    K143368

    Validate with FDA (Live)

    Device Name
    GenIQ
    Date Cleared
    2015-07-29

    (246 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GenIQ is an automated post-processing software option that is indicated for use on dynamic magnetic resonance imaging data sets to generate parametric image intensity variations over time. This dynamic change in signal intensity is used to calculate functional parameters related to tissue flow and leakage of the contrast agent from the intravascular to the extracellular space.

    GenIQ provides information that when interpreted by a trained physician, can be useful for assessing tissue vascular properties.

    Device Description

    GenIQ is a software application used for the pharmacokinetic analysis of Dynamic Contrast Enhanced (DCE) MRI data sets. The application is used to perform a General Kinetic Model (GKM)-based pharmacokinetic modeling of DCE-MRI data. The goal of GenIQ is to extract functional parameters describing tissue vascular properties such as forward and backward transfer constants, plasma volume, and volume of extra-cellular space.

    AI/ML Overview

    Acceptance Criteria and Study for GenIQ

    The provided document describes the GenIQ, an automated post-processing software option for dynamic Magnetic Resonance Imaging (MRI) data sets. It calculates functional parameters related to tissue flow and contrast agent leakage.

    1. Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics like sensitivity, specificity, accuracy, or AUC. Instead, it states that "Simulated use testing was performed on digital phantom data referenced by Quantitative Imaging Biomarkers Alliance (QIBA). This validation demonstrated good implementation of the General Kinetic Model."

    This implies the acceptance criterion for the phantom study was the accurate implementation of the General Kinetic Model (GKM) as validated against QIBA reference data. The reported performance is that the "validation demonstrated good implementation" of this model.

    For clinical data, the document states, "anonymized MR contrast-enhanced images were used as clinical datasets to validate the GenIQ application." Again, specific performance metrics against an acceptance criterion are not detailed, but the overall conclusion is that "GE Healthcare considers the GenIQ application to be as safe, as effective, and performance is substantially equivalent to the predicate device." This suggests the clinical validation aimed to demonstrate performance comparable to its predicate (Philips MR Permeability Software - K130278), which would inherently imply meeting certain effectiveness and safety standards.

    Table of Acceptance Criteria and Reported Device Performance (Inferred):

    Acceptance Criteria CategorySpecific Metric/TargetReported Device Performance
    Phantom Data ValidationAccurate implementation of the General Kinetic Model (GKM) as referenced by QIBA."Demonstrated good implementation of the General Kinetic Model."
    Clinical Data ValidationPerformance comparable to the predicate device (Philips MR Permeability Software - K130278) in terms of efficacy and safety for assessing tissue vascular properties."Considered to be as safe, as effective, and performance is substantially equivalent to the predicate device."

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size for Test Set:
      • Digital Phantom Data: Not specified.
      • Clinical Datasets: Not specified, only described as "anonymized MR contrast-enhanced images."
    • Data Provenance: The document does not explicitly state the country of origin.
      • Digital Phantom Data: Referenced by the Quantitative Imaging Biomarkers Alliance (QIBA).
      • Clinical Data: Described as "anonymized MR contrast-enhanced images." The study is likely retrospective as it used existing "anonymized" images.

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    This information is not provided in the document. For the digital phantom data, the ground truth is inherently defined by the QIBA reference data and the mathematical model itself. For the clinical datasets, the method of establishing ground truth (e.g., expert consensus, pathology) and the number/qualifications of experts are not described.

    4. Adjudication Method (Test Set)

    This information is not provided in the document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC comparative effectiveness study is not explicitly mentioned. The validation focused on the software's ability to implement the GKM and its equivalence to a predicate device. The document states, "GenIQ provides information that when interpreted by a trained physician," implying human-in-the-loop, but there's no study comparing human readers with and without AI assistance to quantify an effect size.

    6. Standalone Performance

    The evaluation primarily describes the "GenIQ application" as an automated post-processing software option, suggesting a focus on its standalone (algorithm-only) performance in calculating pharmacokinetic parameters from DCE-MRI data. While the output is "interpreted by a trained physician," the validation described (GKM implementation on phantom data, and validation on clinical datasets) focuses on the algorithm's accuracy in producing these parameters.

    7. Type of Ground Truth Used

    • Digital Phantom Data: The ground truth was based on the Quantitative Imaging Biomarkers Alliance (QIBA) reference data and the theoretical correctness of the General Kinetic Model (GKM). These are essentially simulated/known data sets designed to test the mathematical implementation.
    • Clinical Datasets: The document does not specify the type of ground truth used for the clinical validation. It only states that images were used "to validate the GenIQ application." This could imply comparison to a reference standard established by expert consensus, other imaging modalities, or clinical outcomes, but it is not detailed.

    8. Sample Size for the Training Set

    This information is not provided in the document. The document describes testing and validation, but not the development or training of the software.

    9. How Ground Truth for the Training Set was Established

    This information is not provided in the document as the training set details are absent.