    Intended Use

    The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

    Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.

    With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.

    When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e. scoliosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    The CD HORIZON® SPIRE™ Plate is a posterior, single level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis; trauma; and/or tumor.

    In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

    Device Description

    The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® Plates, staples, and connecting components, as well as implant components from other Medtronic spinal systems which can be rigidly locked into a variety of configurations with each construct being tailor-made for the individual case.

    A subset of CD HORIZON® Spinal System components may be used for posterior pedicle screw fixation in pediatric cases. These constructs may be comprised of a variety of shapes and sizes of rods (ranging in diameter from 3.5mm to 6.35mm), hooks, screws, CROSSLINK® Plates and connecting components. Similarly to the CD HORIZON® implants used in adult cases, these components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

    The purpose of this submission is to modify Medtronic's CD HORIZON® Spinal System to add additional components to the system, specifically a Tab Extender and Extender Cap.

    AI/ML Overview

    The provided document is a 510(k) Summary for the CD HORIZON® Spinal System, specifically for the addition of Tab Extender and Extender Cap components. It states explicitly that no clinical testing was performed for this submission.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, and ground truth establishment cannot be provided as these elements are associated with clinical studies that were not conducted for this device modification.

    Instead, the submission relies on the concept of substantial equivalence to a previously cleared predicate device (CD HORIZON® Spinal System K122862, cleared on 10/03/2012). The company performed a risk analysis and engineering rationale to demonstrate that the new components do not introduce new issues of safety or effectiveness.

    Here's a summary of what the document does provide in relation to the prompts:

    Information RequestedProvided by Document
    1. Table of acceptance criteria and device performanceNot applicable. No acceptance criteria or performance metrics derived from a study are provided because no clinical testing was performed. The basis for clearance is substantial equivalence.
    2. Sample size and data provenance (test set)Not applicable. No clinical test set. The submission relies on a risk analysis and engineering rationale for the modified components and comparison to the predicate device.
    3. Number and qualifications of experts (ground truth)Not applicable. No clinical ground truth was established by experts for a test set in this submission.
    4. Adjudication method (test set)Not applicable. No clinical test set requiring adjudication was used.
    5. MRMC comparative effectiveness study (AI vs. human)Not applicable. This device is a spinal implant system, not an AI-assisted diagnostic or treatment tool. No such study was conducted.
    6. Standalone (algorithm only) performance studyNot applicable. This device is a physical spinal implant system, not a software algorithm.
    7. Type of ground truth usedNot applicable. No clinical ground truth was established. The basis for approval is substantial equivalence to a predicate device, supported by non-clinical risk analysis and engineering rationale.
    8. Sample size for the training setNot applicable. No training set was used as no machine learning or AI algorithm is involved.
    9. How ground truth for training set was establishedNot applicable. No training set was used.

    Key takeaway: The regulatory clearance for these new components (Tab Extender and Extender Cap) for the CD HORIZON® Spinal System was based on demonstrating substantial equivalence to an existing, legally marketed predicate device (K122862) through non-clinical testing (risk analysis and engineering rationale), rather than through clinical studies.