Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K132690

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-02-25

    (181 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Abdomen, Cardiac, Small Organ (breast, testes, thvroid), heart soft tissue, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Pediatric, Fetal, Cephalic, and Ob/Gyn.

    The device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal/QB; Abdominal (GYN & Urology); Pediatric; Small Organ (breast, testes, thyroid); Cardiac (adult & pediatric); Peripheral Vascular, Musculo-skeletal Conventional & Superficial, Transrectal and Transvaginal.

    This device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Abdomen; Pediatric; Small Organ (breast, testes, thyroid); heart soft tissue; Peripheral Vascular; Musculo-skeletal; Ob/Gyn and Urology.

    The device is a general-purpose ultrasonic intended for use by a qualified physician for evaluation of Abdomen, Cardlac, Small Organ (breast, thyroid), heart soft tissue, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Pediatric, Fetal, Cephalic, Ob/Gyn, Trans-rectal, Trans-vaginal, Urology.

    Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal, Abdominal, Pediatric, Small Organ (Breast, Thyroid, Testes), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional), Musculo-skeletal (Superficial), Ob/GYN, Cardiac Adult, Cardiac Pediatric, Peripheral vessel.

    Device Description

    The UP600 diagnostic doppler ultrasound system is a compact and portable diagnostic ultrasound device, have integrated preprogrammed color ultrasound imaging system, capable of producing a resolution intended for clinical diagnostic imaging applications. The user interface includes a specialized control kevboard and color 15-inch LCD display. The digital architecture with progressive dynamic receive focusing allows the system to maximize the utility of all imaging transducers to enhance the diagnostic utility provided by the system. The exam dependent default setting allows the user to have minimum adjustment for imaging the in depth soft-menu control allows the advanced user to set the system for different situations. The architecture allows system integration to a variety of upgrade-able options and features.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification Summary for the BenQ Medical Technology UP600 Diagnostic Doppler Ultrasound System. This document does not describe specific acceptance criteria and a study proving the device meets those criteria in the typical format expected for AI/ML device evaluations. Instead, it focuses on demonstrating substantial equivalence to predicate devices, which is a different regulatory pathway.

    Therefore, many of the requested elements (sample sizes for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth types) are not applicable or extractable from this specific document, as it predates the widespread regulatory expectations for AI/ML device evaluations.

    Here's a breakdown of what can be extracted, along with explanations for what cannot:

    1. Table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria (e.g., specific sensitivity, specificity, or image quality metrics) or report performance against such criteria. Instead, it relies on demonstrating comparable technological characteristics and intended uses to legally marketed predicate devices. The "performance" aspect is implicitly shown by comparison to these predicate devices rather than through specific quantitative metrics presented for the UP600 itself.

    CategoryAcceptance Criteria (Not explicitly stated as such, but inferred from comparison)Reported Device Performance (as presented in the document)
    Compliance to StandardsMeeting relevant safety standards.UP600: NEMA UD 2, NEMA UD 3, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 10993-1. (Also explicitly states compliance with IEC601-1 and IEC60601-2 under safety compliance in comparison table).
    Indications for UseComparable to predicate devices.UP600: General-purpose ultrasonic imaging instrument for evaluation of Abdomen, Cardiac, Small Organ (breast, testes, thyroid), heart soft tissue, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Pediatric, Fetal, Cephalic, Ob/Gyn, Trans-rectal and Trans-vaginal. (Matches or expands upon predicate devices, with specific modes enabled for each application per transducer, as shown in the "Indications for Use" tables for each transducer).
    Device DesignSimilar architecture to predicate devices allowing for comparable functionality.UP600: 128 Channel all digital beam former. Compact and portable. User interface includes specialized control keyboard and color 15-inch LCD display. Progressive dynamic receive focusing. Wide band all digital demodulation. Native resolution digital scan converter. Features: 2D 8-mode, M-mode, Harmonic Image, Color, Power Doppler, Pulse wave Doppler, and CW. Supports 2D/M-mode, 2D/Pulsed Doppler, Triplex 2-D/CD/Pulsed Doppler, Dual B and Color in real time. (Similar to GE Voluson I (128 channel) and more advanced than CGMC OPUS 5000 (64 channel)).
    Measurement FunctionsComprehensive measurement capabilities.UP600: Distance; area; Volume; circumference; Heart Rate; calipers; velocity; PI, RI, Cardiac, OB/GYN and Vascular package. (Comparable to predicate devices).
    Acoustic OutputBelow applicable FDA limits.UP600: Track 3; MI, TIS, TIC, TIB. Derated Ispta: 720mW/cm2 maximum, TIS/TIB/TIC:0.1-4.0 Range, Mechanical Index: 1.9 Maximum, or Derated Isppa: 190 W/cm² max. (Identical to predicate devices).
    Patient Contact MaterialsBiocompatibility.UP600: RTV664+Ultrason S2010 silicon rubber complies with ISO10993-5. (Identical to CGMC OPUS 5000, different from GE RTV630A/B silicon rubber but also complying with ISO 10993-5).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. As a premarket notification for a general ultrasound system based on substantial equivalence, the submission primarily focuses on technical specifications, safety standards, and comparison to existing devices rather than clinical study data with specific test sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. No specific clinical test set requiring expert ground truth is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document. No specific clinical test set requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A Multi-Reader, Multi-Case (MRMC) comparative effectiveness study was not performed, nor is it applicable for this type of device submission. The UP600 is a diagnostic ultrasound system, a hardware device, and does not incorporate AI/ML functionality that would assist human readers in image interpretation. Therefore, there is no "AI vs without AI assistance" comparison.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance evaluation of an algorithm was not performed, nor is it applicable. The UP600 is a diagnostic ultrasound system, not an algorithm, and is intended for use by a qualified physician (human-in-the-loop).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not provided in the document. No specific ground truth for performance evaluation is mentioned, as the focus is on substantial equivalence.

    8. The sample size for the training set

    This information is not provided in the document. This device is a hardware system, not an AI/ML algorithm that would typically have a training set.

    9. How the ground truth for the training set was established

    This information is not provided in the document. Not applicable for this hardware device.


    Summary of the Study (or lack thereof, in the context of AI/ML evaluation):

    The BenQ Medical Technology UP600 Diagnostic Doppler Ultrasound System's acceptance was based on a demonstration of substantial equivalence to legally marketed predicate devices (BenQ Medical Technology UP600 K121983, GE Healthcare Voluson i K053435, and CGMC OPUS 5000 K102989). This is the primary "study" or justification presented in the 510(k) summary.

    The document highlights the following to support substantial equivalence:

    • Comparable Indications for Use: The UP600 and its transducers are intended for the same clinical applications as the predicate devices.
    • Similar Technological Characteristics: The UP600 shares core operational principles (applying high voltage burst to piezoelectric material, detecting reflected echoes, constructing 2D B-mode, Doppler color, and Doppler spectrum images) and many features (display modes, measurements, transducer types) with the predicate devices. Some features, like the 128-channel digital beam former, are even identical to one predicate (GE Voluson I) and an improvement over another (CGMC OPUS 5000).
    • Compliance with Safety Standards: The device meets recognized national and international safety standards (NEMA UD 2, NEMA UD 3, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 10993-1).
    • Equivalent Acoustic Output: The acoustic output levels are identical to the predicate devices and stated to be below applicable FDA limits ("Track 3; MI, TIS, TIC, TIB. Derated Ispta: 720mW/cm2 maximum," etc.).
    • Biocompatible Patient Contact Materials: Materials used for patient contact are compliant with ISO10993-5 (a standard for biological evaluation of medical devices).

    The 510(k) summary concludes that "The UP600 is substantially equivalent in safety and effectiveness to the predicate systems." This implies that by meeting the technical and safety specifications and having comparable intended uses to already cleared devices, it is deemed safe and effective without the need for extensive clinical performance studies that would typically involve test sets, expert ground truth, or AI-specific evaluations.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1