510(k) Data Aggregation

    K Number
    K141991

    Validate with FDA (Live)

    Date Cleared
    2015-03-31

    (252 days)

    Product Code
    Regulation Number
    876.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These instruments (sphincterotomes and guidewires) have been designed to be used with an Olympus endoscope for papillotomy using high-frequency current. The preloaded guidewire is used for guiding and exchanging endoscopic accessories for billary duct, including but not limited to the common bile, cystic, pancreatic and right and left hepatic ducts.

    Device Description

    The subject devices consist of the papillotomy knife for endoscopic sphincterotomy. The predicate device consists of a closed quidewire lumen, and the subject devices consist of a partially open quidewire lumen. The partially open guidewire lumen is called the C-Channel design, and the exchange operation of the quidewire is added.

    The subject Single Use Preloaded Sphincterotome V(Distal Wireguided) has a preloaded guidewire, and the subject Single Use Sphincterotome V(Distal Wireguided) does not have a preloaded guidewire. The preloaded guidewire of the subject device is a minor modification of the product from Terumo Corporation that has been cleared via premarket notification, K091417.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device (Single Use Preloaded Sphincterotome V and Single Use Sphincterotome V). It primarily focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a detailed study proving the device meets specific acceptance criteria in the context of AI or diagnostic performance. Therefore, many of the requested items related to AI device performance, sample sizes for training/test sets, expert ground truth, and MRMC studies cannot be extracted from this document.

    However, I can extract information regarding non-clinical performance and the general approach to demonstrating safety and effectiveness.

    Here's an attempt to answer based on the provided text, recognizing the limitations:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance
    Basic PerformanceConfirmed that the subject device works as intended.
    Risk Analysis (ISO 14971:2007)Design verification tests and their acceptance criteria were identified and performed, demonstrating an acceptable risk profile.
    Applicable Standards ComplianceThe device was tested against and complies with the following standards: - IEC 60601-1 - IEC 60601-1-2 - IEC 60601-2-2 - IEC 60601-2-18 - ISO 10993-1 - ISO 10993-5 - ISO 10993-10 - ISO 10993-11 - ISO 11135-1 - ISO 14971 - ASTM F-1980-07
    Comparison to Predicate DeviceNo significant changes in intended use, method of operation, material, or design that could affect safety or effectiveness compared to the predicate device. Minor modification of the guidewire does not affect safety and effectiveness.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not specified. The document refers to "Performance testing" but does not detail the number of devices tested or the specific test parameters.
    • Data Provenance: Not applicable in the context of diagnostic data. The testing mentioned is engineering/performance testing of the physical medical device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable: This document describes a physical medical device (sphincterotome) and its engineering performance and safety, not an AI or diagnostic device that requires expert-established ground truth for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable: This pertains to clinical or diagnostic studies, not the engineering performance testing described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: This document is for a physical surgical accessory, not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable: This document is for a physical surgical accessory, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For the non-clinical performance testing, the "ground truth" would be established engineering specifications, performance standards as defined in ISO and IEC documents, and the functional characteristics of the predicate device. It's not a diagnostic "ground truth" like pathology or expert consensus.

    8. The sample size for the training set

    • Not Applicable: This document does not describe a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    • Not Applicable: This document does not describe a machine learning or AI device that requires a training set.

    In summary: The provided 510(k) notification focuses on the manufacturing, design, and non-clinical performance (e.g., electrical safety, biocompatibility, sterilization, functional testing) of a physical medical device to demonstrate substantial equivalence to a legally marketed predicate. It does not contain information relevant to AI/ML device acceptance criteria or studies involving diagnostic performance, expert ground truth, or statistical measures like sample size for AI training/test sets.