Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K134038

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-03-20

    (79 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Camber Spine Technologies Diagon Oblique Cage is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Camber Spine Technologies Diagon Oblique Cage is to be used with autologous bone graft and implanted via an open transforaminal or posterior approach.
    Camber Spine Technologies Diagon Oblique Cage implants are to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

    Device Description

    Camber Spine Technologies, Diagon Oblique Cage is a device for interbody fusion of the anterior column of the spine. These cages are hollow so that bone can grow through the device, fusing the adjacent bony surfaces.
    Camber Spine Technologies, Diagon Oblique Cage is a hollow device with texture on two opposing convex sides, and is offered in various lengths, widths, heights and shapes. Camber Spine designed the Camber Spine Technologies, Diagon Oblique Cage to be placed through a transforaminal or posterior approach and to address vertebrae in the lumbosacral region of the spine.
    The Camber Spine Technologies, Diagon Oblique Cage is made from Implant grade (PEEK) per ASTM2026 (Solvay Zeniva ZA-500) which is a radiolucent material with embedded tantalum x-ray markers as specified in ASTM F560.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Camber Spine Technologies Diagon Oblique Cage:

    It's important to note that the provided document is a 510(k) summary for a medical device (an intervertebral fusion cage), not an AI/software device. Therefore, many of the requested categories related to AI performance, ground truth, and expert evaluation are not applicable. I will address the applicable sections and indicate where information is not present due to the nature of the device.


    Acceptance Criteria and Device Performance Study for Camber Spine Technologies Diagon Oblique Cage

    This document describes the non-clinical testing performed to demonstrate that the Camber Spine Technologies Diagon Oblique Cage is substantially equivalent to its predicate devices. The studies focus on mechanical properties relevant to an intervertebral fusion device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Test Performed)Reported Device Performance (Met Criteria?)
    Static Compression Test per ASTM F2077Met the acceptance criteria
    Static Compression Shear Test per ASTM F2077Met the acceptance criteria
    Subsidence Test per ASTM F2267Met the acceptance criteria
    Static Expulsion TestMet the acceptance criteria

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes (N) for each of the non-clinical tests conducted (Static Compression, Static Compression Shear, Subsidence, Static Expulsion).

    Data provenance is not applicable in the context of these non-clinical, in-vitro mechanical tests. These tests are conducted in a laboratory setting on manufactured devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. The "ground truth" for mechanical performance tests of a physical device is determined by the results of the standardized tests themselves, measured against predefined engineering specifications and ASTM standards, not by human expert interpretation in the way AI ground truth is established.

    4. Adjudication Method for the Test Set

    This is not applicable to non-clinical mechanical testing. Test results are compared directly against established ASTM standards and internal specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the performance of diagnostic imaging or AI algorithms where human interpretation is involved. The device in question is a physical implant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable. The device is a physical intervertebral cage, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical tests is established by engineering specifications and standardized test methods (ASTM standards F2077, F2267). The device's performance is objectively measured against the criteria outlined in these standards and internal design requirements.

    8. The Sample Size for the Training Set

    This is not applicable. There is no concept of a "training set" for the type of non-clinical mechanical testing described for this physical medical device. Manufacturing processes are validated, but this is distinct from training an AI model.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable, as there is no training set for this type of device and testing.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1