Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K131958

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-01-09

    (196 days)

    Product Code
    Regulation Number
    886.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BLIS is indicated for the folding and injection of Bausch + Lomb intraocular lenses identifying the BLIS in their approved labeling.

    Device Description

    The Bausch + Lomb Iniector System is used for folding and delivering validated IOLs into the eye. The system is composed of two items: a single-use, sterile, disposable cartridge and a reusable handpiece. The handpiece has two elements: a plunger and a body. The cartridge snaps into the handpiece.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Bausch + Lomb Injector System

    This document describes the acceptance criteria and the study that demonstrates the Bausch + Lomb Injector System (BLIS) meets these criteria, based on the provided FDA 510(k) summary (K131958).

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state numerical acceptance criteria in a dedicated table format. Instead, it describes various tests performed to ensure the device's safety and effectiveness. The "acceptance criteria" are implied to be successful completion and conformity to relevant standards and established good practices for medical devices of this type. The "reported device performance" is the successful outcome of these tests.

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    Cleaning ValidationManual and automated cleaning processes must be effective in preparing the handpiece for sterilization.Cleaning validation covered the manual and automated cleaning processes described in the labeling for the Bausch + Lomb Injector System Handpiece. (Implies successful validation).
    Sterilization ValidationAutoclave methods for handpiece and ethylene oxide (EtO) process for cartridge must achieve sterility.Sterilization validation included the autoclave methods described in the labeling for the Bausch + Lomb Injector System Handpiece and the ethylene oxide process for the Bausch + Lomb Injector System Cartridge. (Implies successful validation).
    Sterilization Residual TestingEthylene oxide residuals must be within acceptable limits for patient safety.Ethylene oxide residual testing for a device equivalent to the Bausch + Lomb Injector System Cartridge was included in the 510(k). (Implies compliance with limits).
    Shipping and Handling ValidationDevice integrity and functionality must be maintained after exposure to typical shipping and handling conditions.Shipping and handling validations were performed on both the Bausch + Lomb Injector System Handpiece and Cartridge. (Implies successful validation).
    Stability TestingThe cartridge must maintain its integrity and functionality over its specified shelf life.Stability testing validated the Bausch + Lomb Injector System Cartridge over the shelf life of the device. (Implies successful validation).
    BiocompatibilityDevice components in contact with the patient (handpiece and cartridge) must be biocompatible, showing no adverse biological reactions.Biocompatibility testing was performed on the Bausch + Lomb Injector System Handpiece and cartridge, and both were found to be biocompatible.
    Mechanical Properties (Bench Testing)Device components must meet the mechanical requirements outlined in ISO 11979-3.Bench testing was performed on both components of the Bausch + Lomb Injector System to demonstrate compliance with ISO 11979-3, Mechanical Properties.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided 510(k) summary does not specify numerical sample sizes for the test sets used in cleaning, sterilization, shipping, stability, biocompatibility, or bench testing. It indicates that these tests "were performed" or "were included," implying sufficient samples for validation.

    The data provenance is internal to the manufacturing and testing processes of Bausch + Lomb. The tests were non-clinical (bench-top and laboratory-based) rather than involving human subjects. Therefore, specific country of origin or retrospective/prospective distinctions as would apply to clinical data are not mentioned.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    N/A. This device is an injector system, not an AI or diagnostic device that requires expert ground truth for interpretation of output. The ground truth for its performance is derived from standardized testing methodologies and compliance with industry standards (e.g., ISO 11979-3).

    4. Adjudication Method

    N/A. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation or diagnosis, such as those evaluating AI for medical imaging. This device's testing is focused on physical and biological properties.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. An MRMC study is not applicable to an intraocular lens injector system. These studies are typically performed for diagnostic devices where human readers interpret medical data (e.g., images) and the impact of AI assistance on their performance is measured. The BLIS is a surgical tool.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Study was Done

    Yes, in a conceptual sense. All the tests described (cleaning, sterilization, shipping, stability, biocompatibility, bench testing) assess the device's inherent properties and functionality without human intervention in the performance of the core mechanical/biological tests. The device's operation, however, does require a human operator to physically use it. The tests evaluate the "algorithm" of the device's design and materials.

    7. The Type of Ground Truth Used

    The ground truth used for the non-clinical tests is objective physical measurements and biological assays that conform to established industry standards and regulatory requirements. This includes:

    • Sterility assurance levels
    • Residual levels of sterilants
    • Material properties (strength, durability, etc.) as defined by ISO 11979-3
    • Biocompatibility endpoints as per relevant ISO standards (e.g., ISO 10993 series).

    8. The Sample Size for the Training Set

    N/A. This is not an AI/machine learning device, so there is no concept of a "training set" in the context of algorithm development.

    9. How the Ground Truth for the Training Set Was Established

    N/A. As there is no training set for an algorithm, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1