Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133192

    Validate with FDA (Live)

    Device Name
    MECTALIF TIPEEK
    Date Cleared
    2014-01-30

    (105 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MectaLIF implants in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of non-operative treatment prior to treatment with the devices.

    Device Description

    MectaLIF TiPEEK is characterized by different sizes of PEEK-OPTIMA LT1 (Polyetheretherketone) implants with a Titanium coating and Tantalum Markers. MectaLIF TiPEEK consists of the Oblique Interbody Fusion Device and Posterior Interbody Fusion Device. The Oblique and Posterior Interbody Fusion Devices are used to replace a degenerative disc in order to restore the height of the spinal column structure. MectaLIF TIPEEK can be applied with common surgical techniques such as PLIF (Posterior Lumbar Intervertebral Fusion) and TLIF (Transforaminal Lumbar Intervertebral Fusion). The devices are intended to be used in combination with posterior fixation (e.g. Pedicle Screw System) as well as an autogenous bone graft. The materials of MectaLIF TiPEEK are as follows: Implant: PEEK-OPTIMA LT1: Implant Grade Polyetheretherketone (ASTM F 2026) with commercially pure titanium (CPTi, ASTM F 1580) coating and Marker: Tantalum (ISO 13782 / ASTM F 560).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the MectaLIF TiPEEK device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that the MectaLIF TiPEEK "passed all requirements" for the performance tests. The specific quantitative acceptance criteria or threshold values are not provided in the text. However, the tests performed indicate the areas where the device's performance was evaluated against industry standards.

    Test CategorySpecific Test (ASTM Standard)Acceptance Criteria (Not explicitly stated in document)Reported Device Performance
    Mechanical StrengthStatic axial compression(Implied by standard conformance)Passed all requirements
    Dynamic axial compression(Implied by standard conformance)Passed all requirements
    Static shear compression(Implied by standard conformance)Passed all requirements
    Dynamic shear compression(Implied by standard conformance)Passed all requirements
    Static Tensile strength(Implied by standard conformance)Passed all requirements
    Static Shear strength(Implied by standard conformance)Passed all requirements
    Shear fatigue strength(Implied by standard conformance)Passed all requirements
    StabilitySubsidence(Implied by standard conformance)Passed all requirements
    Coating PerformanceAbrasion resistance(Implied by standard conformance)Passed all requirements
    Material AnalysisMicrographical Analysis(Implied by standard conformance)Passed all requirements
    WearWear Analysis(Implied by standard conformance, explicitly mentioned)Passed all requirements
    Wear analysis for particulates(Implied by standard conformance, explicitly mentioned)Passed all requirements

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "MectaLIF TiPEEK was tested using the worst-case device for each of the following tests".

    • Sample Size: For each specified mechanical and material test, the sample size was "the worst-case device." This implies a very limited sample size, likely one or a few devices representing the highest stresses or risks, common in medical device mechanical testing for substantial equivalence.
    • Data Provenance: The study was conducted as part of Medacta International's premarket notification (510(k)) to the FDA. The testing was performed in accordance with FDA Guidance documents. The location of the testing facility is not explicitly stated, but it's likely a laboratory, either internal or third-party, that adheres to relevant international standards (e.g., ISO, ASTM). This is a prospective study in the sense that the testing was performed before market clearance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable to this type of performance testing. The "ground truth" for mechanical and material performance is established by the specified ASTM standards and FDA guidance, which define the test methods, conditions, and expected outcomes (e.g., fracture loads, wear rates, material properties). Expert consensus is typically used for clinical endpoints or diagnostic accuracy, not mechanical device performance.

    4. Adjudication Method for the Test Set

    This is not applicable. As this is mechanical and material performance testing, there is no "adjudication" in the traditional sense of human interpretation of results. The results are quantitative measurements against predefined criteria within the ASTM standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not conducted or mentioned. This type of study is relevant for evaluating the clinical effectiveness or diagnostic accuracy of a device when interpreted by humans, typically in imaging or diagnostic applications. MectaLIF TiPEEK is an implantable intervertebral fusion device, and its performance is assessed through mechanical and material testing, not human interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

    No, a standalone performance study in the context of an algorithm or AI without human-in-the-loop was not conducted or mentioned. The device itself is a physical implant, not a software algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for this performance testing is based on established industry standards and regulatory guidance. Specifically:

    • ASTM Standards: F2077, F2267, F1877, F1978, F1147, F1044, F1160, F1854. These standards define the scientific methods and expected results for evaluating intervertebral body fusion devices and metallic coatings.
    • FDA Guidance Documents: "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device" and "Testing of Metallic Plasma Sprayed Coatings on Orthopedic Implants to Support Reconsideration of Postmarket Surveillance Requirements." These guidances provide regulatory expectations and testing protocols.

    8. Sample Size for the Training Set

    This is not applicable. The MectaLIF TiPEEK is a physical medical device. It does not involve a "training set" in the context of machine learning or artificial intelligence. Its design and manufacturing processes are informed by engineering principles, material science, and previous predicate device designs, but not a data-driven training set.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no "training set" for this device. The design and validation are based on engineering principles and compliance with established standards, not a data-driven training process.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1