510(k) Data Aggregation

    K Number
    K141259

    Validate with FDA (Live)

    Date Cleared
    2014-06-10

    (26 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GRYPHON BR Anchor is intended for:
    Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair
    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
    Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
    Hip: Capsular Repair, Acetabular Labral Repair

    The GRYPHON PEEK Anchor is intended for:
    Shoulder: Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
    Foot/Ankle: Lateral Stabilization, Medial Stabilization
    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
    Elbow: Ulnar or Radial Collateral Ligament Reconstruction
    Hip: Capsular Repair, Acetabular Labral Repair

    Device Description

    The Gryphon™ Anchor with Permacord™ is a suture anchor preloaded on a disposable inserter assembly intended for fixation of suture to bone. Gryphon Anchors are available in absorbable BR and non-absorbable PEEK materials. Permacord suture is non-absorbable. The Gryphon Anchor with Permacord suture is supplied sterile and is for single use only.

    AI/ML Overview

    This looks like a 510(k) submission for a medical device, specifically a suture anchor, and not an AI/ML device. Therefore, the typical acceptance criteria and study design for AI/ML performance evaluation do not apply here.

    The document describes the Gryphon™ Anchor with Permacord™ as a suture anchor preloaded on a disposable inserter assembly, available in absorbable BR and non-absorbable PEEK materials with non-absorbable Permacord suture.

    Here's an analysis based on the provided text, focusing on the information available and noting where AI/ML specific criteria are not applicable:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since this is not an AI/ML device, the "acceptance criteria" are not related to diagnostic accuracy metrics (e.g., sensitivity, specificity). Instead, they relate to safety and performance specifications for a medical implant like a suture anchor. The document states:

    Acceptance Criteria (Implied)Reported Device Performance
    Mechanical Performance (e.g., pull-out strength, in-vitro testing)"Results of performance testing have demonstrated that the proposed devices are suitable for their intended use."
    Material Equivalence"The proposed Gryphon Anchors with Permacord suture have the same anchor materials and design as predicate Gryphon Anchors (K090124, K100012, K103712)."
    Functional Equivalence"The proposed device principal operation is the same as predicate Gryphon Anchors (K090124, K100012, K103712)."
    Suture Equivalence"The Permacord suture is the same suture as referenced in predicate Healix Advance ™ Anchor with Permacord™ (K133794)."

    The study that "proves" the device meets the acceptance criteria is described as "Verification activities were performed on the implant and / or its predicates. Testing assessments include pull out testing and in-vitro testing." This non-clinical testing demonstrated that the device is "substantially equivalent" to predicate devices.

    2. Sample Size for Test Set and Data Provenance:

    This information is not applicable as the described testing is non-clinical performance and material testing for a physical medical device, not an AI/ML diagnostic or predictive model. There isn't a "test set" in the context of patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable. Ground truth, in the context of AI/ML, refers to clinically verified labels or outcomes. For a physical device, testing involves engineering and materials science experts conducting physical tests, not clinical evaluations of patient data for ground truth.

    4. Adjudication Method for the Test Set:

    This information is not applicable for the same reasons as points 2 and 3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    This information is not applicable. MRMC studies are used to evaluate the impact of AI on human reader performance, which isn't relevant for a physical implant device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    This information is not applicable. This device is a physical implant, not an algorithm.

    7. The Type of Ground Truth Used:

    For the performance testing, the "ground truth" would be the objective measurement of mechanical properties (e.g., maximum load at failure for pull-out strength) under controlled laboratory conditions, compared against pre-defined engineering specifications or predicate device performance. It is not expert consensus, pathology, or outcomes data in the AI/ML sense.

    8. Sample Size for the Training Set:

    This information is not applicable as there is no AI/ML algorithm being trained.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as there is no AI/ML algorithm being trained.

    In summary, the provided document describes a 510(k) submission for a physical medical device (suture anchor) and focuses on demonstrating substantial equivalence to predicate devices through non-clinical performance testing and material characterization. The concepts of acceptance criteria, test sets, ground truth, and studies as typically applied to AI/ML devices are not relevant to this submission.