510(k) Data Aggregation

    K Number
    K132256

    Validate with FDA (Live)

    Date Cleared
    2013-08-07

    (19 days)

    Product Code
    Regulation Number
    866.3328
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD Veritor™ System for Rapid Detection of Flu A+B is a rapid chromatographic immunoassay for the direct and qualitative detection of influenza A and B viral nucleoprotein antigens from nasopharyngeal wash, aspirate and swab in transport media samples from symptomatic patients. The BD Veritor System for Rapid Detection of Flu A+B is a differentiated test, such that influenza A viral antigens can be distinguished from influenza B viral antigens from a single processed sample using a single device. The test is to be used as an aid in the diagnosis of influenza A and B viral infections. A negative test is presumptive and it is recommended that these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative test results do not preclude influenza viral infection and should not be used as the sole basis for treatment or other patient management decisions. The test is not intended to detect influenza C antigens.

    Device Description

    The BD Flu A+B test is a chromatographic assay to qualitatively detect influenza A and B viral antigens in respiratory specimens. The patient specimen is mixed in a prefilled unitized tube containing RV Reagent D and added to the test device. RV Reagent C contains mucolytic agents that function to break down mucus in a patient specimen thereby exposing viral antigens and enhancing detection in the assay device. Processed specimens are expressed through a filter tip into a single sample well on the BD Flu A+B test device.

    The specimen is mixed and added to the test device where influenza A or influenza B viral antigens bind to anti-influenza antibodies conjugated to detector particles on the BD Flu A+B test strip. The antigen-conjugate complex migrates across the test strip to the reaction area and is captured by an antibody line on the membrane. The assay utilizes a proprietary enhanced colloidal-gold particle at the test lines as the means for identifying the presence of influenza A or B viral antigens.

    The BD Flu A+B test devices are designed with five spatially-distinct zones including positive and negative control line positions, separate test line positions for the target analytes, and a background zone. The test lines for the target analytes are labeled on the test device as 'A' for Flu A position, and 'B' for Flu B position. The onboard positive control ensures the sample has flowed correctly and is indicated on the test device as 'C'. Two of the five distinct zones on the test device are not labeled. These two zones are an onboard negative control line and an assay background zone. The onboard negative control zone addresses non-specific signal generation and is not labeled on the test device. The remaining zone is used to measure the assay background and is also not labeled.

    The BD Flu A+B assay incorporates an active negative control feature in each test to identify and compensate for sample-related, nonspecific signal generation. The BD Veritor™ System Reader uses a proprietary algorithm which subtracts nonspecific signal at the negative control line from the signal present at both the Flu A and Flu B test lines. If the resultant test line signal is above a pre-selected assay cutoff, the specimen is scored as positive. If the resultant test line signal is below the cutoff, the specimen is scored as negative. Use of the active negative control feature allows the BD Veritor™ System reader to correctly interpret test results that cannot be scored visually because the human eye is unable to accurately perform the subtraction of the nonspecific signal.

    AI/ML Overview

    This 510(k) summary describes a modification to an existing device, the BD Veritor™ System Flu A+B assay Clinical kit, primarily to update labeling with analytical sensitivity data for a new Influenza A strain (A/Anhui/1/2013 H7N9) and add strain reactivity tables. It explicitly states that no new issues of safety and effectiveness were identified for this addition and that the change did not change the intended use of the device or the fundamental scientific technology.

    Therefore, the document does not contain details about a study designed to establish acceptance criteria for the original device or to demonstrate that the original device meets such criteria through clinical performance. Instead, it focuses on the analytical performance related to detecting specific influenza strains to ensure the updated labeling is accurate.

    Given this, I cannot provide a table of acceptance criteria and reported device performance from this document for the overall device's clinical efficacy, nor details about a comprehensive clinical study as typically conducted for initial device clearance. The information provided heavily relates to analytical performance in detecting specific viral strains for labeling purposes, rather than a full clinical performance study for the entire device.

    However, I can extract information related to the analytical performance presented for the labeling update, and other study-related details where available based on the document.

    Here's the breakdown based on the provided text, focusing on what is available:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the traditional sense for clinical performance metrics (like sensitivity/specificity targets). Instead, it discusses the establishment of analytical sensitivity (Limit of Detection - LOD) for specific influenza strains to enrich the product labeling.

    Performance Metric (Analytical)Reported Device Performance
    Analytical Sensitivity (LOD) for A/Anhui/1/2013 H7N9Data was added to the labeling reflecting the device's ability to detect this cultured virus. (Specific LOD values are not provided in this summary, only the fact that data was added).
    Strain Reactivity tablesAdded to labeling, indicating the device's reactivity (detection capabilities) against various influenza strains. (Specific data points are not provided in this summary, only the fact that tables were added).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated for the analytical sensitivity (LOD) and strain reactivity testing.
    • Data Provenance: The new data relates to cultured virus (A/Anhui/1/2013 H7N9) testing. The document also mentions general clinical performance characteristics for the original device were established:
      • For NP washes/aspirates: January – March 2011, United States (predominant strains: A/2009 H1N1, A/H3N2, B/Victoria, B/Yamagata).
      • For NP swabs in transport media: January – April 2012, United States (predominant strains: A/2009 H1N1, A/H3N2, B/Victoria, B/Yamagata).
      • These dates and predominant strains suggest prospective collection of clinical samples during influenza seasons in the US for the original device's performance characterization, but detailed methodology for that original study is not provided in this 510(k) summary. The current 510(k) is about adding analytical data for a new strain.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable/Not provided for the analytical sensitivity and strain reactivity additions. This type of testing typically involves laboratory-based methods rather than expert clinical consensus for "ground truth".

    4. Adjudication Method for the Test Set

    Not applicable/Not provided. The analytical tests would follow standardized laboratory protocols for determining LOD and reactivity, not clinical adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a rapid chromatographic immunoassay, not an AI-assisted diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The BD Veritor™ System Reader uses a "proprietary algorithm" to interpret test results by subtracting non-specific signal. The document states: "Use of the active negative control feature allows the BD Veritor™ System reader to correctly interpret test results that cannot be scored visually because the human eye is unable to accurately perform the subtraction of the nonspecific signal." This implies the "reader" (which performs the algorithm-based interpretation) operates in a standalone manner to provide results, effectively without human-in-the-loop performance for the final interpretation of the test strip signal. Specific performance metrics for this algorithm alone are not detailed in this summary, but its function is described as essential for accurate test interpretation.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)

    For the analytical sensitivity and strain reactivity data added: The ground truth would be the known presence and concentration of the cultured virus or specific viral strains, confirmed by reference laboratory methods.

    For the original clinical performance of the device (referenced for historical context but not the subject of this 510(k) modification): The indication states, "A negative test is presumptive and it is recommended that these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay." This suggests that viral culture or FDA-cleared molecular assays likely served as the ground truth for establishing the original clinical performance characteristics.

    8. The Sample Size for the Training Set

    Not applicable/Not provided. This 510(k) summary is for a modification based on analytical testing, not a new device requiring a training set for model development. The original device's development (which is not detailed here) might have involved internal validation data, but that's not described as a "training set" in this context.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable/Not provided for the reasons stated above.


    Summary of Device Performance (from the document's new information context):

    The core change in this 510(k) is the addition of data for Analytical Sensitivity (LOD) and Strain Reactivity tables for A/Anhui/1/2013 H7N9 to the product labeling. This indicates that the device has been analytically demonstrated to detect this specific novel avian influenza strain in a cultured setting. The document explicitly includes a disclaimer required by the FDA: "Although this test has been shown to detect the novel avian influenza A(H7N9) cultured virus, the performance characteristics of this device with clinical specimens that are positive for the novel avian influenza A(H7N9) virus have not been established." This clarifies that while analytical detection is shown, clinical performance with patient samples of H7N9 was not part of this submission.