510(k) Data Aggregation

    K Number
    K133256

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-01-09

    (78 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iTotal® CR Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The indications for use include:

    • . Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis or osteonecrosis of the knee.
    • . Post traumatic loss of joint function.
    • Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
    • . Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.
    • Revision procedures provided that . anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.
      This implant is intended for cemented use only.

    The ConforMIS Unicondylar Knee Replacement System (iUni) is intended for use in one compartment of the osteoarthritic knee to replace the damaged area of the articular surface in patients with evidence of adequate healthy bone sufficient for support of the implanted components.
    Candidates for unicondylar knee replacement include those with:

    • . Joint impairment due to osteoarthritis or traumatic arthritis of the knee
    • Previous femoral condyle or tibial plateau fracture, . creating loss of function
    • . Valgus or varus deformity of the knee
    • Revision procedures provided that anatomic . landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.
      This implant is intended for cemented use only.

    The ConforMIS iDuo Bicompartmental Knee Repair System is intended for use in patients with severe knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only a single knee compartment, such as unicondylar or patellofemoral prosthesis. The indications for use include restoring joint function and relief of pain due to :

    • . Painful joint disease due to osteoarthritis
    • Traumatic arthritis of the knee .
    • Post traumatic loss of joint function .
    • . Failed osteotomies, hemiarthroplasties and unicondylar implants
      The iDuo Bicompartmental Knee Repair System may be utilized when the medial or lateral condyle and the patellofemoral areas have been affected by one or more of the above noted conditions.
      The iDuo implant is intended for cemented use only.
    Device Description

    ConforMIS knee replacement systems are patient-specific semi-constrained knee implants which consist of a femoral, tibial, and/or patellar components. The products are intended for treatment of severe pain and/or disability of the knee damaged by osteoarthritis or trauma. Using patient imaging (either CT or MR scans), a patient-specific implant is designed that best meets the geometric and anatomic requirements of the specific patient. The femoral components of the devices are manufactured from cobalt chromium molybdenum (CoCrMo) alloy. The tibial component includes a metal tray manufactured from CoCrMo alloy and polyethylene inserts manufactured from UHMWPE or an all polyethylene tibial component. The patellar components are manufactured from UHMWPE.

    AI/ML Overview

    This 510(k) submission describes the addition of Ethylene Oxide (EO) sterilization as an alternative to the existing VHP Gas Plasma sterilization for the ConforMIS iTotal CR Knee Replacement System, ConforMIS iUni Unicondylar Knee Replacement System, and ConforMIS iDuo Bicompartmental Knee Repair System. Therefore, the acceptance criteria and supporting study focus on demonstrating the safety and effectiveness of this new sterilization method.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Performance Measure)Reported Device Performance (Summary of Study Results)
    Sterility Assurance Level (SAL) of 1 x 10⁻⁶Achieved (Sterilization Validation testing to establish a SAL of 1x10-6 was performed).
    Ethylene Oxide (EO) Residual Limits MetAchieved (EO residual testing was performed).
    Product and Packaging Compatibility with Ethylene Oxide SterilizationAchieved (Product and packaging compatibility with ethylene oxide sterilization was confirmed).
    Substantial Equivalence to Predicate DevicesConcluded based on non-clinical testing showing safety and equivalent performance for intended use with the new sterilization method.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a separate "test set" in the context of an algorithm's performance on patient data. Instead, the testing pertains to the physical device and its sterilization process.

    • Sample Size for Sterilization Validation, EO Residual, and Material Compatibility Testing: The exact number of samples (devices and packaging materials) used for each test (sterilization validation, EO residual, product/packaging compatibility) is not explicitly stated in the provided text.
    • Data Provenance: The studies were non-clinical laboratory tests, likely conducted in a controlled lab environment. The country of origin is not specified, but the applicant (ConforMIS Inc.) is based in Bedford, MA, USA. The data is prospective in the sense that it was generated specifically for this submission to validate the new sterilization method.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is not a study involving human interpretation of data for medical diagnosis or prognosis that would require expert-established ground truth. The "ground truth" here is objective scientific measurements related to sterilization efficacy and material properties.

    4. Adjudication Method for the Test Set

    Not applicable, as this is not a study involving subjective assessments that would require an adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This submission concerns device sterilization and material compatibility, not the comparative effectiveness of an AI algorithm in a clinical setting.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This submission does not involve an algorithm or AI.

    7. Type of Ground Truth Used

    The ground truth used for this submission is based on:

    • Objective scientific standards and methodologies: For sterility validation (e.g., demonstrating a SAL of 1x10⁻⁶), EO residual limits (based on established toxicological standards), and material compatibility (physical and chemical integrity of the device and packaging after sterilization).
    • Regulatory requirements and consensus standards: As governed by the FDA and referenced in the submission (e.g., 21 CFR regulations).

    8. Sample Size for the Training Set

    Not applicable. This submission does not involve an AI algorithm with a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.