    Intended Use

    The Advance® Enforcer™ 35 Focal Force PTA Balloon Catheter is intended for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries, including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral, as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the cerebral or coronary vasculature.

    Device Description

    The Advance® Enforcer™ 35 Focal Force PTA Balloon Catheters are over-the-wire catheters that will be available with inflated balloon diameters of 6, 8, 10, 12 millimeters and a balloon length of 4 centimeters. The balloon includes four polymer elements, which provide focal force upon inflation. These elements will aid in opening lesions. The catheters are 5.2 French, dependent upon device specification, and will be available in lengths of 50, 80, or 135 centimeters. The catheters are compatible with a 0.035 inch (0.89 millimeter) diameter wire guide. The catheters will be supplied sterile and are intended for one-time use.

    AI/ML Overview

    The provided document describes the mechanical and performance testing of the Advance® Enforcer™ 35 Focal Force PTA Balloon Catheter to demonstrate its substantial equivalence to a predicate device. This is a medical device, and the evaluation is based on engineering and material science tests, not artificial intelligence. Therefore, many of the requested fields related to AI/expert ground truth/etc. are not applicable.

    Here's the information extracted from the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriterionReported Device Performance
    Balloon Compliance TestingLabeled diameter at nominal pressure under simulated body temperature.Each balloon met its labeled diameter at the nominal pressure. The acceptance criterion was met.
    Balloon Profile TestingDiameters less than maximum outside diameter for intended sheath size.Diameters for each catheter were less than the maximum outside diameter appropriate for the intended sheath size. The acceptance criterion was met.
    Fatigue TestingFrom leakage and damage on inflation after 10 cycles of inflation/deflation (ISO 10555-4).Balloons were free from leakage and damage on inflation, withstanding 10 cycles of inflation/deflation. Conformed to ISO 10555-4. The acceptance criterion was met.
    Balloon Burst TestingBurst at or above minimum rated burst pressure, with all failure modes being linear tears.Balloons burst at or above the minimum rated burst pressure, with all failure modes being linear tears. The acceptance criteria were met.
    Balloon Inflation/Deflation TestingInflate to rated burst pressure within 60 seconds and fully deflate within 60 seconds.Balloons inflated to rated burst pressure within 60 seconds and fully deflated within 60 seconds. The acceptance criteria were met.
    Sheath Compatibility TestingCapable of being inserted and retracted from an appropriately sized sheath without excessive resistance.Catheters were capable of being inserted and retracted from an appropriately sized sheath without experiencing excessive resistance. The acceptance criterion was met.
    Tensile Strength TestingPeak load values in accordance with applicable values of ISO 10555-1.Under proper clinical use of the device, the peak load values were in accordance with the applicable values of ISO 10555-1. The acceptance criteria were met.
    Soft Tip Integrity TestingSoft tip did not separate, kink, accordion, deform, or show other anomalies after clinically relevant model passage.Soft tip did not separate from the catheter, kink, accordion, deform, or show any other anomalies after passage through a clinically relevant model. The acceptance criterion was met.
    Balloon Working Length TestingMatched labeled length within expected tolerance.Balloon working length for the catheters was matched the labeled length within the expected tolerance. The acceptance criterion was met.
    Simulated Use TestingAdequate or better performance for preparation, introduction, pushability, trackability, inflatability, deflatability, and interaction with supporting devices.Devices were adequate or better in terms of preparation, introduction, pushability, trackability, inflatability, deflatability, and interaction with supporting devices.
    Torque StrengthWithstood at least two rotations before failure.Balloon catheters withstood at least two rotations before failure. The acceptance criterion was met.
    Dimensional Verification TestingCatheter inner diameter, catheter length, and catheter profile are all within acceptable tolerances.Catheter inner diameter, catheter length, and catheter profile were all within acceptable tolerances. The acceptance criteria were met.
    Element Effectiveness TestingApplied at least 20% greater stress in a simulated model than a standard PTA balloon (the predicate PTA5) at a given pressure.Balloons applied at least 20% greater stress in a simulated model than a standard PTA balloon (the predicate PTA5) at a given pressure. The acceptance criterion was met.
    Animal TestingAdequate or better performance for preparation, introduction, pushability, trackability, flexibility, radiopacity, inflatability, deflatability, interaction with supporting devices, and inspection after use. No significant differences in arterial impact relative to the predicate device.Devices were adequate or better in terms of preparation, introduction, pushability, trackability, flexibility, radiopacity, inflatability, deflatability, interaction with supporting devices, and inspection after use. No significant differences in arterial impact relative to the predicate device.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify exact sample sizes for each test beyond indicating "each balloon" or "balloons" and "catheters". The data provenance is not explicitly stated in terms of country of origin or retrospective/prospective. These are engineering performance tests, typically performed in a controlled laboratory environment.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a medical device based on physical and mechanical properties, not an AI or diagnostic device requiring expert interpretation for ground truth.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as described above.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an AI algorithm.


    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the engineering tests, the "ground truth" is defined by the physical specifications, industry standards (e.g., ISO 10555-1, ISO 10555-4), and internal design requirements of the device. For animal testing, it involves direct observation of performance and arterial impact.


    8. The sample size for the training set

    Not applicable. This device does not use an AI training set.


    9. How the ground truth for the training set was established

    Not applicable. This device does not use an AI training set.