510(k) Data Aggregation

    K Number
    K131511

    Validate with FDA (Live)

    Date Cleared
    2013-07-05

    (38 days)

    Product Code
    Regulation Number
    878.4750
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Circular Stapler has applications throughout the alimentary tract for the creation of end-to-end, end-to-side and side-to-side anastomoses in both open and laparoscopic surgeries.

    The Disposable Linear Staplers and Reloads have application in the resection or transection of tissue for abdominal, gynecological, pediatric and thoracic surgical procedures.

    The Disposable Hemorrhoidal Staple has application for general surgical treatment of hemorrhoids.

    The Disposable Endoscopic Linear Cutter Staplers and Reloads have applications in general, gynecologic, pediatric and thoracic surgery for resection, transection, and creation of anastomoses. They may be used for transection of liver substance, hepatic vasculature and biliary structures.

    The Disposable Linear Cutter Staplers and Reloads have application in abdominal, gynecological, thoracic and pediatric surgery transection, resection, and the creation of anastomoses.

    The Disposable Curved Cutter Stapler is intended for transection, and/or creation of anastomoses. The instrument has application in multiple open or minimally invasive general (gastrointestinal and skeletal muscle), gynecologic, and thoracic surgical procedures.

    Device Description

    The proposed device, Disposable Circular Stapler is a sterilized and disposable surgical instrument intended to be used throughout the alimentary tract for the creation of end-to-side and side-to-side anastomoses in both open and laparoscopic surgeries. It places a double staggered, circular row of titanium staples upon activation, which was actieved by squeezing the handles firmly as far as they could go. Immediately after formation of the staples, the excess tissue will be resected by the circular knife, and then a circular anastomosis is created. The staple is available in two specifications which are 4.8 mm and 5.2 mm to be used per different thickness of the tissue.

    The proposed device, Disposable Linear Staplers and Reloads is a sterilized and disposable surgical instrument intended to be used in the resection of tissue for abdominal, gynecological, pediatric and thoracic surgical procedures. It places a double staggered row of titanium staples and is available in 30 mm, 45 mm, 60mm and 94 mm staple line length for use in various applications. Four staple sizes (3.8mm ,4.5 mm and 4.8 mm) are available to accommodate various tissue thicknesses. Each stapler could be reloaded no more than 7 times for total 8 firings. The Disposable Linear Staplers and reloads have two configurations: (1) DLS serials are Disposable I he Disposable Linear Business and (2) DLR serials are reloads for Disposable Linear Staplers. Each of them has various specifications.

    The proposed device, Disposable Hemorrhoidal Stapier is a sterilized and disposable surgical The "proposed" Corner and surgical treatment of hemorrhoids. It is a set of instruments that place a double staggered, circular row of titanium staples. Immediately It is a set of insultions and photo a knowlar knife blade resects the excess of compressed mucosa. The aller the formation of supres, and them. Sets are commonly used in the procedures for prolapsed and Hemorrhoids. They are also used for sets are ochanced , and circular or semicircular stapling of anorectic tissue is required.

    The proposed device, Disposable Endoscopic Linear Cutter Staplers and Reloads is a sterilized and I ile proposed unstrument, which has applications in abdominal, gynecologic, pediatic and thoracic surgery for resection, and creation of anastomoses. They may be used for thoracter bargery 200 on of liver substance, hepatic vasculature and biliary structures. It places two, triple-staggered rows of titanium staples and simultaneously divides the tissue from central line. The size of the staples is decided by selecting 3.5 mm, or 4.8 mm reload. The Endoscopic Linear Cutter can be adapted for all of the reload sizes available. The Disposable Endoscopic Linear Cutter Staplers and Reloads have two configurations: (1) DEC The Disposable Endoscopic Linear Cutter Staplers and Reloads, and (2) DER serials are sellas for Disposable Endoscopic Linear Cutter Staplers. Each of them has various specifications. Detail specifications are presented in Section 4.5 Model Description.

    The proposed device, Disposable Linear Cutter Staplers and Reloads is a sterilized and disposable surgical instrument, which has application in abdominal, gynecological, thoracic and pediatric surgery transection, resection, and the creation of anastomoses. It places two double staggered rows of titanium staples and simultaneously cut and divides tissue between the two double rows. The Disposable Linear Cutter Staplers and Reloads are available in 55 between the two 20 mm and 100 mm lengths. Reloads are available in four staple sizes to accommodate उन्द various tissue thicknesses: 3.8 mm, 4.0mm, 4.5mm and 4.8 mm. The Disposable Linear Cutter Staplers and Reloads has two configurations: (1) DLC serials is The Disposable Linear Cutter Staplers Disposable Linear Cutter Staplers with Reloads, and (2) DCR serials is Reloads for Disposable Linear Cutter Staplers. Each of them has various specifications.

    The proposed device, Disposable Curved Cutter Stapler is a sterilized and disposable surgical instrument, which is mainly suitable for low rectal resection and anastomosis of tissues where manual anastomosis is difficult. It places four staggered curved row of titanium staples on the tissue upon activation, and cut the tissue between staple lines.

    AI/ML Overview

    The provided document is a 510(k) Summary for various surgical staplers. It focuses on demonstrating substantial equivalence to predicate devices rather than providing a standalone clinical study report with detailed acceptance criteria and performance data for a new AI/ML device.

    Therefore, the requested information, particularly regarding specifics of a study proving device performance against acceptance criteria, sample sizes for test and training sets, data provenance, expert qualifications, adjudication methods, MRMC studies, or specific AI algorithm performance metrics, is not present in this document.

    This document outlines the intended use, technological characteristics, and comparison to predicate devices for a suite of surgical staplers. The "Non-Clinical Test Conclusion" section {4} states that tests were conducted to verify the device met all design specifications and was substantially equivalent to the predicate device. These tests primarily cover physical performance, material properties, and packaging integrity.

    Here's a table summarizing the acceptance criteria and reported device performance for one of the devices (Disposable Circular Stapler) as presented in the document:

    Table of Acceptance Criteria and Reported Device Performance (Disposable Circular Stapler)

    Acceptance Criteria (Item)UnitAcceptance Range / Predicate Value (Implicitly the criterion)Reported Device Performance (Proposed Device)
    Product Code-GDWGDW
    Regulation No.-21 CFR 878.475021 CFR 878.4750
    Class-IIII
    Intended Use-Creation of end-to-end, end-to-side and side-to-side anastomoses in both open and laparoscopic surgeries.Same
    Operation Principle-ManualManual
    Cutting Mechanism-Circular KnifeCircular Knife
    Safety Mechanism-Indicator for appropriate range of desired closed staple height. Safety Release for preventing from mis-firing.Same
    Outsider DiametermmSimilar to predicate21, 23, 25, 27, 29, 31, 33 mm
    Cutting DiametermmSimilar to predicate12.4, 15, 17, 18.5, 20, 22.2, 24 mm
    Number of Staples-Similar to predicate18, 20, 22, 24, 28, 30
    Closed Staples HeightmmSimilar to predicate2.0, 2.2 mm
    Closed Staples Form-Two staples connected at the topImage: Two staples connected at the top
    Hardness (Circular knife)HV0.2≥ 380≥ 380HV0.2
    Pressure Resistance after SuturingkPa≥ 3.6≥ 3.6kPa
    Staple Material-Unalloyed Titanium conforms to ASTM F 67-06Unalloyed Titanium conforms to ASTM F 67-06
    Stapler Materials-Stainless Steel, PolycarbonateStainless Steel, Polycarbonate
    Sterilization-Similar to predicateIrradiation Sterilized, SAL: 10-6
    Endotoxin LimitEU per Product2020 EU per Product
    Package-Tray with Tyvek PaperTray with Tyvek Paper
    Labeling-Conforms to 21 CFR part 801Conforms to 21 CFR part 801

    Regarding the other requested information:

    1. Sample size used for the test set and the data provenance: This document does not specify sample sizes for specific tests or the provenance of any "data provenance" (e.g., country of origin, retrospective/prospective clinical data) as it's a submission for a medical device (surgical staplers) based on non-clinical performance and substantial equivalence to existing predicate devices, not an AI/ML diagnostic or prognostic tool. The tests mentioned are "Performance Testing, including Physical Specification, Closed Staple Height Dimensions, Pressure Perforrnality Tisting, Mercums Tensile Strength of Staple Line Repair and Force Required to Fire Stapler" {4}.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for surgical staplers like this would typically be established through engineering specifications, material standards, and performance testing against those standards, not expert consensus on interpretations of data.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, an MRMC study was not done as this is not an AI/ML-driven diagnostic device.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is a manual surgical instrument.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for this type of device would be its adherence to established engineering specifications, material standards (e.g., ASTM F 67-06 for titanium), and performance thresholds (e.g., pressure resistance, hardness) determined through non-clinical testing.
    7. The sample size for the training set: Not applicable, as this is not an AI/ML device requiring a training set.
    8. How the ground truth for the training set was established: Not applicable.

    In summary, this 510(k) submission demonstrates substantial equivalence of a physical medical device (surgical staplers) to previously cleared devices based on non-clinical performance testing and design specifications, rather than clinical studies or AI/ML algorithm evaluations which would require the detailed information requested.