510(k) Data Aggregation

    K Number
    K133795

    Validate with FDA (Live)

    Date Cleared
    2014-12-12

    (364 days)

    Product Code
    Regulation Number
    870.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sunmed The Inflation Device is a set of accessorial device. It is composed of an inflation device, and may be composed of a stopcock and a hemostatic valve. The inflation device is intended for use in PTCA or PTA procedures to create and monitor pressure (by connecting the rotating male luer of inflation device with the balloon catheter's female luer) in the balloon and to deflate the balloon. The hemostatic valve is recommended for use during PTCA or PTA procedures in conjunction and/or diagnostic device (e.g. balloon dilatation catheters).Stopcock is used to release and purge any trapped air (if used).

    Device Description

    The Sunmed The Inflation Device is a set of accessorial device. It is composed of an inflation device, and may be composed of a stopcock and a hemostatic valve. The inflation device is 30ml disposable device capable of producing a maximum pressure of 40ATM. Pressure can be monitored via a pressure gauge. The manually operation of the device is achieved by the manipulation of a large handle to drive a piston housed within the body of the device.

    Hemostatic valve is a single Y connector, and may be composed of an insertion tool, and a torquer. The hemostasis valve is designed to provide a port for interventional system. The insertion tool is used to facilitate placement of a guide wire tip through the hemostasis valve. The torque device is designed to hold the guide wire and provide a handle for manipulating.

    AI/ML Overview

    This FDA 510(k) summary for the Sunmed™ Inflation Device (K133795) describes a device intended for use in PTCA or PTA procedures. However, it does not contain the detailed information required to fill out the user's requested template regarding acceptance criteria and a study proving those criteria are met.

    Here's why and what information is available:

    The document is a 510(k) premarket notification, which seeks to demonstrate substantial equivalence to legally marketed predicate devices, not necessarily to prove the device meets specific acceptance criteria through a clinical study in the way a PMA (Premarket Approval) might require.

    Information NOT available in the provided text:

    • A table of specific numerical acceptance criteria (e.g., specific thresholds for pressure accuracy, deflation time, etc.) and the reported device performance against those criteria. The document mentions "non-clinical testing" but does not provide results or specific criteria.
    • Details about a specific study proving acceptance criteria:
      • Sample size used for the test set.
      • Data provenance (country of origin, retrospective/prospective).
      • Number of experts, qualifications, or adjudication method for ground truth.
      • Multi-Reader Multi-Case (MRMC) comparative effectiveness study.
      • Standalone (algorithm only) performance.
      • Type of ground truth used (expert consensus, pathology, outcomes data).
      • Sample size for the training set (if applicable to an AI/algorithm-based device, which this is not).
      • How ground truth for the training set was established.

    What can be extracted and inferred from the document regarding testing:

    1. Nature of Testing: The document states "Summary of Non-Clinical Testing" and explicitly says "Clinical Evaluation was not applicable." This indicates that the device's performance was evaluated through bench testing and biocompatibility studies, not human clinical trials against specific performance endpoints.

    2. Types of Non-Clinical Tests Performed (as listed in section 7):

      • Biocompatibility Testing:
        • Pyrogen Test
        • Endotoxin Test
        • Acute Systemic Toxicity Test (two kinds of solvent)
        • Intracutaneous Reactivity Test
        • In Vitro Cytotoxicity Test
        • Skin sensitization Test (two kinds of solvent)
        • In Vitro Hemolysis Study
        • Complement Activity Test (C3a, SC5b-9)
        • In Vivo Thrombogenicity
      • Package Penetrate Testing
      • Asepsis Testing
      • Aging Testing
      • EtO and ECH Residue Testing
      • Bench Testing
    3. Lack of Reported Performance Data: While these tests were performed, the document does not report the results of these tests or the specific acceptance criteria (e.g., "Pyrogen test result must be < X EU/mL"). It merely lists the tests as having been performed.

    4. Basis for Substantial Equivalence: The conclusion (section 9) is based on the information presented, which includes technological comparison to predicate devices (K102648 and K032840) and the non-clinical testing performed. The device is considered substantially equivalent if its non-clinical performance and technological characteristics are similar enough to predicates that it raises no new questions of safety or effectiveness.

    In summary, based only on the provided text, the requested table and detailed study information cannot be created because the document is a 510(k) summary focused on substantial equivalence rather than a detailed report of performance against specific, quantifiable acceptance criteria. The FDA's decision to clear the device implies that the submitted non-clinical data did meet the required standards for substantial equivalence, but those specific standards and results are not detailed in this summary.