    The Smart Coil is indicated for the embolization of:
    · Intracranial aneurysms

    • · Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
    • · Arterial and venous embolizations in the peripheral vasculature
    Device Description

    The Smart Coil functions to selectively embolize targeted segments of the vasculature by packing a sufficient quantity of bare platinum coils to achieve occlusion in an identical fashion to the Penumbra Coil System. The Smart Coil System consists of three components: a Coil Implant attached to a Detachment Pusher and a Detachment Handle.

    AI/ML Overview

    The Penumbra Smart Coil is a medical device designed for the embolization of intracranial aneurysms, neurovascular abnormalities, and arterial and venous embolizations in the peripheral vasculature. The provided document focuses on demonstrating the substantial equivalence of the Penumbra Smart Coil to predicate devices through a series of non-clinical tests.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document presents several tables detailing acceptance criteria and results for various non-clinical tests. Due to the extensive nature of these tables, they are summarized below by category. Please refer to the original document for the full detailed tables.

    Test CategoryAcceptance Criteria (General)Reported Device Performance (General)
    Biocompatibility
    In vitro CytotoxicitySample extracts must yield cell lysis grade ≤ 2Grade: 0 (Non-toxic)
    SensitizationTest Group shall yield Grade < 1 score on Magnusson and Kligman scale (provided Control Grade < 1)Grade: 0 (No Sensitization response - Non-sensitizing)
    Irritation (Intracutaneous Reactivity)The difference in the mean test article and mean control score must be grade 1.0 or lowerGrade 0.0 or 0.2 difference (Non-irritant)
    Implant StudyOverall interpretation of biocompatibility exhibited by the test article based on clinical observations, and gross and microscopic analysis comparing the test article to the control, shall show the test article to be a non-irritant.Test article is considered a non-irritant (Non-irritant)
    Systemic Toxicity (Acute)Sample extracts must not cause significant biological reaction greater than control (e.g., death in ≥2 animals, toxic signs, >10% weight loss in ≥3 animals).No evidence of systemic toxicity (No deaths, no toxic signs, no weight loss >10% in any animals - Non-toxic)
    Material Mediated PyrogenSample extracts must not cause a total rise in body temperature of ≥ 0.5° CNon-pyrogenic: no animals had a temperature rise ≥ 0.5°C
    Sub-Chronic ToxicityParameters (death, body weight change, clinical signs, hematology, histopathology) will be reviewed for patterns of toxicity.Negative for systemic toxicity (No deaths, no significant differences in body weights, no abnormal clinical signs, no patterns of toxicity in hematology or histopathology - Non-toxic)
    Hemocompatibility (Thrombosis)Device must be non-thrombogenic after 4 hours in vivo when compared to a predicate device.Test device was not thrombogenic when compared to the predicate; exhibited acceptable interaction with blood (Non-hemolytic)
    Hemocompatibility (Coagulation - PT & PTT)Coagulation times must be similar to the predicate or negative control values using analysis of variance.Test article coagulation times are statistically similar to the predicate (p-value=1.000 for PT, p-value=0.105 for PTT)
    Hemocompatibility (Hemolysis)Sample extracts must be nonhemolytic (≤2% hemolytic index).Nonhemolytic (Hemolytic index ≤ 1.25%, Corrected hemolytic index ≤ 0.40%)
    Hemocompatibility (Complement Activation)The concentrations of C3a and SC5b-9 in the test samples are statistically similar to or lower than the predicate.C3a = similar at all time points; SC5b-9 = lower at 30, 60 minutes / similar at 90 minutes.
    Genotoxicity (Ames, Mouse Lymphoma, Mouse Micronucleus)Sample extracts must be non-mutagenic.Non-mutagenic
    Bench-top Testing
    Dimensional / Visual InspectionUnits meet all inspection criteria for release of finished goods (clinically acceptable product).Pass
    Simulated Use/Detachment ReliabilityUnits meet all inspection criteria for release of finished goods (clinically acceptable product).Pass
    Fatigue TestingThe Coil Implant shall retain its secondary shape after being cycled into/out of introducer sheath 5 times.Pass
    Torsion ResistanceThe Coil Implant can be subjected to 8 turns minimum.Pass
    Corrosion ResistanceNo visible corrosion after testing.Pass
    Friction TestingPush/pull friction acceptable through a 0.0165 in. ID microcatheter.Pass
    Distal System Tensile TestCoil Implant and Detachment Pusher joints has sufficient tensile strength.Pass
    Coil Dimensional Inspection (after simulated use testing)Units withstood the simulated use testing.Pass
    SterilizationIdentical acceptance criteria and testing methods as the predicate device (in accordance with EN ISO 11135-1:2007).Tested to be sterile.
    Shelf LifeMinimum one year shelf life (using identical acceptance criteria and testing methods as the predicate device, in accordance with ASTM D4169:2009 and ASTM F2096:2004).Tested to have a minimum one year shelf life.
    MR EnvironmentStandards ASTM F2182-11, ASTM F2052-06, ASTM F2213-06, and ASTM F2119-07 (R 13)Tested in the MR environment to advise the MR Conditional statement in the IFU. (Implies satisfactory performance for MR safety labeling) Specific performance values not in given text.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily describes non-clinical testing (bench-top and in vitro/animal studies) for biocompatibility, physical/mechanical properties, sterilization, and shelf life.

    • Sample Size: The exact sample sizes for each specific test are not explicitly stated in the summary. For example, for the systemic toxicity test, it mentions "Death in 2 or more animals" as an acceptance criterion, but not the total number of animals tested. The biocompatibility tests refer to "samples," "test articles," "animals," and "groups."
    • Data Provenance: Not applicable in the context of clinical data for a device intended for human use. The studies are non-clinical (laboratory and animal studies). The document does not specify the country of origin for the non-clinical data, but it is submitted to the U.S. FDA by a U.S.-based company (Penumbra, Inc., Alameda, CA, USA). The studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices, which are US federal regulations.
    • Retrospective/Prospective: These non-clinical studies are inherently prospective in their design, meaning they were planned and executed to evaluate the device's properties.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable to the provided document. The studies described are non-clinical (biocompatibility, mechanical testing, etc.) and do not involve human subjects or expert clinical evaluation to establish a "ground truth" for diagnostic or treatment efficacy. The "ground truth" for these non-clinical tests would be the measured physical, chemical, or biological responses against predefined scientific or regulatory standards. For example, in histopathology for implant studies, a pathologist would interpret slides, but their role is to report observations against scientific criteria, not establish a "ground truth" for a clinical diagnosis.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods like 2+1 or 3+1 typically refer to a process where multiple clinical readers (e.g., radiologists) review cases and a discrepancy resolution method is employed, often involving a super-reader. The studies described are non-clinical and do not involve human clinical readers or case adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. The Penumbra Smart Coil is a physical neurovascular embolization device, not an AI-powered diagnostic or assistive tool for human readers. The document describes non-clinical testing of the device's material properties, mechanical performance, and biological safety, not its impact on human reader performance in interpreting medical images.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The Penumbra Smart Coil is a physical medical device, not an algorithm. Therefore, "standalone" algorithm performance is not a relevant concept for this device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the non-clinical studies are established by:

    • Scientific and regulatory standards: e.g., EN ISO 10993 for biocompatibility, ASTM standards for material testing, EN ISO 11135-1 for sterilization.
    • Measured physical/chemical properties: e.g., tensile strength, corrosion presence, hemolytic index.
    • Biological responses in controlled in vitro or animal models: e.g., cell lysis grade, sensitization scores, animal physiological responses, histopathological findings.
    • Comparison to predicate devices: For several tests (e.g., hemocompatibility, sterilization, shelf life), performance is compared to or demonstrated using methods identical to predicate devices.

    8. The sample size for the training set

    This question is not applicable. There is no "training set" in the context of physical medical device testing for 510(k) clearance, as this is not a machine learning or AI device. The document describes pre-market non-clinical validation studies.

    9. How the ground truth for the training set was established

    This question is not applicable for the same reason as point 8. No "training set" exists for this type of device submission.