Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K132629

    Validate with FDA (Live)

    Date Cleared
    2013-12-31

    (131 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reprocessed Bladeless Trocar has applications in abdominal, thoracic, and gynecologic minimally invasive surgical procedures to establish a path of entry for endoscopic instruments. The trocar may be used with or without visualization for primary and secondary insertions.

    The Reprocessed Dilating Tip Trocar has applications in thoracic, gynecologic laparoscopy and other abdominal procedures to establish a path of entry for endoscopic instruments.

    The Reprocessed Stability Sleeve has application in thoracic, gynecologic laparoscopy and other abdominal procedures to establish a path of entry for endoscopic instruments.

    Device Description

    Reprocessed Bladeless Trocar: The Reprocessed Bladeless Trocar, is a sterile single patient use instrument consisting of a radiolucent sleeve (cannula) and obturator. The obturator contains a clear, tapered optical element. The 5 mm obturator accommodates an appropriately sized 0° endoscope and provides visibility of individual tissue layers during insertion. The trocar sleeve contains two seals that accommodate instruments 5 mm in diameter. Together, these two seals minimize gas leakage when instruments are inserted or withdrawn through the trocar. A stopcock valve is compatible with standard luer lock fittings and provides attachment for gas insufflation and desufflation. The stopcock is in the closed position when it is parallel to the sleeve.

    Reprocessed Dilating Tip (Bladed) Trocar: The Reprocessed Dilating Tip Trocar is a sterile, single patient use instrument consisting of a radiolucent sleeve (cannula) and obturator. The obturator has a sharp, flat-bladed tip and spring-loaded shield is designed to cover the flat-bladed tip to protect internal structures from puncture or laceration once the abdominal or thoracic cavity has been entered. The trocar sleeve contains two seals that accommodate instruments 5 mm in diameter. Together, these two seals minimize gas leakage when instruments are inserted or withdrawn through the trocar. A stopcock valve is compatible with standard luer lock fittings and provides attachment for gas insufflation and desufflation. The stopcock is in the closed position when it is parallel to the sleeve.

    Reprocessed Stability (Universal) Sleeve (Cannula): The Reprocessed Stability Sleeve is a sterile, single patient use device with a radiolucent sleeve. The trocar sleeve contains two seals that accommodate instruments 5 mm in diameter. Together, these two seals minimize gas leakage when instruments are inserted or withdrawn through the trocar. A stopcock valve is compatible with standard luer lock fittings and provides attachment for gas insufflation and desufflation. The stopcock is in the closed position when it is parallel to the sleeve.

    AI/ML Overview

    The document provided describes the reprocessing of surgical trocars and sleeves, and the performance data focuses on demonstrating that the reprocessed devices are as safe and effective as new (predicate) devices. This is not the type of medical device for which the requested acceptance criteria and study information (AI/radiology performance metrics, expert adjudication, etc.) would typically apply.

    The device is a reprocessed medical instrument, not an AI or diagnostic imaging device. Therefore, the performance evaluation centers on its physical integrity, sterility, and functionality after reprocessing, rather than diagnostic accuracy or human-AI interaction.

    Here's an analysis based on the provided text, highlighting why most of your requested points are not applicable in this context:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table with specific quantitative acceptance criteria for each test and reported pass/fail results. Instead, it lists the types of tests performed.

    Acceptance Criteria CategoryReported Device Performance (Summary)
    BiocompatibilityTested and performs as originally intended.
    Validation of ReprocessingTested and performs as originally intended.
    Sterilization ValidationTested and performs as originally intended.
    Function Test(s)Each individual device is tested for appropriate function of its components prior to packaging. Tested and performs as originally intended.
    Packaging ValidationTested and performs as originally intended.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample sizes for the performance tests (biocompatibility, reprocessing validation, sterilization, function, packaging). It also does not specify the country of origin of the data or whether the studies were retrospective or prospective. Given the nature of these tests (bench and laboratory), these details are less critical than for clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable. The "ground truth" for a reprocessed medical instrument involves objective measurements of sterility, material integrity, and mechanical function, not diagnostic interpretations by human experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical trials, particularly in image interpretation, to resolve discrepancies between readers. This device undergoes bench and laboratory testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. An MRMC study is relevant for diagnostic devices (e.g., AI in radiology). This submission is for reprocessed surgical instruments. Human "improvement" or AI assistance is not a relevant metric here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. This device is a physical surgical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    As mentioned in point 3, the "ground truth" for this device relates to objective physical and biological properties. It includes:

    • Sterility: Absence of viable microorganisms (validated through sterilization methods).
    • Biocompatibility: Absence of adverse biological reactions (tested with accepted standards).
    • Functional Performance: Mechanical integrity and operation of components (e.g., seals, bladed tip, stopcock valve, resistance to leakage, insertion force).
    • Cleaning Efficacy: Absence of residual soil after reprocessing.
    • Packaging Integrity: Maintenance of sterility barrier.

    These are established through standardized laboratory methods and specifications, not expert consensus or pathology in the diagnostic sense.

    8. The sample size for the training set

    This question is not applicable. This device is a physical instrument, not an AI algorithm, so there is no "training set."

    9. How the ground truth for the training set was established

    This question is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1