510(k) Data Aggregation

    K Number
    K141859

    Validate with FDA (Live)

    Date Cleared
    2014-08-01

    (22 days)

    Product Code
    Regulation Number
    866.3980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Influenza A/H5 Subtyping Kit contains reagents and controls of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel and is intended for use in real-time RT-PCR (rRT-PCR) assays on an Applied Biosystems (ABI) 7500 Fast Dx Real Time PCR Instrument in conjunction with clinical and epidemiological information:

    · For the presumptive identification of virus in patients who may be infected with influenza A subtype A/H5(Asian Lineage) from viral RNA in human respiratory specimens and viral culture in conjunction with clinical and epidemiological risk factors;

    · To provide epidemiologic information for surveillance of circulating influenza viruses.

    Device Description

    The Influenza A/H5 Subtyping Kit contains components of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel that is used in real-time RT-PCR (rRT-PCR) assays on the ABI 7500 Fast Dx Real-Time PCR Instrument. The Influenza A/H5 Subtyping Kit consists of oligonucleotide primers and dual-labeled hydrolysis (TaqMan®) probes to be used in rRT-PCR for the in vitro qualitative detection and characterization of human influenza viruses from viral RNA in respiratory specimens from patients presenting with influenza-like illness (ILI).

    The Influenza A/H5 Subtyping Kit is based on technology which is used in many molecular diagnostic assays. rRT-PCR assays are one-tube assays that first reverse-transcribe specific regions of RNA into cDNA copies. The cDNA then serves as a template for a polymerase chain reaction that utilizes a thermocyclic heating and cooling of the reaction to logarithmically amplify a specific region of DNA. The probe anneals to a specific internal target sequence located between the target loci of the forward and reverse primers. During the extension phase of the PCR cycle, the 5' nuclease activity of Tag polymerase degrades any probe molecules hybridized to amplified target sequence, causing the reporter dye to separate from the quencher dye, and generating a fluorescent signal. With each cycle, additional reporter dye molecules are cleaved from their respective probes, increasing the fluorescence intensity. Fluorescence intensity is monitored at each PCR cycle. Amplification of targets is reflected by logarithmic increase in fluorescence over time in comparison to the background signal.

    AI/ML Overview

    This document is a 510(k) premarket notification from the FDA, asserting substantial equivalence for the "CDC Human Influenza Virus Real-time RT-PCR Diagnostic Panel, Influenza A/H5 Subtyping Kit." The document does not contain acceptance criteria or a study proving device performance against such criteria. Instead, it compares the proposed device to a predicate device (K132508) and concludes that the changes are primarily for labeling and do not alter the device's design, technological attributes, or intended use.

    Therefore, I cannot provide the requested information from this document. The document explicitly states:

    • "The changes proposed to the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel subject to this special 510(k) are for labeling purposes only and will not alter the technological attributes of the device." (Page 6)
    • "The changes proposed to the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel do not alter the device's design or technological attributes. In addition, the indications for use and intended use of the CDC Human Influenza Virus Real-Time RT-PCR Diagnostic Panel will remain the same." (Page 9)

    This type of 510(k) submission (a Special 510(k)) is used for modifications to a manufacturer's own legally marketed device where the modification does not affect the device's fundamental scientific technology or its intended use. As such, it relies on the previous clearance of the predicate device (K132508) for performance data rather than presenting new performance studies against acceptance criteria.