510(k) Data Aggregation

    K Number
    K142717

    Validate with FDA (Live)

    Date Cleared
    2015-05-07

    (226 days)

    Product Code
    Regulation Number
    886.5928
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For storage of soft (hydrophilic), hard, and rigid gas permeable contact lenses during chemical disinfection. Use for storage during chemical disinfections only. DO NOT USE WITH HEAT.

    Device Description

    The contact lens cases in submission are intended to be used to store soft (hydrophilic) and hard contact lenses during chemical disinfection only. These are not to be used in heat disinfection.

    There are 2 parts to each contact lens case, the main body/base with two wells, and two screw-on lids. In order to store contact lenses with disinfection solution, the wells for storage need to be leak-proof, which is designed so by the screw construction of the lids.

    The plastic that is used in all four contact lens cases is Acrylonitrile Butadene Styrene (Product Name: Polylac ® PA-707, manufactured by Chi Mei Corporation, please refer MSDS attached: Attachment 1), which is a durable plastic that is insoluble in water and can be used for at least two months continuous use.

    The cat contact lens case is black, with white printed on the lids. The color pigment used in the case is PLASBLAK® UN2014 (Supplier: Cabot Plastics, please refer MSDS attached: Attachment 2). This pigment is insoluble in water.

    The sunglasses contact lens case is white, with black labels on the lids. The color pigment used in the case is DuPont™ Titanium Dioxide Pigment - Plastics Grade/Type : R-350 (Supplier: DuPont, please refer MSDS attached: Attachment 3). This pigment is insoluble in water.

    The 1st contact lens case is green, with clear uncolored lids. There are 2 color pigments used in the green body (1) PV Fast Green GNX (Supplier: Clariant Chemicals (India) Limited, please refer MSDS attached: Attachment 4) and (2) MACROLEX Yellow G Granulate (Supplier: Bayer AG, please refer MSDS attached: Attachment 5). These pigments are insoluble in water.

    The Geeky eyes contact lens case is red with black glasses. The color pigment used in the red case is 13575 Orange Color Powder (Supplier: Summit Color Co., Ltd., please refer MSDS attached: Attachment 6). This pigment is insoluble in water.

    The individual well capacity of each contact lens are: cat contact lens case 2.8ml, sunglasses contact lens case 2.5ml, 1* contact lens case 3ml, geeky eyes contact lens case 2.8ml. The capacities aforementioned are sufficient for a contact lens to be fully soaked in the disinfection liquid.

    At the base of the cat contact lens case and the sunglasses contact lens case, in each well there is L and R embossed to help users differentiate the left and right lenses. The geeky eyes contact lens case also has the L and R markings on the side of the casing outer. The 1d contact lens case does not have L and R markings on it as the two wells are situated one above and one below the disc so it is unlikely to mix up the two wells during use.

    Through use, the material of the contact lens case may come into contact with the user's eyes indirectly. The user will need to put contact lenses with chemical disinfecting liquid into the contact lens case for storage. Therefore we have used a plastic material and colorants that are insoluble in water so that it will not come in contact with one's eyes.

    These contact lens cases are intended for the storage of soft (hydrophilic) and hard contact lenses. The case should ensure the fluids contained do not leak. These cases are for the storage of chemical disinfecting fluids with contact lenses only, they are not to be used in heat disinfection. These are not sold sterile, so the packaging should have instructions written on them to "Rinse the case thoroughly with clean solution before use".

    It is recommended to replace the contact lens case at least every 3 months.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for several contact lens cases. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study to prove acceptance criteria for a novel device. Therefore, much of the requested information regarding acceptance criteria, performance studies, sample sizes, expert ground truth, and MRMC studies is not directly applicable or available in this document.

    However, I can extract the relevant information and indicate where data is not provided for questions that are typically relevant to AI/diagnostic device submissions.

    The Device:

    Ni Hau Industrial Co., Ltd. Contact Lens Cases (Cat Contact Lens Case, Sunglasses Contact Lens Case, 1st Contact Lens Case, Geeky Eyes Contact Lens Case)

    Acceptance Criteria and Reported Device Performance

    The document doesn't define quantitative acceptance criteria in terms of metrics like sensitivity, specificity, accuracy, or other performance characteristics typically associated with AI or diagnostic devices. Instead, the acceptance criteria are based on demonstrating substantial equivalence to a legally marketed predicate device by fulfilling manufacturing, material, and functional requirements.

    The "reported device performance" is implicitly that the device meets the same safety and performance characteristics as the predicate device. The comparison focuses on physical properties, intended use, and biocompatibility.

    Acceptance Criteria CategorySpecific Criteria (Implicit/Explicit)Reported Device Performance (as per document)
    Intended UseFor storage during chemical disinfection of soft (hydrophilic), hard, and rigid gas permeable contact lenses. Not for heat disinfection.Matches predicate device's intended use.
    Device MaterialAcrylonitrile Butadiene Styrene (ABS)All 4 devices use ABS (Polylac ® PA-707). Matches predicate.
    Device ConstructionShaped base with two wells and two screw-on lids for leak-proof storage.All 4 devices have shaped base with two wells and two screw lids. Matches predicate.
    BiocompatibilityMust pass relevant ISO 10993 tests (Cytotoxicity, Eye Irritation, Systemic Toxicity).All 4 devices "PASS the following test: Cytotoxicity (ISO 10993-5), Eye Irritation (ISO 10993-10), Systematic Toxicity (ISO 10993-11)." Matches predicate.
    SterilizationNot sold sterile.Not sold sterile. Matches predicate.
    Well Capacity (Sufficiency)Sufficient volume for contact lenses to be fully soaked in disinfection liquid.Well capacities range from 2.5ml to 3ml, stated as "sufficient for a contact lens to be fully soaked." Predicate device wells are 4.8ml-5.5ml, also sufficient.
    Material InsolubilityPlastic material and colorants used are insoluble in water to prevent contact with eyes.ABS plastic and all specified color pigments (PLASBLAK® UN2014, DuPont™ Titanium Dioxide, PV Fast Green GNX, MACROLEX Yellow G Granulate, 13575 Orange Color Powder) are stated as insoluble in water.
    Leak-proof DesignWells for storage need to be leak-proof.Designed with screw construction of the lids to be leak-proof.
    Recommended ReplacementManufacturers recommendation to replace cases every 3 months.Stated: "It is recommended to replace the contact lens case at least every 3 months."
    LabelingInstructions to "Rinse the case thoroughly with clean solution before use."Packaging should have these instructions.

    Study Information

    The document describes a submission for substantial equivalence based on a comparison to a predicate device, not a de novo study demonstrating the clinical performance of a diagnostic or AI device. Therefore, many of the typical study parameters for such devices are not applicable.

    1. Sample size used for the test set and the data provenance: Not applicable. This is not a study assessing diagnostic performance or algorithm output. The "test set" would implicitly refer to the contact lens case prototypes themselves. The provenance of the materials (ABS plastic, pigments) is provided by manufacturer and product name (e.g., Chi Mei Corporation, Cabot Plastics, DuPont), and the manufacturing location is Taiwan.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a diagnostic output is not established here. Compliance with international standards (e.g., ISO 10993 for biocompatibility) is used as evidence of safety, which would be evaluated by qualified testing laboratories.

    3. Adjudication method for the test set: Not applicable. No diagnostic outputs to adjudicate.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI device.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI device.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): The "ground truth" here is compliance with established device safety and manufacturing standards, and demonstrated equivalence to a legally marketed predicate device. This includes:

      • Material specifications (e.g., ABS composition, insolubility of pigments).
      • Physical design specifications (e.g., screw lids for leak-proofing, well capacity).
      • Biocompatibility test results (passing ISO 10993-5, 10993-10, 10993-11). These lab tests provide objective data as "ground truth" for the material's biological safety.
    7. The sample size for the training set: Not applicable. This is not an AI device.

    8. How the ground truth for the training set was established: Not applicable. This is not an AI device.

    Ask a Question