Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K141593

    Validate with FDA (Live)

    Date Cleared
    2014-11-13

    (153 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unimicro Disposable Retrieval Endo-Pouch is intended for the collection and extraction of tissue, organs and calculi during general and laparoscopic surgical procedures.

    Device Description

    The Unimicro Disposable Retrieval Endo-Pouch is a sterile and single-use specimen container designed for use in retrieving specimens during endoscopic surgery. The Unimicro Disposable Retrieval Endo-Pouch is supplied in a dispending tube for ease of insertion through a standard 10,11or 12mm trocar sheath.

    AI/ML Overview

    The document is a 510(k) premarket notification for the Unimicro Disposable Retrieval Endo-Pouch. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel effectiveness through clinical studies with acceptance criteria based on outcome measures.

    Therefore, the information typically requested in your prompt (such as acceptance criteria for device performance in terms of accuracy, sensitivity, specificity, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and how ground truth for training data was established) is not applicable to this 510(k) submission for a medical device that retrieves tissues, organs, and calculi during surgical procedures.

    Instead, the submission relies on demonstrating similarity in intended use, design, principles of operation, materials, performance, and sterilization to a predicate device. The "acceptance criteria" here refers to the successful completion of non-clinical (bench) testing to ensure the device performs as expected and meets safety requirements.

    Here's a breakdown of the relevant information from the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document states: "A series of preclinical physical, mechanical and biocompatibility tests were performed to assess the safety and effectiveness of the Unimicro Disposable Retrieval Endo-Pouch. All the test results demonstrate the performance of Unimicro Disposable Retrieval Endo-Pouch meets the requirements of its predefined acceptance criteria and intended uses."

    While the specific acceptance criteria and detailed quantitative results for each test are not provided in this summary document, the general categories of tests and the conclusion are stated.

    Category of TestAcceptance Criteria (Implicit)Reported Device Performance
    Physical TestsDevice functions as intended without physical failure during simulated use or applicable stress.Meets requirements of predefined acceptance criteria and intended uses.
    Mechanical TestsDevice maintains structural integrity and performs mechanical actions (e.g., deployment, retrieval strength) as designed.Meets requirements of predefined acceptance criteria and intended uses.
    Biocompatibility Tests (ISO 10993 series)Device materials are non-toxic, non-irritating, and non-sensitizing as per the specified ISO standards: ISO 10993-1:2009 (Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process) ISO 10993-5:2009 (Tests for in vitro cytotoxicity) ISO 10993-10:2010 (Tests for irritation and skin sensitization) ISO 10993-12:2012 (Sample preparation and reference materials)Meets requirements of predefined acceptance criteria (by adhering to ISO standards) and intended uses.
    Sterilization (EO Sterilized)Device achieves and maintains sterility.EO Sterilized (implies successful sterilization validation).

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified. Standard practice for non-clinical bench testing generally involves a sufficient number of units to demonstrate reproducibility and robustness, but specific numbers are not typically included in a 510(k) summary.
    • Data Provenance: Not explicitly stated, but these are bench tests conducted by the manufacturer, Unimicro Medical Systems (ShenZhen) Co., Ltd., in China. The data would therefore be prospective in the sense that the tests were conducted specifically for this submission, but not in the clinical sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This is a non-clinical device that collects specimens. There is no "ground truth" to be established by clinical experts in the context of the described bench tests. The tests evaluate physical, mechanical, and biological properties according to engineering and biocompatibility standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. As these are non-clinical bench tests evaluating device properties, there is no need for expert adjudication of results in the way it would be applied to a diagnostic imaging algorithm. Test results are typically objective measurements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is a device for specimen retrieval, not a diagnostic AI device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a medical device, not an algorithm. Bench testing evaluates the device itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not Applicable. For non-clinical bench testing, the "ground truth" or reference standard would be the pre-defined engineering specifications, material properties, and biocompatibility standards (e.g., ISO 10993 suite). These are objective criteria.

    8. The sample size for the training set:

    • Not Applicable. There is no "training set" as this is a physical medical device, not a machine learning model.

    9. How the ground truth for the training set was established:

    • Not Applicable. (See point 8).

    Summary of Equivalence (rather than direct performance against clinical criteria):

    The submission demonstrates substantial equivalence by comparing the Unimicro Disposable Retrieval Endo-Pouch to the predicate device, Unimax Specimen Retrieval System (K103510), across several parameters:

    • Intended Use: Identical ("receptacle for the collection and extraction of tissue, organs and calculi during general and laparoscopic surgical procedures.")
    • Design: Both consist of a flexible polymer bag and an introducer structure that fits through a trocar port.
    • Principles of Operation: Implied to be identical given design and intended use.
    • Materials: Both use "Various polymer Stainless."
    • Performance: Supported by the statement "The result of bench testing indicates that the new device is as safe and effective as the predicate device" and that the non-clinical tests "demonstrate the performance... meets the requirements of its predefined acceptance criteria and intended uses."
    • Sterilization: Both are EO Sterilized.
    • Biocompatibility: Both adhere to ISO 10993 standards (with slightly updated versions for the proposed device, reflecting current standards).

    In conclusion, this 510(k) submission establishes equivalence through non-clinical testing against predefined engineering and biocompatibility standards, rather than through clinical studies with performance metrics like sensitivity/specificity or expert-adjudicated ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1