    UFill Bulk Fill Posterior Flowable Composite is indicated for use (i) in fillings with layer thicknesses up to 4mm in Class I and II cavities, (ii) in cavity lining- as a first (bottom layer) in Class I and II cavities.

    Device Description

    UFill Bulk Fill Posterior Flowable Composite is a light cure radiopaque posterior composite for use in the Bulk-Fill Technique. The flowable viscous material is suitable for filling layers of up to 4mm in thickness and as a cavity lining (bottom layer) for both Class I and II restorations. Placed in a unit dose needle capsule containing 0.25g of material, the device is delivered by a standard dental unit-dose tip applicator dispenser gun, and offered in a universal dentin shade that is formulated and manufactured without Bis-GMA, HEMA, or TEG-DMA.

    AI/ML Overview

    The provided document describes the "UFill Bulk Fill Posterior Flowable Composite" and its substantial equivalence to a predicate device. However, it does not contain information about acceptance criteria or a study that uses human experts to establish ground truth for a test set, nor does it discuss AI or machine learning. The device is a dental restorative material, and the performance testing mentioned is physical testing per ISO 4049-2009.

    Therefore, I cannot provide the requested information. The document focuses on demonstrating substantial equivalence based on material composition, intended use, and physical properties, not on diagnostic performance or human-AI interaction.

    The information present in the document is limited to:

    • Device Description: A light-cure radiopaque posterior composite for bulk-fill technique.
    • Predicate Device: 3M-Filtek Bulk Fill Flowable Restorative (K120453).
    • Indications for Use: Fillings with layer thicknesses up to 4mm in Class I and II cavities, and as a cavity lining (bottom layer) in Class I and II cavities.
    • Technological Characteristics: UDMA and BDDMA as resin components, silanized strontium aluminum boron silicate glass and silanized silica fillers (approx. 77% by weight), light-activated cure.
    • Performance Testing: "Physical testing per ISO 4049-2009, Dentistry -- Polymer-based restorative materials was conducted with all criteria for success satisfied."
    • Safety Testing: Evaluated per AAMI/ANSI/ISO 10993-1:2009.
    • Risk Management: Evaluated per ISO 14971:2007.