Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K140944

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-08-22

    (130 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Phoenix® Atherectomy System is intended for use in atherectomy of the peripheral vasculature. It is not intended for use in coronary, carotid, iliac or renal vasculature.

    Device Description

    The AtheroMed Phoenix Atherectomy System is a sterile, single-use device designed for atherectomy of the peripheral vasculature. The Phoenix Atherectomy System has two main components: the Phoenix Catheter and the Phoenix Handle. The Phoenix Catheter is a flexible, over-the-wire (OTW), front-cutting Catheter that continuously captures and clears debulked plaque proximally through the Catheter and Handle into a collection reservoir that resides outside the patient. For use, the Phoenix Catheter is inserted into the Phoenix Handle incorporates a self-contained battery-powered motor designed to drive and rotate the cutter of the Phoenix Atherectomy Catheter at its specified rotational speed. The device is activated by an ON/OFF slider switch on the top of the Handle. An optional Wire Support Clip can also be used to clip a guidewire torque device in a fixed position relative to the Handle. The Catheter, Handle, and Wire Support Clip are each packaged separately as sterile, single-use components of the Phoenix Atherectomy System. There are three different models of the Phoenix Catheter. Two Phoenix Catheter models track directly over the guidewire with no tip deflection capability. These models are available in 1.8mm and 2.2mm tip diameter sizes. A third, 2.4mm tip diameter model has a Catheter tip design that can be deflected and rotated by the user so that the cutter can eccentrically debulk to a larger diameter than the Catheter's 2.4mm distal cutter. The controls for deflection and rotation are housed in the Phoenix Handle when the Catheter is inserted into the Handle. All three Phoenix Catheter models are compatible with commercially available 0.014" exchange length (260cm or greater) guidewires, and all use the same Phoenix Handle. This 510(k) includes modifications to the cutter and distal assembly of the 1.8mm and 2.2mm tip diameter Phoenix Catheter models, as well as modifications to the Catheter and proximal chassis design, to further optimize the device design and manufacturability of the device. Table 9-1 summarizes the subject modifications relative to the predicate device. Additionally, the Phoenix Guidewire has been added as a compatible guidewire in the product labeling.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes the acceptance criteria implicitly through the various tests conducted and the conclusion that the device "meet defined design requirements" and "performs in a manner equivalent to the predicate Phoenix Atherectomy System." Since this is a 510(k) submission for modifications to an already cleared device, the primary acceptance criterion is substantial equivalence to the predicate device.

    Acceptance CriterionReported Device Performance (Summary)
    Overall Equivalence to Predicate DeviceModified Phoenix Atherectomy System performs in a manner equivalent to the predicate Phoenix Atherectomy System with identical intended use.
    Dimensional and Visual InspectionMet defined design requirements.
    Simulated Use PerformanceMet defined design requirements.
    Comparative Predicate Testing in Simulated LesionMet defined design requirements.
    Cutter Torque Chain Torque-to-Failure TestMet defined design requirements.
    Functional Outer Shaft Torque TestMet defined design requirements.
    Knob to Shaft TestingMet defined design requirements.
    Catheter Drive Train Stress TestMet defined design requirements.
    Cutter Stall TestMet defined design requirements.
    Temperature Rise of Catheter During Simulated UseMet defined design requirements.
    Corrosion TestMet defined design requirements.
    Kink Bend Radius TestMet defined design requirements.
    Guidewire CompatibilityMet defined design requirements.
    Sheath CompatibilityMet defined design requirements.
    Sheath Flow RateMet defined design requirements.
    Catheter Trackability in Below-the-Knee AnatomyMet defined design requirements.
    Shelf LifeMet defined design requirements.
    Biocompatibility (Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Hemolysis, Prothrombin Time Assay, Partial Thromboplastin Time Assay, In Vitro Platelet Aggregation Assay, Complement Activation, Pyrogenicity)Met defined design requirements.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each individual test. It mentions "Comparative Predicate Testing in Simulated Lesion," implying a test set was used, but the size or characteristics of the "simulated lesion" are not detailed.

    The data provenance is from laboratory testing and simulations ("Simulated Use," "Comparative Predicate Testing in Simulated Lesion," "Temperature Rise of Catheter During Simulated Use," "Catheter Trackability in Below-the-Knee Anatomy"). No information is provided about patient data, animal studies, or specific geographic origins. Given the tests listed, the data is likely prospective in nature, as tests were performed specifically to evaluate the modified device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The tests described are primarily engineering and bench-top performance evaluations, rather than clinical studies requiring expert interpretation of patient data.

    4. Adjudication Method for the Test Set

    This information is not applicable/provided. The evaluations are based on objective engineering measurements and laboratory results, not on subjective interpretations from multiple readers requiring an adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or reported for this device. This document describes a 510(k) submission for an atherectomy system, which is a surgical tool, and the testing focuses on its physical and performance characteristics, not on diagnostic accuracy or interpretation by human readers. Therefore, the concept of "improving human readers with AI vs without AI assistance" is not relevant here.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This concept is not applicable to this device. The Phoenix Atherectomy System is a physical medical device (catheter-based system), not an AI algorithm. Its performance is intrinsically tied to human operation during a medical procedure.

    7. The Type of Ground Truth Used

    The "ground truth" for the various tests appears to be defined by engineering specifications, industry standards, and the performance characteristics of the predicate device. For instance, in "Cutter Torque Chain Torque-to-Failure Test" or "Kink Bend Radius Test," the ground truth would be the pre-defined acceptable limits of torque or bend radius. For "Biocompatibility," the ground truth is established by relevant ISO standards or guidelines for material safety.

    8. The Sample Size for the Training Set

    The concept of a "training set" is not applicable to this type of device. Training sets are typically used for machine learning or AI models. This document describes the evaluation of a physical medical device.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no training set in the context of this device's evaluation, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1