Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K132123

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-03-20

    (253 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TRC-Pex: The TRC-Pex is a temporary root canal sealer for use in the treatment of root canals, following a pulpectomy or for apexegenesis or apexification.
    Radic-Sealer: The Radic-Sealer is a root canal sealing of root canals following established endodontic procedures and may be used in conjunction with the auxiliary materials in the root canal (i.e. gutta percha points). The Radic-Sealer is intended for use by qualified healthcare personnel trained in its use.

    Device Description

    TRC-Pex: TRC-Pex is an immediately available pre-mixed root canal filling material based on Calcium Hydroxide and lodoform. This pre-mixed temporary filling material maintains a constant flow, which makes it easy to inject into the canal. Depending on apicoectomy or root canal therapies, it can be used as a temporary root canal filling material. The device is contained in a plastic syringe and the system includes a plunger, disposable tips, a rubber, a protective cap, an indo stop and a holder for direction control of the tip.
    Radic-Sealer: Radic-Sealer is a root canal sealing & filling material based on epoxy-amine resin. Radic-Sealer is contained in a dual syringe which makes it easy to mix and inject. It makes permanent obturation of root canals using adhesion properties between root canal walls and the root canal filling materials. The base and catalyst are contained separately in the chambers of two separate plastic syringes. The product consists of a mixing tip with a mixing joint, a protective cap, a mixing pad, and a plunger.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the TRC-Pex and Radic-Sealer devices:

    This document is a 510(k) summary for medical devices, which focuses on demonstrating substantial equivalence to predicate devices rather than proving a new device's safety and effectiveness from scratch through extensive clinical trials. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are framed within the context of showing equivalence to existing, legally marketed devices.

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly derived from the standards and properties of the predicate devices. The study aims to show that the new devices meet the established performance parameters defined by these standards.

    Acceptance Criteria (Based on ISO Standards)Reported Device Performance (TRC-Pex & Radic-Sealer)
    ISO 6876: RadiopacityConfirmed to protocols and recommended values
    ISO 6876: LiquidityConfirmed to protocols and recommended values
    ISO 6876: Setting timeConfirmed to protocols and recommended values
    ISO 6876: Solubility testConfirmed to protocols and recommended values
    ISO 6876: Film ThicknessConfirmed to protocols and recommended values
    ISO 6876: FlowabilityConfirmed to protocols and recommended values
    ISO 10993-5: CytotoxicitySimilar biocompatibility properties (comparative test with predicate)
    ISO 10993-10: Oral Mucous Irritation & SensitizationBiocompatibility data provided
    ISO 10993-11: Short-term systemic toxicity (Oral)Biocompatibility data provided
    Other Bench Testing: Shelf lifeConfirmed
    Other Bench Testing: AppearanceConfirmed
    Other Bench Testing: Volume/weight specConfirmed
    Other Bench Testing: Packaging testsConfirmed

    Study Proving Acceptance Criteria:

    The study conducted was a series of non-clinical performance tests and biocompatibility tests. The document states:

    • "Performance testing that confirmed to the protocols and recommended values described in ISO 6876 was performed."
    • "Biocompatibility data were provided to demonstrate substantial equivalence to the predicate devices."
    • "Side-by-Side cytotoxicity testing was performed that compared the TRC-Pex and a predicate device, and the test results showed that they have similar biocompatibility properties."
    • "The bench and biocompatibility tests provided evidence that the chemical and physical properties of TRC-Pex and Radic-Sealer are substantially equivalent to the predicate devices."

    Additional Information Not Present in the Document:

    The provided text describes a 510(k) submission for dental materials (root canal filling and sealing materials). For this type of device, the performance evaluation primarily relies on bench testing and biocompatibility studies as per ISO standards, which is what the document reports. It does not involve studies typically associated with AI/software devices or diagnostic image analysis. Therefore, many of the requested points below are not applicable or cannot be answered from the provided text.


    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated for each test, but standard protocols for material testing (e.g., ISO 6876, ISO 10993 series) involve a specified number of samples for each property being evaluated (e.g., usually 3-5 samples per test condition for material science, but this is not detailed here).
    • Data Provenance: The tests were conducted by KM Corporation, likely in a laboratory setting. No country of origin for "data" in the sense of patient data is applicable. These are material properties tests.
    • Retrospective or Prospective: Not applicable. These are laboratory-based material characterization tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. For material properties and biocompatibility, "ground truth" is established by adherence to standardized testing methods (e.g., ISO standards) and instrument measurements, not expert consensus in the human interpretation sense.

    4. Adjudication method for the test set:

    • Not applicable. Material property tests are objective measurements, not subject to human adjudication or interpretation disputes.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI/imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is not an AI/software device.

    7. The type of ground truth used:

    • Bench Test Ground Truth: Measured physical and chemical properties according to ISO standards.
    • Biocompatibility Ground Truth: Cellular responses (e.g., cytotoxicity) and tissue reactions according to ISO 10993 standards and comparison to established predicate device performance.

    8. The sample size for the training set:

    • Not applicable. These materials are not "trained" in the machine learning sense. The "training" might refer to the development and refinement process of the material formulations, which is not detailed here.

    9. How the ground truth for the training set was established:

    • Not applicable for this type of device submission.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1