510(k) Data Aggregation

    K Number
    K132897

    Validate with FDA (Live)

    Date Cleared
    2013-12-11

    (86 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CLYDESDALE® Spinal System is designed to be used with autogenous bone graft to facilitate interbody fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The CLYDESDALE® Spinal System is used for patients diagnosed with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via a minimally invasive lateral approach.

    Device Description

    The CLYDESDALE® Spinal System is intended to help provide support in the intervertebral body space during fusion of vertebral bodies in the lumbar spine. This system is intended to be used with supplemental fixation. The CLYDESDALE® Spinal System consists of PEEK cages of various widths and heights, which include tantalum markers. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allow them to be packed with autogenous bone graft. The purpose of this 510(k) submission includes additional implant sizes and trials for the CLYDESDALE® Spinal System. The subject implants will be offered in 18mm, 22mm, 26mm widths, 0°, 8°, 12°, 18° of lordosis and 8mm, 10mm, 12mm, 16mm. 18mm. 20mm heights, and in the same lengths as the predicate CLYDESDALE® Spinal System, K100175 (S.E. 06/02/2010).

    AI/ML Overview

    The information provided pertains to the CLYDESDALE® Spinal System, which is an intervertebral body fusion device. The acceptance criteria and the study that proves the device meets those criteria are detailed below.

    1. Table of Acceptance Criteria and Reported Device Performance

    Tests PerformedApplicable StandardsAcceptance CriteriaReported Device Performance
    Static Compression-ShearASTM F2077Not explicitly stated, but implies meeting or exceeding predicate device performance and safety."The subject implants met the acceptance criteria."
    Compression-Shear FatigueASTM F2077Not explicitly stated, but implies meeting or exceeding predicate device performance and safety."The subject implants met the acceptance criteria."

    The report states, "Based on the results, the subject implants demonstrated that they are as safe, as effective and perform as well as the predicate device(s)." This indicates that the acceptance criteria were met by showing equivalency to predicate devices in mechanical testing.

    2. Sample Size Used for the Test Set and the Data Provenance

    This submission did not involve clinical testing. The tests conducted were non-clinical mechanical tests on the device itself. Therefore, there is no "test set" in the context of patient data, nor is there data provenance in terms of country of origin or retrospective/prospective nature. The tests were performed on the medical devices (implants).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. As this was a non-clinical mechanical study, no human experts were involved in establishing a "ground truth" for a test set of patient data. The ground truth was established by the specified mechanical testing standards.

    4. Adjudication Method for the Test Set

    Not applicable. No human readers or adjudication method were involved as this was a non-clinical mechanical study.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted as this was a non-clinical mechanical study. No human readers or AI were involved in the evaluation of the device's performance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This device is a physical implant, not an algorithm or AI system. Therefore, a standalone algorithm performance study was not performed.

    7. Type of Ground Truth Used

    The ground truth for this study was based on established mechanical testing standards and the performance of legally marketed predicate devices. Specifically, the device's performance was compared to the requirements outlined in the FDA's guidance document, "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device," and demonstrated substantial equivalence to predicate devices.

    8. Sample Size for the Training Set

    Not applicable. This submission concerns a physical medical device, not an AI or machine learning model. There is no concept of a "training set" in this context.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for a physical device, no ground truth needed to be established for it.