    Device Name
    THE DOME S6C
    Date Cleared
    2014-02-28

    (95 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dome Sx line, model Dome S6c is intended to be used in displaying and viewing medical images for review and analysis by trained medical practitioners.

    Model Dome S6c is not intended for mammography use.

    Device Description

    The Dome model S6c is a widescreen 30-inch 6 megapixel color monitor. Its sleek design, high brightness and bezel free separation between two 3MP displays makes Dome model S6c ideal for read of patient images acquired from multiple medical imaging modalities.

    AI/ML Overview

    The provided text describes NDS Surgical Imaging's Dome® S6c color monitor, a 6-megapixel display intended for displaying and viewing medical images. The submission focuses on demonstrating substantial equivalence to a predicate device, the Dome model C3 color.

    However, the document does not contain the following information regarding acceptance criteria and performance studies:

    • A table of acceptance criteria and reported device performance (beyond basic specifications and regulatory compliance). The document primarily compares the technical specifications to a predicate device for the purpose of demonstrating substantial equivalence, not against specific performance acceptance criteria for diagnostic tasks.
    • Sample size used for the test set and data provenance.
    • Number of experts used to establish ground truth and their qualifications.
    • Adjudication method for the test set.
    • Whether a multi-reader multi-case (MRMC) comparative effectiveness study was done, or any effect size of human reader improvement with AI assistance. This device is a display monitor, not an AI-powered diagnostic tool.
    • Whether a standalone performance study was done. This device is a display monitor, not an AI-powered diagnostic tool.
    • The type of ground truth used.
    • The sample size for the training set.
    • How the ground truth for the training set was established.

    The core of the submission is a comparison of technical specifications and regulatory compliance, and not a clinical performance study using patient data and expert ground truth.

    Based on the provided text, here is what can be extracted:

    The device is a medical display monitor, and the "acceptance criteria" discussed are primarily related to its technical specifications and compliance with safety and EMI standards, rather than diagnostic performance metrics. The study presented is a technical comparison for substantial equivalence.

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" in the context of a performance study with diagnostic outcomes. Instead, it compares the technical specifications and regulatory approvals of the Dome® S6c with a predicate device (K032638 Dome model C3 color). The argument for acceptance is based on demonstrating substantial equivalence in these technical and regulatory aspects.

    Characteristic ItemsPredicate device K032638 (Dome C3 Color)Reported Device Performance (Dome® S6c K133609)
    Intended UseDisplaying and viewing medical images for review and analysis by trained medical practitioners.Displaying and viewing medical images for review and analysis by trained medical practitioners. (Not for mammography use)
    Screen diagonal20.8" Diagonal (528 mm)30" Diagonal (762 mm)
    Resolution1536 x 2048 pixels (portrait) / 2048 x 1536 pixels (landscape)Dual 1536 x 2048 pixels (portrait) (future implementation) / Dual 1640 x 2048 pixels (portrait) / 3280 x 2048 pixels (landscape) (Equivalent resolution of dual 3MP radiology displays, described as 6MP)
    Pixel pitch0.207 mm0.197 mm
    Contrast ratio600:1 - typical1000:1 - typical
    Brightness235 cd/m² typical800 cd/m² typical
    Pixel rise/fall time50 ms30 ms typical
    Viewing angle (CR 10:1)+/- 85° (170°) horizontal / +/- 85° (170°) vertical+/- 85° (170°) horizontal / +/- 85° (170°) vertical
    Stabilization, calibration, automated QABuilt in RightLight rear sensor for backlight stabilization. Factory calibrated. QA through manual external photometer.Built in RightLight rear sensor for backlight stabilization. Factory calibrated. QA through embedded RightCheck front sensor and manual external photometer.
    Look up table10/12 bit10/12 bit
    Digital Video Input1x DVI Revision 1.0 single channel connector1x DVI Revision 1.0 digital dual-channel connector
    Input Formats (Landscape)2048x1536 (8 bits per pixel, 24 bit color)3280x2048 (8 or 10 bit per pixel, 24 bit or 30 bit color)
    Input Formats (Portrait)1536x2048 (8 bits per pixel, 24 bit color)2x 1536x2560 or 2x 1640x2048 (8 or 10 bit per pixel, 24 bit or 30 bit color)
    Power Supply100240 VAC 5060Hz100240 VAC 5060Hz Universal Auto Switching with Medical Approvals and PFC
    SafetyEN 60601-1 Part 1. General requirements for safetyIEC 60601-1: 2005
    EMIEN 55011 Class BEN 60601-1-2: 2007
    Primary Claim of EquivalenceThe Dome model C3 color, a 3MP display.The Dome model S6c ("6MP display device, equivalent to the resolution of dual 3MP radiology displays"). The difference is that S6c is "one package that replaces two Dome model C3 color side by side displays in use." The device meets IEC/EN60601-1 and 60601-1-2 standards for safe design and operation.

    2. Sample size used for the test set and the data provenance: Not applicable. This is a technical comparison for a display device, not a study of diagnostic performance using a test set of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a display monitor, which does not incorporate AI or directly assist human readers in diagnostic interpretation in the manner of an AI algorithm.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a display device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. The "ground truth" for this submission is adherence to technical specifications and regulatory standards.

    8. The sample size for the training set: Not applicable. This is a display device, not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established: Not applicable.