Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K132610

    Validate with FDA (Live)

    Date Cleared
    2013-09-04

    (15 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The G52SP+/G52S+ 5MP Grayscale LCD Display is intended to be used in displaying and viewing digital images, including digital mammography, for review and analysis by trained medical practitioners.

    Device Description

    The G52SP+/G52S+ 5MP Grayscale LCD display is a display system for medical viewing, with high resolution 2560 x 2048, built-in brightness stabilization circuit, front sensor and ambient light sensor, stable brightness and persistent calibration can be guaranteed. The display can support both landscape and portrait mode. The anti-reflection coated protective screen can prevent display from damage under hard using conditions, make the clean and disinfect easier.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical display, not a study evaluating a device's performance against acceptance criteria. Therefore, most of the requested information regarding acceptance criteria, study design, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for a device or algorithm is not present in this document.

    This document describes a medical display (G52SP+/G52S+ 5MP Grayscale LCD Display) and argues for its substantial equivalence to a predicate device (Barcoview, Nio 5MP-21") based on technical specifications and intended use. It is a regulatory submission for market clearance, not a performance study.

    However, I can extract the following information that is somewhat relevant to the request:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't present "acceptance criteria" in the typical sense of a performance study with quantitative thresholds for diagnostic accuracy. Instead, it compares the proposed device's specifications to those of a predicate device to argue for "substantial equivalence." The "performance" refers to technical specifications rather than clinical performance.

    Comparison Item (ID)Proposed Device (G52SP+/G52S+) PerformancePredicate Device (Nio 5MP-21) PerformanceDiscussion of Difference & Substantial Equivalence Claim
    Intended UseDisplaying and viewing digital images, including digital mammography, for review and analysis by trained medical practitioners.Displaying and viewing digital images, including digital mammography, for review and analysis by trained medical practitioners.Same; SE not affected.
    Panel Size and Type21.3", TFT LCD display21.3", TFT LCD displaySame; SE not affected.
    Pixel Pitch0.165 mm0.165 mmSame; SE not affected.
    Native Resolutions2560×20482560×2048Same; SE not affected.
    Brightness1200 cd/m²700 cd/m²Proposed device brightness is higher, stated as "better than" the predicate, thus SE not affected.
    Contrast Ratio1200:1800:1Proposed device contrast ratio is higher, stated as "better than" the predicate, thus SE not affected.
    Network InterfaceUSB(1 Up, 2 Downstream)USB(1 Up, 2 Downstream)Same; SE not affected.
    Active Display Size422mm x338mm(HxV)422mm x338mm(HxV)Same; SE not affected.
    DimensionsVaried (e.g., 398.5mm (W) x 483.0mm (H) x 70.8mm (D) without Stand)Varied (e.g., 382mm x 488mm x 114mm without Stand)Comparable; SE not affected.
    Operating Temperature0°C ~ 40°C0°C ~ 40°CSame; SE not affected.
    Transport/Storage Temperature-20°C ~ 60°C-20°C ~ 60°CSame; SE not affected.
    Operating Relative Humidity15% ~85%8% ~80%(non-condensing)Comparable; SE not affected.
    Transport/Storage Relative Humidity10% ~90%5% ~95%Comparable; SE not affected.
    Power Capacity<80W95WBoth comply with IEC 60601-1 and IEC 60601-1-2; SE not affected.
    Input VoltageDC12V/5.0A100~250vBoth comply with IEC 60601-1 and IEC 60601-1-2; SE not affected.
    UsabilityButton operation, LED indicatorButton operation, LED indicatorSame; SE not affected.
    Mode of operationContinuous operationContinuous operationSame; SE not affected.
    Biocompatibility EvaluationDevice does not contact patients; evaluation not needed.Device does not contact patients; evaluation not needed.Same; SE not affected.
    SterilizationNot needed.Not needed.Same; SE not affected.
    Electrical, Mechanical, Thermal SafetyEvaluated per IEC 60601-1.Evaluated per IEC 60601-1.Same; SE not affected.
    EMC EvaluationComplying with IEC 60601-1-2.Complying with IEC 60601-1-2.Same; SE not affected.

    Missing Information:

    • 2. Sample sized used for the test set and the data provenance: Not applicable, as this is a device specification comparison, not a clinical or performance study involving a test set of data.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document does not describe a study involving human readers or AI.
    • 6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This document describes a display monitor, not an algorithm.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. The "ground truth" here is the documented specifications of the predicate device for comparison.
    • 8. The sample size for the training set: Not applicable. This is not an AI/algorithm device.
    • 9. How the ground truth for the training set was established: Not applicable. This is not an AI/algorithm device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1