510(k) Data Aggregation

    K Number
    K140816

    Validate with FDA (Live)

    Date Cleared
    2014-07-08

    (98 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    The tested chemotherapy drugs and their breakthrough detection times are as follows:

    Test Chemotherapy Drug and ConcentrationAverage BDT
    Fluorouracil, 50.0mg/ml (50,000ppm)>240 min.
    Etoposide (Toposar), 20.0mg/ml (20,000ppm)>240 min.
    Cyclophosphamide (Cytoxan), 20mg/ml (20,000ppm)>240 min.
    Carmustine (BCNU), 3.3mg/ml (3,300ppm)2.3 min.
    Thiotepa, 10.0mg/ml (10,000ppm)30.9 min.
    Paclitaxel (Taxol), 6.0mg/ml (6,000ppm)>240 min.
    Doxorubicin Hydrochloride, 2.0mg/ml (2,000ppm)>240 min.
    Dacarbazine (DTIC), 10.0mg/ml (10,000ppm)>240 min.
    Cisplatin, 1.0mg/ml (1,000ppm)>240 min.

    Please note that Carmustine and Thiotepa have extremely low permeation times of 30.9 minutes and 2.3 minutes, respectively.

    Device Description

    Powder Free Nitrile Examination Glove, Tested for Use with Chemotherapy Drugs (Blue) are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of nitrile rubber materials and are powder free. The physical and performance characteristics of the devices meet all requirements of ASTM standard D 6319-10 Standard Specification for Nitrile Examination Gloves for Medical Application.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Powder Free Nitrile Examination Glove, Tested for Use with Chemotherapy Drugs (Blue), based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicStandard/Acceptance CriteriaReported Device Performance
    Product CodeLZALZA
    Intended UseDisposable non-sterile device for medical purposes, worn on hand/finger to prevent contamination.Meets
    Chemotherapy Drugs (Breakthrough Detection Time - BDT)Fluorouracil, Etoposide, Cyclophosphamide, Paclitaxel, Doxorubicin HCl, Dacarbazine, Cisplatin: >240 minFluorouracil (50.0mg/ml): >240 min Etoposide (20.0mg/ml): >240 min Cyclophosphamide (20mg/ml): >240 min Paclitaxel (6.0mg/ml): >240 min Doxorubicin HCl (2.0mg/ml): >240 min Dacarbazine (10.0mg/ml): >240 min Cisplatin (1.0mg/ml): >240 min
    Carmustine & Thiotepa: Permeation times should be reported.Carmustine (3.3mg/ml): 2.3 min Thiotepa (10.0mg/ml): 30.9 min
    LabelingNo special labeling claims, no hypoallergenic claims.Meets
    Device MaterialsNitrile compoundNitrile Compound
    ColorBlueBlue
    Tensile strength (before/after aging)ASTM D6319-10 (Meets)Meets
    Ultimate elongation (before/after aging)ASTM D6319-10 (Meets)Meets
    Freedom from pinholesASTM D6319-10, ASTM D5151-06 (2011), 21CFR800.20 (Meets)Meets
    Dimensions (Overall length, Width, Palm, Finger thickness)ASTM D6319-10 (Meets)Meets
    Residual powderASTM D6319-10, ASTM D6124 (Meets)Meets
    Primary skin irritation testISO 10993-10 (Not an irritant)Not an irritant
    Dermal sensitization assayISO 10993-10 (Not a sensitizer)Not a sensitizer
    Resistance to PermeationASTM D6978-05 (Specific breakthrough times for chemotherapy drugs)Carmustine: 2.3 minutes Thiotepa: 30.9 minutes Other specified drugs: >240 minutes

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for the individual tests (e.g., tensile strength, pinholes, biocompatibility, resistance to permeation). However, it indicates that the device's physical and performance characteristics meet the requirements of ASTM D 6319-10 and ASTM D 6978-05, which are standards that specify testing methodologies and, implicitly, sample sizes (e.g., AQL levels for pinholes).

    The data provenance is from non-clinical tests performed to demonstrate substantial equivalence to a predicate device. The country of origin for the device manufacturer is CHINA (Hebei Province). The testing itself (e.g., ASTM standards, ISO standards) is generally applicable internationally. The document does not specify if the data is retrospective or prospective, but given it's part of a 510(k) submission, it would be considered prospective in the sense that the device was specifically tested to demonstrate compliance for this submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable to this type of device and study. The "ground truth" for glove performance is established by standardized testing protocols (e.g., ASTM, ISO) that yield quantitative measurements (e.g., tensile strength values, breakthrough detection times, pass/fail for pinholes, irritation/sensitization ratings). There isn't a need for expert consensus in the way there would be for image interpretation or disease diagnosis.

    4. Adjudication Method for the Test Set

    This is not applicable. As mentioned above, the acceptance criteria are based on objective measurements against established standards, not on subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not conducted. This type of study is relevant for diagnostic devices where human readers (e.g., radiologists) interpret results, and the study aims to assess the impact of an AI tool on their performance. For a medical glove, performance is evaluated through physical, chemical, and biological testing as outlined above.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This is not applicable. The device is a physical product (a glove), not an algorithm or AI system. Its performance is inherent to the product itself, not an independent algorithmic output.

    7. The Type of Ground Truth Used

    The ground truth used for this device is based on:

    • Standardized Test Results: Compliance with established industry standards such as ASTM D6319-10 (for physical properties like tensile strength, elongation, dimensions, freedom from holes, residual powder) and ISO 10993-10 (for biocompatibility).
    • Quantitative Measurement: For chemotherapy drug permeation, the "ground truth" is the measured breakthrough detection time (BDT), determined according to ASTM D6978-05.

    8. The Sample Size for the Training Set

    This is not applicable. The device is a physical product, not a machine learning model. Therefore, there is no "training set" in the context of AI/ML. The manufacturing process of the glove and its materials are designed to meet the specifications, and the testing ensures these specifications are met for production batches.

    9. How the Ground Truth for the Training Set was Established

    This is not applicable for the same reasons as #8.