510(k) Data Aggregation

    K Number
    K133310

    Validate with FDA (Live)

    Date Cleared
    2014-01-10

    (74 days)

    Product Code
    Regulation Number
    882.5950
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Axium ™ Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae. The Axium™ Detachable Coils are also indicated for arterial and venous embolizations in the peripheral vasculature.

    Device Description

    The Axium™ Detachable Coil System consists of a platinum coil secured to a composite delivery wire and is compatible with a 2-marker band micro catheter and a mechanical detachment system. The Axium™ Detachable Coil System consists of three (3) components: 1 -- Implantable Coil, 2 - Implant Delivery Pusher, 3 - Instant Detacher. The Instant Detacher is packaged and sold separately. Axium™ coil configurations include bare platinum, PGLA enlaced platinum and Nylon enlaced platinum coils. Bare configuration coils are non-fiber-enlaced, whereas PGLA and Nylon are enlaced with PGLA and Nylon fibers respectively. This submission expands the size offerings of the Axium™ Detachable Coil System by adding one (1) new model number to the Axium bare product portfolio with a new diameter and coil length combination (SKU QC-1.5-1-Helix). Currently cleared sizes for Axium™ bare under K081465 already include a coil diameter of 1.5mm and a coil length of 1cm; however, not in combination.

    AI/ML Overview

    This document describes a 510(k) submission for the Axium™ Detachable Coil System, which is a neurovascular embolization device. The submission is for expanding the size offerings of an already cleared device.

    Here's an analysis of the provided information concerning acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The provided document describes the performance data in textual form, stating that "The following test(s) was performed to support the addition of the new size offering" and then lists several categories of bench testing. It does not explicitly present a table of acceptance criteria with corresponding reported device performance values. Instead, it states that these tests were "adopted from existing test data for currently cleared Axium™ coil sizes" and that the conclusion is based on "successful completion of non-clinical testing; a thorough assessment of existing test data; as well as identical principles of operation, materials of construction, dimensions, packaging, and indications for use."

    Therefore, the acceptance criteria are implicitly that the new size offering (SKU QC-1.5-1-Helix) performs comparably to the previously cleared Axium™ coils in the listed bench tests. The reported device performance is that it did successfully meet these implied criteria.

    Implied Acceptance Criteria and Reported Performance:

    Acceptance Criteria CategoryReported Device Performance
    Dimensions - 1st Loop Outer DiameterSuccessfully met (implied by "successful completion of non-clinical testing")
    Dimensions - LengthSuccessfully met (implied by "successful completion of non-clinical testing")
    Coil DeformationSuccessfully met (implied by "successful completion of non-clinical testing")
    FrictionSuccessfully met (implied by "successful completion of non-clinical testing")
    Fatigue After Knotting and Reliability DetachmentSuccessfully met (implied by "successful completion of non-clinical testing")
    Torque ResponseSuccessfully met (implied by "successful completion of non-clinical testing")
    Force Transfer - ImplantSuccessfully met (implied by "successful completion of non-clinical testing")
    Coil Tensile - PolypropyleneSuccessfully met (implied by "successful completion of non-clinical testing")
    Coil Tensile - Implant/Weld (Implant Coil to Coil Shell Weld)Successfully met (implied by "successful completion of non-clinical testing")
    Coil Tensile - AssemblySuccessfully met (implied by "successful completion of non-clinical testing")
    ParticulateSuccessfully met (implied by "successful completion of non-clinical testing")
    Pusher DimensionsSuccessfully met (implied by "successful completion of non-clinical testing")
    Marker RadiopacitySuccessfully met (implied by "successful completion of non-clinical testing")
    Tip BucklingSuccessfully met (implied by "successful completion of non-clinical testing")
    Detachment Zone StiffnessSuccessfully met (implied by "successful completion of non-clinical testing")
    Kink ResistanceSuccessfully met (implied by "successful completion of non-clinical testing")
    Hypotube and Weld Tensile StrengthSuccessfully met (implied by "successful completion of non-clinical testing")
    Force Transfer - PusherSuccessfully met (implied by "successful completion of non-clinical testing")
    Pusher ElongationSuccessfully met (implied by "successful completion of non-clinical testing")
    MRI CompatibilitySuccessfully met (implied by "successful completion of non-clinical testing")

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily refers to "bench testing" and "existing test data" for currently cleared devices. It does not specify the sample size for the new SKU (QC-1.5-1-Helix) or the provenance (country of origin, retrospective/prospective) of the data. Given it's a 510(k) for an additional size, it's highly likely this is internal R&D data from the company, likely prospective for the new SKU but leveraging retrospective data for comparison to the predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This submission does not involve clinical studies with human expert evaluation for establishing ground truth in the traditional sense of a diagnostic or AI device. The testing is bench-based and focuses on physical and mechanical properties. Therefore, these questions are not applicable to this type of submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is bench testing, not a clinical study requiring adjudication of expert opinions.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This submission is for a physical medical device (detachable coil system) and does not involve AI or a multi-reader, multi-case study. This question is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This submission is for a physical medical device and does not involve an algorithm or standalone performance testing in this context. This question is not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the bench testing, the "ground truth" would be the established engineering and materials specifications and performance characteristics derived from recognized standards and internal quality control protocols for the predicate devices. These are quantitative and measurable physical properties.

    8. The sample size for the training set

    This is not an AI/machine learning device; therefore, there is no training set in the conventional sense. The "training data" equivalent in this context would be the extensive manufacturing specifications, material properties, and performance data accumulated from the predicate devices' development and testing.

    9. How the ground truth for the training set was established

    As there is no training set for an AI algorithm, this question is not applicable. The "ground truth" for the overall device's performance is established through rigorous engineering design, material science characterization, and comprehensive bench testing against established industry standards and internal acceptance criteria for medical devices.