510(k) Data Aggregation

    K Number
    K132584

    Validate with FDA (Live)

    Date Cleared
    2013-12-04

    (110 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PERIMETER® C Spinal System is intended to be used for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc who have had six weeks of nonoperative treatment. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. Additionally, the PERIMETER® C Spinal System implants are to be used with autogenous bone graft and supplemental fixation and implanted via an open, anterior approach.

    Device Description

    The PERIMETER® C Spinal System consists of spacers/cages of various widths and heights, which can be inserted between two cervical bodies to give support and correction during cervical interbody fusion procedures. Additionally, this implant has six degrees of lordosis and the superior and inferior surfaces of the implant are designed with teeth which interact with the surface of the vertebral endplates to aid in resisting expulsion. The hollow geometry of the implants allows them to be packed with autogenous bone graft and is to be used with supplemental fixation in all procedures.

    The purpose of this submission is to include additional interbody cages manufactured from medical grade titanium alloy (Ti-6A1-4V ELI) and designed with lateral windows. The lateral windows allow for visibility of bone graft placement. The subject device is offered in a non-sterile form.

    AI/ML Overview

    The Medtronic PERIMETER® C Spinal System is a medical device for anterior cervical interbody fusion procedures. The study demonstrates substantial equivalence to existing predicate devices based on non-clinical mechanical testing and Finite Element Analysis (FEA), following FDA guidance document, "Class II Special Controls Guidance Document: Intervertebral Body Fusion Devices".

    1. Acceptance Criteria and Device Performance:

    Test PerformedApplicable StandardAcceptance Criteria (Implicitly: Meet/Exceed Predicate Performance)Reported Device Performance
    Static Torsion TestingASTM F2077 (Test Methods for Intervertebral Body Fusion Devices)Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.
    Static Compression Bending TestingASTM F2077 (Test Methods for Intervertebral Body Fusion Devices)Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.
    Static Compression Shear TestingASTM F2077 (Test Methods for Intervertebral Body Fusion Devices)Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.
    Dynamic Torsion TestingASTM F2077 (Test Methods for Intervertebral Body Fusion Devices)Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.
    Dynamic Compression Bending TestingASTM F2077 (Test Methods for Intervertebral Body Fusion Devices)Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.
    Dynamic Compression Shear TestingASTM F2077 (Test Methods for Intervertebral Body Fusion Devices)Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.
    Subsidence TestingASTM F2267 (Standard Test Method for Measuring Load Induced Subsidence of the Intervertebral Body Fusion Device under Static Axial Compression)Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.
    Expulsion TestingDRAFT ASTM F-04.25.02.02Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.

    The study concludes that "Based on the results the subject intervertebral devices demonstrated that they are as safe and effective as the predicate device(s)."

    2. Sample Size and Data Provenance for Test Set:

    No specific sample sizes for mechanical testing are provided in the document. The testing was non-clinical (laboratory-based) and not based on patient data, so "data provenance" in terms of country of origin or retrospective/prospective is not applicable.

    3. Number of Experts and Qualifications for Ground Truth (Test Set):

    Not applicable. This was a non-clinical, mechanical engineering study, not involving human expert assessment for ground truth.

    4. Adjudication Method for Test Set:

    Not applicable. As this was a non-clinical mechanical test, there was no adjudication method involving human interpretation of results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study was done, as this was a non-clinical mechanical engineering study and did not involve human readers or AI assistance.

    6. Standalone Performance Study:

    Yes, a standalone performance study was done for the device in the sense of mechanical testing of the device itself (algorithm only without human-in-the-loop performance). The tests listed above are all standalone tests of the device's mechanical properties.

    7. Type of Ground Truth Used:

    The "ground truth" for the non-clinical tests was based on the performance requirements specified in the FDA guidance document "Class II Special Controls Guidance Document: Intervertebral Body Fusion Devices" and the applicable ASTM standards (ASTM F2077, ASTM F2267, and DRAFT ASTM F-04.25.02.02). The device's performance was compared against these established engineering standards and implicitly, against the performance of predicate devices.

    8. Sample Size for Training Set:

    Not applicable, as this was a non-clinical mechanical engineering study, not involving machine learning or a "training set" of data in the conventional sense.

    9. How Ground Truth for Training Set Was Established:

    Not applicable, as there was no training set in the context of this non-clinical mechanical study.