510(k) Data Aggregation

    K Number
    K132583

    Validate with FDA (Live)

    Date Cleared
    2014-02-11

    (179 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EZ-10 Intraosseous Infusion System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.

    Device Description

    The EZ-IO Intraosseous Infusion System including previously cleared K091140 EZ-MIO Distal Tibia, EZ-10 Distal Tibia; EZ-10 (formerly Vidaport) Intraosseous Infusion System ; EZ-IO Humeral Head and K101026 Powered PD IO Infusion System; EZ-IO Humeral Head; Powered PD-IO is designed to allow the user to insert a needle set consisting of a stylet and catheter through the cortex of the bone to a desired depth within the bone marrow to facilitate intraosseous infusion of desired fluids for vascular access. All system materials are medical grade. Needle sets are single-use and composed of 304 stainless steel with polycarbonate hubs and available in 15 mm (3-39 kg); 25 mm (40 kg or over) and 45 mm (40 kg or over). Black lines on the needle set catheter serve as depth markers. The needle sets connect to the driver/drill/ manual handle shaft with a magnetic disc. The reusable cordless driver/drill is powered by lithium batteries with a battery-power indicator light. The EZ-MIO consists of a manual handle device which is primarily used by the armed forces.

    Clinicians locate anatomical landmarks and clean the insertion site. Using the cordless driver or manual handle with needle set attached the needle set is pressed through the soft tissue to the outer cortex of the bone. Depth markers on the catheter must be visible prior to powering driver or manually inserting the needle set to ensure placement within the medullary space. If using the powered driver clinicians then squeeze the driver trigger and apply moderate, steady pressure. Trigger is released when a sudden "give" or "pop" is felt, which indicates entry into medullary space; the catheter will not always be inserted to the hub. If using the manual handle, clinicians apply downward pressure and rotation through the bone until a change in pressure is felt as a "give" or "pop" or desired depth is reached. For both methods after insertion of the needle set, the driver unit is detached from the needle set, leaving the stylet and cannula firmly seated in the bone. The stylet is then separated and removed from the catheter by turning the stylet hub counter clockwise leaving a standard Luer lock catheter securely seated in the bone. The catheter Luer lock permits attachment of standard syringes and IV tubing for administration of medications and fluids.

    This submission requests a label change only to the 25 mm needle set from "40 kg or over" to" 3 kg or over" utilizing the same technique and devices previously cleared via 510(k)s K091140 and K101026.

    AI/ML Overview

    The provided text describes the EZ-IO Intraosseous Infusion System and a 510(k) submission for a label change. However, it does not contain information about acceptance criteria, specific studies proving the device meets those criteria, device performance metrics, sample sizes, expert qualifications, or ground truth establishment relevant to an AI/ML device.

    The submission is for a label change for a physical medical device (intraosseous needle) to expand its approved weight range, not for a software-as-a-medical-device (SaMD) or an AI/ML component. The "study" mentioned refers to cadaveric pediatric studies demonstrating safety for the expanded needle length, rather than performance metrics of an AI model.

    Therefore, I cannot fulfill your request for the specific AI/ML related information (such as reported device performance, sample sizes for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, or ground truth details) because this information is not present in the provided document.

    The document states:

    • "This submission requests a label change only to the 25 mm needle set from '40 kg or over' to '3 kg or over' utilizing the same technique and devices previously cleared via 510(k)s K091140 and K101026."
    • "The labeling change requires no new technology to facilitate the safe application of the product."
    • "There have been no changes to the design or components of the Vidacare devices..."

    This indicates that no new performance studies (especially not for an AI/ML component) were conducted or are being presented here. The "performance data" mentioned in the summary refers to existing data and clinical observations supporting the expanded weight range of the physical needle.