Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K143424

    Validate with FDA (Live)

    Device Name
    UTAH CVX-RIPE
    Date Cleared
    2015-08-28

    (270 days)

    Product Code
    Regulation Number
    884.4260
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UTAH CVX-RIPE™ is intended to mechanically improve the favorability of the cervix of pregnant patients at term gestation, in which induction of labor is medically indicated.

    Device Description

    The UTAH CVX-RIPE™ is a double balloon catheter. The tip is placed through the vagina and into the cervical canal. The distal balloon is positioned against the internal cervical os and inflated with saline. The proximal balloon is positioned against the external cervical os and inflated with saline, providing gradual mechanical cervical dilation by simultaneously applying pressure on the internal and external cervical os. The materials used to manufacture the device, and which have patient contact, are all medical grade silicone.

    The UTAH CVX-RIPE™ consists of an 18 Fr., 42 cm silicone catheter with a silicone tip and two silicone balloons. Each balloon is independently inflated via a separate inflation lumen through color coded check valves located at the proximal end of the device. The check valves are marked accordingly, "U" for the distal uterine balloon and "V" for the proximal vaginal balloon. A malleable stainless steel stylet is included for ease of placement. The stylet is placed through a third separate lumen that is closed at the distal tip. The stylet is preloaded in the catheter.

    AI/ML Overview

    This document is a 510(k) premarket notification for the UTAH CVX-RIPE™ cervical ripening balloon catheter. It asserts substantial equivalence to a predicate device, the Cook® Cervical Ripening Balloon (K131206).

    Here's an analysis of the provided text in relation to your request:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state "acceptance criteria" with numerical or performance targets in a tabular format that is typically expected for, say, diagnostic accuracy or predictive power. Instead, it lists mechanical performance tests and states that the device successfully met the requirements of these tests.

    Acceptance Criteria (Implied)Reported Device Performance
    Balloon Deflation ReliabilityBalloons deflate properly after use.
    Balloon IntegrityBalloons do not burst at maximum recommended inflation volume and duration under simulated use conditions.
    Balloon OveruseBalloons do not leak and properly deflate when using an inflation volume and duration greater than maximum recommendations.
    Balloon Volume MaintenanceBalloons do not leak.
    Balloon Response to TractionDistal balloon does not pull out from the patient under simulated use conditions.
    Bond StrengthBonded connections do not fail during simulated use conditions.
    Stylet PunctureStylet does not puncture through the catheter during simulated use.
    BiocompatibilityMeets the requirements of ISO 10993-1.
    Shelf LifeEstablished based on results of an accelerated aging study.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify a separate "test set" in the context of clinical data for performance evaluation. The performance tests mentioned are mechanical/bench tests. For these mechanical tests, the sample sizes are not explicitly stated. The document does not refer to any human-patient data collection (retrospective or prospective) for this 510(k) submission, as it relies on substantial equivalence to a predicate device based on similar intended use and technological characteristics, supported by bench testing. Therefore, there is no information on data provenance (e.g., country of origin).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable and not provided. The "test set" here refers to mechanical performance tests, not clinical data requiring expert human assessment for ground truth.

    4. Adjudication Method for the Test Set:

    This is not applicable as the "test set" refers to mechanical performance tests, not a clinical study involving human readers or adjudicators.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

    This type of study was not done. The submission is for a mechanical medical device (a balloon catheter), not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not applicable. The device is a physical medical device, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the mechanical performance tests, the "ground truth" would be the engineering specifications and expected physical behavior validated through laboratory testing. For example, for "Balloon Deflation Reliability," the ground truth is simply whether the balloons did, in fact, deflate properly according to specified parameters. There is no biological or patient-related "ground truth" in this context.

    8. The sample size for the training set:

    This is not applicable. There is no machine learning or AI component to this device that would require a "training set." The training referenced in the document would be for healthcare professionals on how to use the device, not a dataset for an algorithm.

    9. How the ground truth for the training set was established:

    This is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1