Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K140667

    Validate with FDA (Live)

    Date Cleared
    2014-12-15

    (272 days)

    Product Code
    Regulation Number
    876.5130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Well Lead Hydrophilic Silicone Foley Catheter is intended for use for bladder management including urine drainage, collection and measurement. The Catheter is passed through the urethra during urinary catheterization and into the bladder to drain urine. The Three-way Catheter provides a lumen that is used for bladder irrigation.

    Device Description

    The Well Lead Hydrophilic Silicone Foley Catheter is sterile, single patient use, urinary drainage catheter that is made from silicone and with a hydrophilic coating. The Catheter is available in two types, Two-way and Three-way Silicone Foley Catheter, it is supplied in French size ranging from 6 to 26 and balloon size 1.5cc to 30cc. The 6Fr, 8Fr and 10Fr catheters are for pediatric, and others are for adult.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the "Well Lead Hydrophilic Silicone Foley Catheter." It details the device, its intended use, and substantial equivalence to predicate devices. Crucially, it outlines the performance requirements and testing conducted to demonstrate compliance.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Flow Rate Through Drainage LumenMet performance requirements per testing conducted according to ASTM F623-99 and/or Well Lead testing/acceptance criteria.
    Balloon Integrity (Resistance to Rupture)Met performance requirements per testing conducted according to ASTM F623-99 and/or Well Lead testing/acceptance criteria.
    Inflated Balloon Response to PulloutMet performance requirements per testing conducted according to ASTM F623-99 and/or Well Lead testing/acceptance criteria.
    Balloon Volume MaintenanceMet performance requirements per testing conducted according to ASTM F623-99 and/or Well Lead testing/acceptance criteria.
    Balloon Size and Shaft SizeMet performance requirements per testing conducted according to ASTM F623-99 and/or Well Lead testing/acceptance criteria.
    Deflation Reliability (Failure to Deflate)Met performance requirements per testing conducted according to ASTM F623-99 and/or Well Lead testing/acceptance criteria.
    Coefficients of FrictionMet performance requirements per testing conducted according to ASTM F623-99 and/or Well Lead testing/acceptance criteria.
    Coating effect on base material after agingApplication of the coating has no adverse effect on the base material of the shaft or balloon.
    BiocompatibilityPassed biocompatibility testing per ISO 10993-1 (Cytotoxicity, Implantation, Irritation, Sensitization, Systemic Toxicity).

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each of the performance tests. It states that "Testing datas and results are included in this submission," but doesn't elaborate on the number of catheters tested for each criterion.

    The data provenance is from Well Lead Medical Co. LTD (the manufacturer) and is part of their 510(k) submission to the FDA. It is prospective testing conducted by the manufacturer specifically to demonstrate compliance for their device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this type of device and testing. The acceptance criteria are based on objective physical and material properties, defined by industry standards (ASTM F623-99, ISO 10993-1) and the manufacturer's internal criteria, not on expert interpretations of medical images or conditions.

    4. Adjudication method for the test set

    This information is not applicable. The tests performed are objective, measurable physical properties, not subjective assessments requiring adjudication by multiple experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. This device is a urological catheter, not an AI-powered diagnostic tool requiring human interpretation. Therefore, an MRMC study is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. This is a physical medical device, not an algorithm, so "standalone performance" in the context of AI is not relevant. The performance refers to the physical and material properties of the catheter itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is established by objective physical measurements and adherence to recognized industry standards.

    • For performance criteria like "Flow Rate Through Drainage Lumen" or "Balloon Volume Maintenance," the ground truth is defined by the numerical specifications and limits set by ASTM F623-99 and/or Well Lead's internal acceptance criteria.
    • For biocompatibility, the ground truth is determined by the results of specific assays as mandated by ISO 10993-1 (e.g., cell viability for cytotoxicity, inflammatory response for implantation, etc.).

    8. The sample size for the training set

    This information is not applicable. This is a physical medical device, not a machine learning model, so there is no concept of a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1