Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133345

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-04-11

    (163 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    REVEEL is designed as an organ and tissue retractor for the use in endoscopic surgical procedures to elevate organs and tissue to provide better access as well as visualization of surgical sites.

    Device Description

    REVEEL is a single use device intended for mobilizing and maneuvering organs and tissue during endoscopic surgical procedures. It is comprised of a proximal handle, a rigid shaft, and a distal retraction surface.

    AI/ML Overview

    The "REVEEL ENDOSCOPIC RETRACTOR" is a medical device and therefore does not involve AI or algorithms in its operation. The provided documents focus on its substantial equivalence to predicate devices based on design, materials, and non-clinical testing.

    Therefore, many of the questions related to acceptance criteria for an AI/algorithm-based device and its study are not applicable (N/A) to this product.

    Here's a breakdown of the requested information based on the provided text, indicating where the information is not relevant for this type of device:


    Description of the Acceptance Criteria and Study for the REVEEL ENDOSCOPIC RETRACTOR

    The REVEEL ENDOSCOPIC RETRACTOR is a single-use manual device intended for mobilizing and maneuvering organs and tissue during endoscopic surgical procedures. Its acceptance criteria and supporting studies are based on demonstrating substantial equivalence to legally marketed predicate devices, focusing on design, functional performance through bench and pre-clinical testing, and safety aspects like biocompatibility and sterilization.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance (Summary)
    Substantial Equivalence to PredicatesDemonstrate substantial equivalence to PRETZELFLEX (K123110), AUTO SUTURE ENDO RETRACT II (K914190), and NATHANSON (K942002) in terms of: - Sterilization method - Packaging - Product architecture - Retraction surface - Patient contact materials - Mechanism of operation - Method of insertion - Incision site and sizeThe REVEEL ENDOSCOPIC RETRACTOR is stated to be substantially equivalent to the identified predicate devices across all these characteristics.
    Bench Testing- Ensure device performs as intended - Evaluate mechanical integrity and functionalityBench testing was performed to ensure the device performs as intended. Specific performance metrics are not detailed but imply successful completion of these tests.
    Simulated Transportation & Shelf Life- Validate packaging integrity post-transport - Confirm device integrity over shelf lifeSimulated transportation and shelf life testing were conducted, implying successful validation. Specific results are not detailed.
    Sterilization Validation- Ensure device sterilitySterilization validation was performed, implying successful validation. Specific methods or results are not detailed.
    Biocompatibility Testing- Confirm patient contact materials are safe - Comply with ISO 10993 standardsBiocompatibility testing was performed, indicating compliance with safety standards for patient contact. Specific results are not detailed.
    Pre-Clinical Testing (Functional Performance)- Demonstrate functionality in real-life simulation (insertion, actuation, device mobilization and retraction, visualization of gastroesophageal junction, de-actuation and removal)Use of porcine and cadaveric models demonstrated that the device is capable of functioning as intended for all listed aspects. This confirms the device's ability to perform its intended use.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Specific sample sizes for bench tests, simulated transportation, shelf life, sterilization, and biocompatibility are not provided in the summary. For pre-clinical testing, "porcine and cadaveric models" were used, but the exact number of models is not specified.
    • Data Provenance: The studies were conducted by RETRACTION LIMITED, located in Hong Kong. The data provenance (country, retrospective/prospective) related to the specific test execution is not explicitly detailed beyond the company's location. The studies are by nature prospective as they are part of a premarket notification for a new device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is Not Applicable (N/A). The device is a manual surgical retractor, not an AI/diagnostic device that requires expert-established ground truth for performance evaluation in the context of image interpretation or similar tasks. The "ground truth" for this device's performance is its physical functionality and safety as demonstrated through engineering tests and pre-clinical simulations. The "experts" involved would be engineering and medical professionals conducting the tests, but their specific roles and qualifications are not detailed as they would be for ground-truthing diagnostic AI.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is Not Applicable (N/A). Adjudication methods like 2+1 or 3+1 are typically used in studies where a panel of experts review and reach consensus on diagnostic outcomes (e.g., in medical imaging). This is not relevant for the physical and functional testing of a surgical retractor.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is Not Applicable (N/A). MRMC studies are used for evaluating the performance of diagnostic imaging systems, often involving human readers and AI. The REVEEL ENDOSCOPIC RETRACTOR is a manual surgical tool and does not involve AI, human readers, or diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is Not Applicable (N/A). The device is a manual tool and does not employ any algorithms or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device primarily relies on:

    • Engineering specifications and standards: For bench testing, mechanical properties, and material safety.
    • Performance as "intended": In pre-clinical models (porcine and cadaveric), the "ground truth" for functionality is the observable success of the device in performing retraction, mobilization, and providing visualization, as determined by the surgical operators/evaluators participating in the simulation. This is essentially functional verification against its stated intended use.
    • Regulatory compliance: For sterilization and biocompatibility.

    8. The sample size for the training set

    This question is Not Applicable (N/A). The REVEEL ENDOSCOPIC RETRACTOR is a physical medical device, not an AI algorithm, and therefore does not have a "training set."

    9. How the ground truth for the training set was established

    This question is Not Applicable (N/A). As there is no training set for an AI algorithm, no ground truth needed to be established for it.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1