510(k) Data Aggregation

    K Number
    K141425

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-09-11

    (104 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    755 nm:
    The Cynosure Elite+ Laser is indicated for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowting when measured 6, 9, or 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I-VI) including tanned skin. It is also indicated for the treatment of vascular lesions, benign pigmented lesions, and wrinkles.

    1064 nm:
    The Cynosure Elite+ Laser is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangectasia, rosacea, venus lake, leg veins, spider veins and poikiloderma of civatte; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis. The laser is also intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, sebortheic keratoses, nevi, chloasma, verrucae, skin tags, keratosis and plaques. The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles. Additionally, the laser is indicated for the treatment of pseudofolliculitis barbae (PFB) and for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

    Device Description

    Cynosure Elite+ laser has an Alexandrite crystal rod as the lasing medium, with a wavelength of 755 nm. The 1064 nm wavelength is generated from the Nd:YAG laser head.

    Laser activation is by either by finger or foot switch. The overall weight of the laser is 180 lbs and the size is 41 x 15 x 25 in (HxWxD).

    Electrical equipment is 208/240 VAC, single phase, 30A, 50/60 Hz.

    The modifications to this device are four handpieces allowing for a larger treatment area; 18 mm, 20 mm, 22 mm and 24 mm. The 1064 Nd:YAG laser did not require any modifications to support the new handpieces. Minor modifications, including rod diameter change and lamp fill pressure change, were made to the Alexandrite laser to increase the energy to support the larger spot sizes. The software continues to control the energy based on the handpiece attached. This prevents the increase in available energy from the laser from increasing the fluence of the existing spot sizes.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Cynosure Elite+ Laser, seeking clearance for a modified device. It focuses on demonstrating substantial equivalence to a predicate device, not on presenting novel clinical study data to prove the device meets specific acceptance criteria in a clinical setting.

    Therefore, the requested information regarding acceptance criteria and a study proving the device meets those criteria, specifically concerning clinical performance metrics, cannot be fully extracted from this document in the way it would be for a direct clinical performance study.

    However, based on the provided text, I can infer the acceptance criteria relate to the engineering and software performance specifications and that the study proving acceptance was non-clinical testing.

    Here's what can be extracted and inferred:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Inferred from predicate device specifications)Reported Device Performance (Elite+ K141425)
    Laser Type: Alexandrite and Nd:YAG LasersAlexandrite and Nd:YAG Lasers
    Wavelength: 755 nm & 1064 nm755 nm & 1064 nm
    Maximum Fluence (755nm): 60 J/cm²60 J/cm²
    Maximum Fluence (1064nm): 300 J/cm²300 J/cm²
    Repetition Rate (755nm): 1 to 5 Hz1 to 5 Hz
    Repetition Rate (1064nm): 1 to 10 Hz1 to 10 Hz
    Pulse Duration: 0.1-300 ms0.1-300 ms
    Spot Sizes (mm): (Predicate: 3, 5, 7, 10, 12.5, 15)3, 5, 7, 10, 12, 15, 18, 20, 22, 24 (The key modification)
    Software Functionality: Meet software specificationsTesting confirmed that the software specifications were met.
    Overall Performance: Device performs within the same specifications of the predicate deviceTesting confirmed the performance of the Elite+ Laser meets the product system requirements, which is based on the predicate device.

    Important Note: The "acceptance criteria" here are engineering and operational parameters demonstrating equivalence to the predicate device, not clinical efficacy or safety endpoints for the expanded spot sizes. The document explicitly states "None" for clinical tests submitted.

    The Study that Proves the Device Meets the Acceptance Criteria:

    The document describes non-clinical tests as the study proving the device meets its performance requirements.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified. As this was non-clinical engineering and software testing, the "sample size" would refer to the number of devices or test cycles, which is not detailed.
    • Data Provenance: Not specified, but it would be internal testing conducted by Cynosure Incorporated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this was non-clinical engineering and software testing. Ground truth was established by engineering specifications and direct measurement/validation against those specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable as this was non-clinical engineering and software testing. Adjudication methods are typically for human interpretation in clinical studies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is a laser system, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is a medical laser system, not an algorithm. The "standalone" performance here refers to the laser's physical and software operation, which was tested non-clinically.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the non-clinical tests, the ground truth was based on engineering specifications, direct physical measurements, and software requirement validations.

    8. The sample size for the training set:

    • Not applicable. This device is a laser system, not an algorithm that requires a "training set" in the context of machine learning.

    9. How the ground truth for the training set was established:

    • Not applicable for the reason above.