510(k) Data Aggregation

    K Number
    K140336

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-06-25

    (135 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OPTIMIS FUSION YAG: photodisruption of ocular tissue using light energy emitted by a Nd: YAG Laser, including discission of posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy.
    OPTIMIS FUSION SLT: Selective Laser Trabeculoplasty
    OPTIMIS FUSION YAG/SLT: photodisruption of ocular tissue using light energy emitted by a Nd: YAG Laser, including discission of posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy; and Selective Laser Trabeculoplasty

    Device Description

    The QUANTEL MEDICAL OPTIMIS FUSION is an ophthalmic surgical laser designed for performing photodisruption of ocular tissue using laser energy emitted by a Nd:YAG laser including discission of the posterior capsule of the eye (posterior capsulotomy), discission of the pupillary membranes (pupillary membranectomy), and iridotomy/iridectomy; and selective laser Trabeculoplasty.

    The Family of OPTIMIS FUSION Ophthalmic Lasers. Delivery Device and Accessories consists of the following models:

    1. OPTIMIS FUSION YAG a Nd: YAG Laser providing-switched laser pulses at a wavelength of 1064 nm for use in photodisruption of ocular tissue (posterior capsulotomy, pupillary membranectomy, iridotomy/iridectomy). The 1064 treatment beam delivers 4 ns, 0.3 -10mJ adjustable and selectable single, double or triple pulse of energy. It is conditioned trough beam shaping optics to generate a photodisruption micro pulse of plasma at a precision adjustable location relative to the visual focal plane (located at slit lamp center of rotation) and along the slit lamp objective lens axis. A twin aiming beam is also focused by the slit lamp objective to a converging 10 um spot located at the focal point of the lens. The focal point of photodisruption is adjustable 150um in the posterior direction and -150um in the anterior direction by the physician relative to this convergence of the twin aiming beams.
    2. OPTIMIS FUSION SLT a Nd: YAG Laser providing Q-switched frequency doubled pulses at a wavelength of 532 nm for use in Selective Laser Trabeculoplasty. The treatment beams delivers a 4 nsec. 0.3-2mJ adjustable single pulse of energy. The aiming and treatment beams are coaxial with each other and focussed by the slit lamp objective to a 400um spot at the focal point of the lens.
    3. OPTIMIS FUSION YAG/SLT a Nd: YAG Laser providing Q-switched laser pulses at a wavelength of 1064 nanometers for use in photodisruption or O-switched frequency doubled pulses at a wavelength of 532 nm for use in Selective Laser Trabeculoplasty. depending upon the mode selected. The OPTIMIS FUSION YAG/SLT contains two aiming beam modules that produce a single beam for the 532nm mode and a dual beam for the 1064nm mode, respectively.

    For each OPTIMIS FUSION model, the physician controls delivery of laser energy from the OPTIMIS FUSION control display unit and activates the treatment laser beam with a footswitch or joystick pushbutton. In addition, a laser slit lamp adaptor may be coupled to each of the above OPTIMIS FUSION models and connected to a currently cleared QUANTEL MEDICAL 532nm photocoagulator (VITRA (K04236), SUPRA (K07776), VITRA MULTISPOT (K122251)) to allow to the physician to use the Slit Lamp Adaptor to deliver 532 nm continuous wave laser energy for retinal photocoagulation.

    The OPTIMIS FUSION produces short, individual pulses of focused laser light with wavelengths of either 1064 nm or 532 nm. depending on the selected operational mode. Using a slit lamp microscope and aiming beam, the pulsed light is accurately targeted on a structure within the patient's eye.

    When the photodisruptor mode is selected, the treatment wavelength is 1064nm. A twinaiming beam targets the area of tissue disruption. The energy contained within a single short pulse is concentrated by focusing to a very small spot size so that plasma formation occurs at the focal point. This creates an acoustic wave which disrupts nearby tissue.

    When the SLT mode is selected, the treatment wavelength is 532nm. A coaxial aiming beam targets the trabecular meshwork via a contact lens. The SLT treatment laser provides a low energy, short pulse of laser light that produces a thermal effect in pigmented cells in the trabecular meshwork.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Quantel Medical OPTIMIS FUSION Ophthalmic Laser Systems:

    Disclaimer: This 510(k) summary focuses on demonstrating substantial equivalence to existing predicate devices, rather than proving novel clinical efficacy or accuracy against a defined ground truth in a clinical study. Therefore, many of the typical elements sought for evaluating AI/diagnostic device performance will not be explicitly present.


    Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not delineate specific, quantifiable "acceptance criteria" in the format of a diagnostic performance study (e.g., sensitivity, specificity, AUC thresholds). Instead, the acceptance criteria are implicitly tied to demonstrating substantial equivalence to predicate devices. This means the device is deemed acceptable if its technological characteristics and intended use are similar enough to previously cleared devices that it raises no new questions of safety or effectiveness.

    The "device performance" reported is also in terms of technological characteristics and similarity to predicate devices, rather than clinical outcome or diagnostic accuracy metrics.

    Feature / CriterionOPTIMIS FUSION Performance (vs. Predicate)
    Indications for UseIdentical to Lumenis Family of Selecta Ophthalmic Laser Systems. YAG model indications are similar to Carl Zeiss Meditec VISULAS YAG III (Optimis Fusion YAG also indicates pupillary membranectomy). SLT model indications are identical to Quantel Medical Solutis.
    Technological Characteristics (General)Same basic design: Ophthalmic surgical laser for photodisruption (YAG) and Selective Laser Trabeculoplasty (SLT). Emits Nd:YAG laser energy. Multiple configurations (SLT, YAG, YAG/SLT combined).
    SLT Model (OPTIMIS FUSION SLT)- Laser Type: Q-Switched frequency doubled Nd: YAG laser - Wavelength: 532 nm (Same as Selecta SLT, Solutis) - Pulse Mode: Single pulse (Same as Selecta SLT, Solutis) - Energy Range: 0.3-2.0 mJ (Same as Selecta SLT; Solutis is 0.2-2.0 mJ - considered minor difference) - Spot Size in Air: 400 microns (Same as Selecta SLT, Solutis) - Focus Angle: < 3 degrees (Same as Selecta SLT, Solutis) - Max Energy Density at Min Exposure: 1.6 J/cm² (Same as Selecta SLT, Solutis) - Repetition Rate: 2.0 Hz (Same as Selecta SLT, Solutis) - Pulse Duration: 4 ns (Same as Solutis; Selecta SLT is 3 ns - considered minor difference) - Max Average Power: 4 mW (Same as Selecta SLT, Solutis)
    YAG Model (OPTIMIS FUSION YAG)- Laser Type: Q-Switched Nd: YAG laser - Wavelength: 1064 nm (Same as Selecta 1064, VISULAS YAG III) - Pulse Mode: Pulsed with single spot, user selectable bursts of 1, 2, or 3 pulses (Same as Selecta 1064, VISULAS YAG III) - Energy Range: 0.3-10.0 mJ (Same as Selecta 1064, VISULAS YAG III) - Spot Size in Air: 10 microns (Same as Selecta 1064, VISULAS YAG III) - Focus Angle: 16 degrees (Same as Selecta 1064, VISULAS YAG III)
    YAG/SLT Model (OPTIMIS FUSION YAG/SLT)Combines features/specs of both OPTIMIS FUSION SLT and OPTIMIS FUSION YAG systems, with unchanged individual specs, similar to Lumenis Selecta Duet.
    Aiming System635 nm red diode (Same as predicate systems).
    Delivery SystemSlit lamp (Same as predicate systems). Uses CSO SL980 slit lamp (cleared in K992836).
    Validation/Verification (General Statement)Laboratory testing was conducted to validate and verify that the Family of OPTIMIS FUSION Ophthalmic Laser Systems, Delivery Device and Accessories met all design specifications and was substantially equivalent to the predicate devices.

    Study Details (Based on the Provided 510(k) Summary)

    The provided document describes a 510(k) Premarket Notification submission, which is primarily focused on demonstrating "substantial equivalence" to legally marketed predicate devices, not on a de novo clinical study to prove efficacy or accuracy. As such, the study details you've requested regarding AI/diagnostic device performance are largely not applicable (N/A) or not explicitly detailed in this type of submission.

    Here's what can be inferred or stated as N/A:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
      • N/A. This submission relies on "laboratory testing" to verify design specifications and demonstrate substantial equivalence comparing technological characteristics. It does not describe a clinical "test set" of patient data for evaluating diagnostic performance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
      • N/A. As there is no clinical "test set" with a diagnostic ground truth, expert consensus for ground truth establishment is not relevant to this submission.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      • N/A. No clinical test set.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      • No. This is a medical device (laser system) without an AI component for interpretation or diagnosis. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
      • N/A. This is a physical laser system, not an algorithm, so "standalone algorithm performance" is not relevant. The device operates under direct physician control ("human-in-the-loop" in the sense that a human physically operates it).
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc):
      • N/A. The "ground truth" for this type of submission is the performance and safety established for the predicate devices. The new device demonstrates "ground truth" of its own performance through "laboratory testing" measured against its own design specifications, and the comparison of these specifications to the predicates.
    7. The sample size for the training set:
      • N/A. This device does not use machine learning, therefore there is no "training set" in the AI sense.
    8. How the ground truth for the training set was established:
      • N/A. No training set for an AI algorithm.

    Summary of the Study Conducted (Based on the 510(k) Text):

    The "study" conducted for the Quantel Medical OPTIMIS FUSION Ophthalmic Laser Systems, as described in the 510(k) summary, consists of:

    • Laboratory Testing: This was performed to validate and verify that the device met its own design specifications (e.g., wavelength, energy range, spot size, pulse duration, repetition rate, maximum average power).
    • Substantial Equivalence Analysis: A comparative analysis of these design specifications and the Indications for Use against legally marketed predicate devices (Lumenis Selecta Family, Carl Zeiss Meditec VISULAS YAG III, Quantel Medical Solutis). The primary goal was to demonstrate that the new device's technological characteristics and intended use are sufficiently similar to the predicates, and that any differences do not raise new questions of safety or effectiveness. The conclusion was that differences were "extremely minor in significance and don't affect the safety or efficacy of the devices."