Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K140237

    Validate with FDA (Live)

    Date Cleared
    2015-01-08

    (343 days)

    Product Code
    Regulation Number
    868.5630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The disposable nebulizer set is used to administer various aerosol treatments to adult patients in hospital environments. It is not intended for transport use. The Disposable Nebulizer Set is for less than 24 hours cumulative use.

    The disposal nebulizer set is used to administer various aerosol treatments to adult patients in hospital environments. It is not intended for transport use. It is intended for adult patients consistent with the indications for the aerosol medication The Disposable Nebulizer Set is for single use for less than 24 hours cumulative use

    Device Description

    The disposal nebulizer set is designed and manufactured by SUZHOU WEIKANG MEDICAL APPARATUS CO., LTD. The disposal nebulizer sets are made of PVC; they are supplied as sterile. It is for single use for less than 24 hours cumulative use. The nebulizer set consists of a hollow tube connected to a nebulizer jar/bottle, which holds the liquid medication; a compressed gas source respiratory size aerosolized liquids into gasses that are delivered directly to the patient for breathing.

    The Weikang Disposable Nebulizer Set is a light weight portable aerosol Nebulizer which uses a compressor that causes compressed air to flow at high velocity through a liquid medicine to turn it into an aerosol which is inhaled by the patient.

    AI/ML Overview

    This FDA 510(k) summary describes the Suzhou Weikang Medical Apparatus Co., Ltd Disposable Nebulizer Set (K140237). It is a premarket notification for a medical device, which means the device is being compared to existing legally marketed predicate devices to show "substantial equivalence."

    Based on the provided document, here's a breakdown of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document establishes substantial equivalence by comparing the proposed device's performance attributes to two predicate devices: Headstar Medical Small Volume Jet Nebulizer (K121770) and Miller (VixOne) (K926055). The acceptance criteria are implicit in the performance characteristics of the predicate devices. The proposed device aims to meet or be comparable to these established performance levels.

    AttributePredicate Device (Miller - K926055)Proposed Device (Suzhou Weikang - K140237)Acceptance Criteria (Implied)Does it meet?
    Intended UseA handheld, pneumatic nebulizer designed to aerosolize prescription drugs for inhalation by a patient. Its use is indicated whenever a healthcare professional administers or prescribes medical aerosol products to a patient using a small volume nebulizer.The disposal nebulizer set is used to administer various aerosol treatments to adult patients in hospital environments. It is not intended for transport use.Similar intended use for administering aerosol treatments.Yes (for adult patients in hospital settings)
    MechanismPneumatic (gas powered) jet nebulizerAir compressorSimilar mechanism for aerosol generation.Yes (both use compressed air)
    Raw materialsUnknownNebulizer Bottle: PP, Mouth Piece: PP, Oxygen Tubing: PVC, Corrugated Tubing: PVC, CPA: PPBiocompatible materials, suitable for medical use.Yes (material specifications provided)
    SterilizationEOEOSterilization method is equivalent.Yes
    PopulationAdult and pediatric patientsAsthma Patient, adult patientSubset of predicate population (adult patients).Yes (meets for adult patients)
    Anatomic siteMouthMouthSame route of administration.Yes
    Environment of UseHome or hospitalHospitalSubset of predicate environment.Yes (meets hospital environment)
    Flow rate6-8 l/min4-8 l/minOverlap in flow rate range.Yes
    Air pressure (Kpa)Unknown700Comparable if within acceptable therapeutic range.Yes (value provided and presumably acceptable)
    Total dose delivered by device (µg)Albuterol sulfate: 1557±113; Ipratropium bromide: 276±5; Cromolyn sodium: 6227±945Albuterol sulfate: 1483±46; Ipratropium bromide: 284±8; Cromolyn sodium: 6387±361Values are similar and within acceptable manufacturing/therapeutic variation.Yes
    Particle size (MMAD) (µm)Albuterol sulfate: 2.1; Ipratropium bromide: 1.53±0.06; Cromolyn sodium: 1.97±0.06Albuterol sulfate: 2±0.1; Ipratropium bromide: 1.5±0.13; Cromolyn sodium: 2.2±0.33Values are similar and within acceptable therapeutic range for aerosol delivery.Yes
    Geometric standard deviationAlbuterol sulfate: 3.33±0.48; Ipratropium bromide: 3.45±1.23; Cromolyn sodium: 2.88±0.33Albuterol sulfate: 3.28±0.28; Ipratropium bromide: 3.13±0.3; Cromolyn sodium: 2.59±0.58Values are similar.Yes
    Total respirable dose (0.5-5 µm) (µg)Albuterol sulfate: 786±63; Ipratropium bromide: 146±9; Cromolyn sodium: 4001±506Albuterol sulfate: 737±22; Ipratropium bromide: 141±13; Cromolyn sodium: 4075±320Values are similar and within acceptable therapeutic range.Yes
    BiocompatibilityISO 10993-5 and ISO 10993-10ISO 10993-5 and ISO 10993-10Meets specified international standards for biocompatibility.Yes

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state a "test set" in the context of an AI/algorithm study. Instead, the "testing" involves a series of non-clinical studies (performance bench testing and biocompatibility testing) to demonstrate that the new device has "similar performance characteristics and technological characteristics" to the predicate devices.

    • Sample Size for performance testing (Total Dose, Particle Size, Respirable Dose): Not explicitly stated in terms of number of runs or units tested, but numerical results with standard deviations are provided, implying multiple measurements were taken for each parameter.
    • Data Provenance: The data appears to be prospective in the sense that it was generated by Suzhou Weikang Medical Apparatus Co., Ltd specifically for this 510(k) submission, comparing their new device's performance to the predicate devices. The country of origin for the testing itself is not explicitly stated, but the manufacturer is based in China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable to this document. The device is a medical apparatus (a nebulizer), not an AI algorithm. Therefore, there's no "ground truth" established by experts in the context of diagnostic interpretation or AI model validation. The ground truth for performance measures (e.g., total dose, particle size) is physical measurement against established laboratory standards.

    4. Adjudication method for the test set:

    This is not applicable as there is no diagnostic image/data interpretation by experts that would require an adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is a nebulizer, a physical medical apparatus, not an AI-powered diagnostic tool requiring human-in-the-loop performance studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable as the device is not an algorithm.

    7. The type of ground truth used:

    For the performance characteristics (e.g., total dose, particle size, respirable dose), the "ground truth" is established through physical measurements and adherence to recognized scientific and engineering principles for nebulizer performance testing. For biocompatibility and sterilization, the ground truth is established by compliance with international standards (ISO 10993 series and ISO 11135-1).

    8. The sample size for the training set:

    This is not applicable. There is no AI/machine learning model, and thus no "training set."

    9. How the ground truth for the training set was established:

    This is not applicable as there is no AI/machine learning model.


    Summary of the Study Proving Acceptance Criteria:

    The study proving the device meets the acceptance criteria is a series of non-clinical bench tests and compliance assessments described under "Summary of Non-Clinical studies". These studies evaluate the physical and material characteristics of the Suzhou Weikang Disposable Nebulizer Set.

    • Biocompatibility: The device was tested for biocompatibility according to ISO 10993-5 (In vitro cytotoxicity) and ISO 10993-10 (Irritation and skin sensitization), and also complies generally with ISO 10993-1.
    • Sterilization: The sterilization process (Ethylene Oxide) complies with ISO 11135-1, and residual ethylene oxide levels comply with ISO 10993-7.
    • Performance Testing: The device's performance characteristics, such as total dose delivered, particle size (MMAD), geometric standard deviation, and total respirable dose for specific medications (Albuterol sulfate, Ipratropium bromide, Cromolyn sodium), were measured and compared to the values reported for the predicate device (Miller/VixOne K926055). The results showed comparable performance, demonstrating substantial equivalence in key therapeutic delivery metrics.
    • Other characteristics: Manufacturing materials, intended use, mechanism, target population, anatomic site, environment of use, flow rate, and air pressure were also compared to the predicate devices and found to be sufficiently similar or equivalent.

    The conclusion is that based on these non-clinical studies and comparisons, the proposed device is "Substantially Equivalent" (SE) to the legally marketed predicate devices, implying it is as safe and effective. No clinical studies were required to support this claim of substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1