Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133456

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-09-24

    (316 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Power-TrialysisTM Short-Term Dialysis Catheter, with a third internal lumen for intravenous therapy, power injection of contrast media, and central venous pressure monitoring, is indicated for use in attaining short-term (less than 30 days) vascular access for hemodialysis, hemoperfusion, and apheresis treatments. The catheter is intended to be inserted in the jugular, femoral, or subclavian vein as required. The maximum recommended infusion rate is 5 mL/sec for power injection of contrast media.

    Device Description

    Power-Trialysis™ Short-Term Dialysis catheters are made of thermosensitive polyurethane, which softens when exposed to body temperature. The catheter is divided into three separate lumens permitting continuous blood flow. Both the venous (blue) and the arterial (red) lumens may be used for hemodialysis, hemoperfusion, and apheresis treatments. The distal (purple) lumen is completely independent from the two dialysis lumens and may be used for intravenous therapy, power injection of contrast media, and central venous pressure monitoring. The distal lumen can also be accessed for blood draws and infusion of medications.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the Power-Trialysis Short-Term Dialysis Catheter (K133456). This document focuses on demonstrating substantial equivalence to a previously cleared predicate device (K083675) rather than establishing novel performance criteria for a new type of device. As such, the structure of the provided information does not perfectly align with a typical AI/software device study seeking to establish performance against defined acceptance criteria.

    The "acceptance criteria" here are generally that the new device performs as safely and effectively as the predicate device. The study described is a series of engineering and performance tests comparing the subject device to the predicate device to confirm this substantial equivalence.

    Here's an attempt to structure the information based on your request, with clarifications where the provided text doesn't directly map to the question:


    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a substantial equivalence submission for a physical medical catheter, the "acceptance criteria" are generally based on meeting the performance established by the predicate device and relevant industry standards. The reported performance confirms these criteria were met.

    Acceptance Criteria CategorySpecific Test/Performance RequirementReported Device Performance (Summary)
    Material BiocompatibilityHemolysis testing: In accordance with ASTM F1841:1997 (R 2005)Device met acceptance criteria.
    Material/Device Integrity (Chemical Exposure)Leak, Static Burst, Tensile, Shaft to Hub Tensile testing: Following chemical conditioningDevice met acceptance criteria.
    Device Integrity (Mechanical Stress)Occluded Power Injection and Assembly Leak Post Occluded Power Injection testingDevice met acceptance criteria to address lower static burst threshold of modified device.
    Device Integrity (Repeated Use)Assembly Burst testing: Post multiple power injectionsDevice met acceptance criteria, material did not weaken.
    Functional Performance (Dialysis)Dialysis Flow testing: Satisfy established flow rate requirements under labeled procedural pressuresCatheter configurations satisfied established flow rate requirements.
    Functional Performance (Structural Integrity)Catheter Collapse for Dialysis Lumens testing: Will not collapse at established pressure requirementsCatheter lumens did not collapse at required pressures.
    Functional Performance (Power Injection)Gravity Flow testing: Determine flow rate of normal saline through power injectable lumenPerformed; results implicitly met acceptance criteria for flow.
    Functional Performance (Power Injection)Catheter Collapse for Power Injectable Lumen testing: Determine flow rate through power injectable lumen at vacuum pressurePerformed; results implicitly met acceptance criteria for flow under vacuum.
    Usability/CompatibilityGuidewire Fit testing: Ensure use compatibility between catheters and guidewiresEnsured use compatibility.
    Clinical Monitoring CapabilityCentral Venous Pressure Monitoring testing: Verify capabilities in simulated environmentCapabilities verified in simulated environment.
    Sterilization EfficacySterilization Validation: Added length does not adversely impact sterilant residual levels (ISO 10993-7: 2008)Added length did not adversely impact sterilant residual levels.
    Overall EquivalenceDemonstrates substantial equivalence to predicate device (K083675) across indications for use, technological characteristics, and safety/performance.Subject device met all predetermined acceptance criteria and demonstrated substantially equivalent performance to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes for each specific test (e.g., how many catheters were tested for hemolysis or burst strength). It refers to "verification and validation activities" and "performance data" but does not quantify the number of units tested.

    • Sample Size for Test Set: Not explicitly stated for each individual test. Implied to be sufficient for engineering validation studies following design controls (21 CFR §820.30).
    • Data Provenance: The studies were conducted by Bard Access Systems, Inc. in the US, as part of their 510(k) submission. These are prospective engineering and performance tests on manufactured devices, not retrospective clinical data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is generally not applicable to a 510(k) submission for a physical device like a catheter. The "ground truth" for the performance tests (e.g., flow rate, burst pressure, tensile strength) is based on measurable physical properties and objective engineering standards (e.g., ISO, ASTM). The "experts" involved would be the engineers, technicians, and quality control personnel conducting and reviewing these tests, whose qualifications would be in engineering, materials science, or related fields. The document does not specify the number or specific qualifications of these individuals.


    4. Adjudication Method for the Test Set

    Not applicable in the context of physical performance testing against objective standards. The results of tests like burst pressure or hemolysis are quantitative and compared directly to predetermined specifications or standards, without the need for expert adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging AI systems where human reader performance is being evaluated with and without AI assistance. The Power-Trialysis Short-Term Dialysis Catheter is a physical medical device, not a diagnostic AI tool.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not applicable. The device is a physical catheter, not an algorithm.


    7. The Type of Ground Truth Used

    The "ground truth" for the device's performance is based on:

    • Objective Engineering Standards: Adherence to established international and national standards (e.g., ASTM F1841, ISO 10555-1, ISO 10993-7).
    • Predicate Device Performance: Performance data from the legally marketed predicate device (K083675) serves as a benchmark for substantial equivalence.
    • Design Specifications: Internal design requirements and specifications for the catheter's physical and functional properties.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical device, and the concept of a "training set" is generally used for machine learning models, not for engineering validation of a catheter. The manufacturing process for these devices would involve process validation and quality control, which are distinct from machine learning training.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no "training set" in the machine learning sense, there is no ground truth established for it.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1