Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K140955

    Validate with FDA (Live)

    Device Name
    MTA2.1 MATERIAL
    Manufacturer
    Date Cleared
    2014-06-11

    (58 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MTA2.1 MATERIAL is indicated for dental procedures that contact pulp and periradicular tissues, as well as obturation and sealing of root canals.

    Device Description

    The MTA2.1 MATERIAL is designed and developed for dental clinicians to use in contact with vital pulp tissue and periradicular tissue, including sealing and obturation of root canals. This material can be used for dental procedures contacting pulpal or periradicular tissue such as: Pulp capping, Cavity lining, Base material in a cavity, Pulpotomies, Root-end filling, Apexification, Perforation repair, Root resorption, and Obturation (pulpectomy) in primary or secondary teeth. The dentist will mix the powder and water-based gel and place the mixed MTA2.1 MATERIAL into the space created by the procedure. The procedure may be part of caries treatment, root canal treatment, or periapical surgery. These procedures are performed on primary and secondary dentition and are color stable over time. MTA2.1 MATERIAL is tinted yellow, pink and gray.

    AI/ML Overview

    The provided document describes the Avalon Biomed Inc.'s MTA2.1 MATERIAL and its substantial equivalence determination to a predicate device (MTA2 ROOT AND PULP MATERIAL, K122892). This submission is for a material, not an AI/ML powered medical device, therefore, many aspects of the requested information (e.g., sample size for test set, number of experts, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable.

    However, I can extract the acceptance criteria and the studies performed to demonstrate the device meets those criteria, focusing on the available non-clinical performance data.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the MTA2.1 MATERIAL are based on established dental material standards, specifically ADA 57 and ISO 9917. The document states that the MTA2.1 MATERIAL meets these standards.

    Acceptance Criterion (Standard)Reported Device Performance
    Radiopacity (ADA 57)Meets standard
    Solubility (ADA 57)Meets standard
    Dimensional Stability (ADA 57)Meets standard
    Film Thickness (ADA 57)Meets standard
    Flow (ADA 57)Meets standard
    Working Time (ADA 57)Meets standard
    Setting Time (ADA 57)Meets standard
    Compressive Strength (ISO 9917)Conforms to standard
    Leaching of Arsenic (ISO 9917)Conforms to standard
    Leaching of Lead (ISO 9917)Conforms to standard
    Washout StabilityConfirmed stability
    Discoloration StabilityConfirmed color stability
    Biocompatibility (Cytotoxicity)Evaluated, passed

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified. The document refers to "bench testing" and "design validation" for these non-clinical tests, implying laboratory experiments. The specific number of samples for each test (e.g., radiopacity, solubility) is not detailed.
    • Data Provenance: The tests were conducted as part of the device's "design validation." This implies laboratory testing performed by or for the manufacturer (Avalon Biomed Inc.). The country of origin of the data is not explicitly stated beyond the company being in Bradenton, FL, USA, and the submission being to the US FDA. The studies are prospective in the sense that they were conducted specifically for this regulatory submission to evaluate the material's properties against predefined standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. The "ground truth" for material properties like radiopacity, solubility, or compressive strength is established through standardized laboratory measurement methods as defined by ADA 57 and ISO 9917, rather than expert consensus.

    4. Adjudication Method for the Test Set

    Not applicable for material property testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically relevant for interpretative diagnostic devices where human readers evaluate cases with and without AI assistance. This submission is for a dental material, not an AI system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, this is not applicable as the device is a dental material, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for the material's performance relies on standardized measurement methods specified by ADA 57 and ISO 9917. These standards define the procedures and expected ranges for various physical and chemical properties of dental root canal filling and pulp materials. For example:

    • Radiopacity: Measured according to the method described in ADA 57.
    • Solubility, Dimensional Stability, Film Thickness, Flow, Working Time, Setting Time: Measured according to the methods described in ADA 57.
    • Compressive Strength, Leaching of Arsenic and Lead: Measured according to the methods described in ISO 9917.
    • Biocompatibility: Evaluated through cytotoxicity tests.

    8. The Sample Size for the Training Set

    Not applicable. This is a dental material, not an AI/ML model that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. (See #8)

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1