI system that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1