Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K141854

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-10-29

    (477 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OSUNG INSTRUMENT CASSETTE (EFCCL1-F) is intended to hold instruments and accessories during storage and steam sterilization. This case cassette is intended to be used with a sterilized wrap in order to maintain sterility.

    This case cassette is intended to be used with an FDA cleared sterilization wrap in order to maintain sterility.

    Validated sterilization parameters:

    Cycle TypeTemperatureExposure TimeDry Time
    Gravity Steam132℃ (270°F)15 minutes30 minutes
    Device Description

    The OSUNG INSTRUMENT CASSETTE is designed to handle reusable dental instruments for sterilization and storage. This device is intended to hold instruments and accessories during storage and steam sterilization, and to be used with a sterilized wrap in order to maintain sterility.

    The OSUNG INSTRUMENT CASSETTE is comprised of a base, lid, silicone mat/rail, holders, and spring lock system.

    The internal insert cassette has the ability to hold individualized pieces and accessories which include scissors, needle, elevators, forceps, pincers, retractors, mallets etc.

    AI/ML Overview

    The provided text does not describe an AI/ML powered device, but rather a medical device for sterilization—an "Instrument Cassette." Therefore, many of the requested fields (such as MRMC study, effect size of human improvement with AI, training set size, etc.) are not applicable and cannot be found in the document.

    However, I can extract information related to the acceptance criteria and the study that proves the device meets them based on the provided text.

    Acceptance Criteria and Device Performance for Instrument Cassette (EFCCL1-F)

    1. A table of acceptance criteria and the reported device performance

    Performance Test / Acceptance CriteriaReported Device Performance
    Half cycle sterilization validation at 132°C (per ISO 17665-1, ISO 17665-2)No growth at half cycle.
    Dry Time Validation (per ANSI/AAMI ST79 & A1, A2, A3, A4)The minimum weight difference was -0.05% (Pass).
    Cleaning Validation (per AAMI TIR 30)Results are not explicitly stated in the summary table but implied to be compliant with "All of tests demonstrated compliance to both safety and effectiveness requirements that apply to the device."
    Limits of Reuse (per FDA Guideline: "Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" & AAMI TIR12, AAMI TIR30)Results are not explicitly stated in the summary table but implied to be compliant with "All of tests demonstrated compliance to both safety and effectiveness requirements that apply to the device."
    Overall Safety and EffectivenessDemonstrated compliance to both safety and effectiveness requirements, including sterilization performance, maintenance of sterility, and reusability.
    Substantial EquivalenceSubstantially equivalent to the predicate device, Miltex Thompson Cassettes (K101653), based on intended use, indications for use, technological characteristics, and performance data.
    Validated Sterilization ParametersGravity Steam: 132°C (270°F) temperature, 15 minutes exposure time, 30 minutes dry time.

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample sizes used for the performance tests (e.g., number of cassettes tested, number of cycles performed). Data provenance is also not detailed regarding country of origin or whether studies were retrospective or prospective, beyond stating "Validation test for steam sterilization of ultrasonic cleaning and limit of reuse test for the OSUNG INSTRUMENT CASSETTE (EFCCL1-F) has been carried out as the results were shown by test reports."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a physical instrument cassette requiring performance testing against standards, not expert interpretation of outputs to establish ground truth like in an AI/ML context.

    4. Adjudication method for the test set

    Not applicable. There is no mention of adjudication for establishing ground truth, as the performance tests are objective measurements against established standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-powered device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an AI-powered device.

    7. The type of ground truth used

    The "ground truth" or acceptance criteria for this device are based on established industry standards and FDA guidelines for sterilization, cleaning validation, and reusability of medical devices. Specifically, the following standards were used:

    • ISO 17665-1(2006)
    • ISO 17665-2(2009)
    • ANSI/AAMI ST79(2010) & A1(2010) & A2(2011) & A3(2012) & A4(2013)
    • AAMI TIR 30 (2011)
    • FDA Guideline: "Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling (2011)"
    • AAMI TIR12(2010)

    8. The sample size for the training set

    Not applicable. This is not an AI-powered device subject to machine learning training.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI-powered device subject to machine learning training.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1