510(k) Data Aggregation

    K Number
    K141594

    Validate with FDA (Live)

    Date Cleared
    2015-08-10

    (423 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unimicro Trocar kit,Model: Auto-Locking Trocar, and Bladeless Trocar has application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments.

    Device Description

    The Auto-Locking Trocar has application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments. The Auto-Locking Trocar is available in three (3) sizes: 5mm, 10mm and 12mm. The cannula assembly has a universal seal ,a valve,and a stopcock.This device has a bladed tip with an internal shield, which is designed to cover the cutting edges once the body cavity has been entered.

    The Hasson Trocar has application in a variety of endoscopic procedures to provide a port for entry for endoscopic instruments. The Hasson Trocar is available in two (2) sizes: 10mm and 12mm.This device has a blunt tip, which is designed for open Laparoscopy. The cannula assembly has a fixation device, a universal seal, a valve, and a stopcock.

    The Bladeless Trocar has application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments. The Bladeless Trocar is available from 5-12mm: The cannula assembly (10mm to 12mm type) has a universal seal, a valve, and a stopcock. The cannula assembly (5mm type) has a stopcock.

    AI/ML Overview

    The document is a 510(k) summary for the Unimicro Trocar Kit. It describes the device, its intended use, and a comparison to a predicate device to establish substantial equivalence.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (implied/general)Reported Device Performance
    Obturator CompatibilityMet requirements of compatibility with the cannulaAll test results demonstrate Unimicro Trocar kit meet the requirements of its pre-defined acceptance criteria
    Insertion & Cannula StabilityMet requirements for stable insertion and cannula positionAll test results demonstrate Unimicro Trocar kit meet the requirements of its pre-defined acceptance criteria
    Air LeakageDevice effectively prevents air leakageAll test results demonstrate Unimicro Trocar kit meet the requirements of its pre-defined acceptance criteria
    Biological Safety (ISO 10993 series)Biocompatibility requirements metTests conducted in accordance with ISO 10993-1, ISO 10993-5, ISO 10993-7, ISO 10993-10, ISO 10993-12
    Sterilization (ISO 11135-1)Sterilization effectiveness metTests conducted in accordance with ISO 11135-1

    Note: The document states "All the test results demonstrate Unimicro Trocar kit meet the requirements of its pre-defined acceptance criteria and intended uses." However, the specific numerical or qualitative acceptance criteria for each performance test (Obturator Compatibility, Insertion & Cannula Stability, Air Leakage) are not explicitly stated in this summary. It only indicates that they were met.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated. The document mentions "A series of safety tests were performed" and "The performance testing conducted on subject device and predicate device," but does not provide specific numbers for the devices tested.
    • Data Provenance: The tests were non-clinical ("non-clinical testing") and likely conducted in-house or by a third-party lab affiliated with the manufacturer, Unimicro Medical Systems (ShenZhen) Co., Ltd. The manufacturer is based in Shenzhen, Guangdong Province, China. The data would therefore be retrospective in the context of the 510(k) submission, meaning the tests were completed before the submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This device is a medical instrument (trocar kit), not a diagnostic algorithm or image analysis device that requires expert ground truth for its performance evaluation in the way medical AI would. The performance is assessed through engineering and biocompatibility testing.

    4. Adjudication Method for the Test Set

    Not applicable. As this is not a diagnostic or AI-driven device, there is no "ground truth" established by human experts that would require adjudication. Performance is determined by objective physical and biological tests against established technical standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. The Unimicro Trocar Kit is a physical medical device, not an AI or diagnostic tool that would involve human readers or image interpretation.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. The Unimicro Trocar Kit is a physical medical device, not an algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for the performance of the Unimicro Trocar Kit is established through adherence to recognized international standards (e.g., ISO 10993 series for biological evaluation, ISO 11135-1 for sterilization) and pre-defined engineering acceptance criteria for functional performance (e.g., obturator compatibility, insertion/cannula stability, air leakage). This is a technical ground truth, not an "expert consensus," "pathology," or "outcomes data" in the typical sense for diagnostic devices.

    8. Sample Size for the Training Set

    Not applicable. This device does not involve machine learning or AI, and therefore does not have a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for this device.