Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133634

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-12-23

    (27 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Spectrum Turbo-Ject® Peripherally Inserted Central Venous Catheter (PICC) Sets and Trays are intended for short- or long-term use for venous pressure monitoring, blood sampling, administration of drugs and fluids, and for use with power injectors for delivery of contrast in CT studies. The catheter is impregnated with the antimicrobials minocycline and rifampin to help provide protection against catheter-related bloodstream infections (CRBSIs). The Spectrum Turbo-Ject® PICC is indicated for multiple injections of contrast media through a power injector. The maximum pressure limit setting for Power Injectors used with the Turbo-Ject PICC may not exceed 325 psi and the flow rate may not exceed the maximum flow rate indicated.

    Device Description

    The proposed Spectrum Turbo-Ject® PICCs are radiopaque polyurethane peripherally inserted central venous catheters for short- or long-term use, and can be inserted through a Peel-Away® introducer, or over-the-wire. The proposed devices are minimally tapered 4.0 Fr and 5.0 Fr single and double lumen catheters. The set components may include the PICC, obturator, Peel-Away® introducer, entry needles, wire guide, and other convenience components. The set is supplied sterile and is intended for one-time use.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Cook Medical Spectrum Turbo-Ject® Peripherally Inserted Central Venous Catheter Set. It outlines the device's intended use, comparison to predicate devices, and the testing conducted to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Tensile TestingPeak load value > 10 NPeak load value was greater than 10 N
    Dynamic Pressure TestingCatheters do not fail during simulated useCatheters did not fail during simulated use
    Static Failure Pressure TestingStatic failure pressure at or above the acceptance criterionStatic failure pressure was at or above the acceptance criterion
    Liquid Leakage TestingCatheters do not leak liquidCatheters did not leak liquid
    Air Leakage TestingCatheters do not exhibit air leakageCatheters did not exhibit air leakage
    Antimicrobial TestingMinimum zone of inhibition of 15 mmCatheters met the predetermined acceptance criteria

    Note: For "Static Failure Pressure Testing" and "Antimicrobial Testing", the specific numerical acceptance criteria (other than >15mm for antimicrobial, which is also stated as "predetermined") are not explicitly defined in the document, but the document states the device met them.

    2. Sample Size and Data Provenance

    The document does not explicitly state the sample sizes used for each specific test (e.g., number of catheters tested for tensile strength or leakage).
    The data provenance is retrospective, as the submission is for modifications to previously cleared devices (K081690, K072625, K132334, K132885). The testing was conducted to demonstrate substantial equivalence to these predicate devices. The country of origin of the data is not specified, but the applicant address is in Bloomington, IN, USA.

    3. Number of Experts and Qualifications

    The document does not mention the involvement of experts to establish ground truth for a test set. The tests describe physical and performance characteristics of the device, which are typically evaluated through laboratory testing against predefined engineering specifications.

    4. Adjudication Method

    Not applicable. This type of submission (510k for a medical device) typically relies on laboratory testing and engineering specifications rather than expert adjudication of a test set, which is more common in diagnostic device or AI algorithm evaluations.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was mentioned. The device is a "Peripherally Inserted Central Venous Catheter Set," a physical medical device, not a diagnostic imaging or AI-assisted system that would typically involve human readers.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. This is a physical medical device, not an algorithm or AI system.

    7. Type of Ground Truth Used

    The "ground truth" for this device's performance is based on engineering specifications and established criteria for medical device safety and effectiveness. For example:

    • Tensile strength: Measured against a force (10 N).
    • Leakage: Absence of liquid/air leakage.
    • Pressure: Ability to withstand specified pressure (325 psi).
    • Antimicrobial activity: Measured zone of inhibition (15 mm).
      These are objective, quantitative measurements against predefined standards.

    8. Sample Size for the Training Set

    Not applicable. As a physical medical device, there is no "training set" in the context of machine learning or AI. The design and manufacturing process would involve internal validation and verification activities, but these are not referred to as training sets.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this type of device. The specifications and acceptable performance limits are derived from industry standards (e.g., ISO 10555-1:1995), regulatory requirements, and the performance characteristics of predicate devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1