Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K130860

    Validate with FDA (Live)

    Date Cleared
    2013-04-25

    (28 days)

    Product Code
    Regulation Number
    864.7470
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bio-Rad VARIANT™ II Hemoglobin A Ic Program is intended for the percent determination of hemoglobin A Ic in human whole blood using ion-exchange high-performance liquid chromatography (HPLC). The Bio-Rad VARIANT II Hemoglobin Alc Program is intended for Professional Use Only. For in vitro diagnostic use. Measurement of percent hemoglobin A1c is effective in monitoring long-term glucose control in individuals with diabetes mellitus.

    The VARIANT™ II ß-thalassemia Short Program is intended for the separation and area percent determinations of hemoglobins A2 and F, and as an aid in the identification of abnormal hemoglobins in whole blood using ion-exchange high-performance liquid chromatography (HPLC). The Bio-Rad VARIANT II ß-thalassemia Short Program is intended for use only with the Bio-Rad VARIANT II Hemoglobin Testing System. For in vitro diagnostic use. Measurement of the percent hemoglobin A2 and F are effective in screening of ß-thalassemia (i.e., hereditary hemolytic anemias characterized by decreased synthesis or more types of abnormal hemoglobin polypeptide chains).

    Device Description

    The VARIANT II Hemoglobin Testing System is a fully automated, highthroughput hemoglobin analyzer. The VARIANT II Hemoglobin Testing System provides an integrated method for sample preparation, separation and determination of the relative percent of specific hemoglobin in whole blood. It consists of two modules - the VARIANT II Chromatographic Station (VCS) and the VARIANT II Sampling Station (VSS). In addition, a personal computer is used to control the VARIANT System using Clinical Data Management (CDM) Software.

    A personal computer (PC) is used to control the VARIANT II Hemoglobin Testing System using Clinical Data Management (CDM™) software. The CDM software supports import of sample information from and export of patient results to a Laboratory Information System (LIS). Control results are displayed on Levy-Jennings Charts and are exportable to Unity Real Time™.

    AI/ML Overview

    This K130860 submission is a Special 510(k) for a device modification, meaning the changes are to existing predicate devices (VARIANT II Hemoglobin A1c Program and VARIANT II ß-thalassemia Short Program) and aim to demonstrate substantial equivalence without impacting the core performance specifications, intended use, or operating principles. The modifications primarily involve software and firmware updates, along with a PC Board replacement.

    Therefore, the study focuses on verification and validation (V&V) testing to ensure the modified device remains safe, effective, and substantially equivalent to its predicate. It does not present new performance data against specific acceptance criteria for diagnostic accuracy as would be expected for a novel device or a device with significant performance changes. Instead, it asserts that the changes do not affect the previously established performance claims.

    Here's an analysis based on the provided text, focusing on how the device meets acceptance criteria related to its modifications:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a device modification submission where the performance specifications are stated to be unchanged from the predicate, the acceptance criteria are implicitly that the modified device's performance is at least as good as the predicate device and that the modifications do not introduce new risks or degrade existing performance. The "performance" reported is the outcome of the verification/validation and risk management processes.

    Acceptance Criteria (Implicit for Device Modification)Reported Device Performance (as stated in the submission)
    No change to performance specifications"When compared to the predicate device, there are no changes to the performance specifications, intended or indications for use, or operating principles.""No change or impact, claims transferred from predicate device." (for both programs)
    No adverse impact on product safety and effectiveness"Risk Analysis and Verification/Validation testing results demonstrate that the changes do not affect product safety, effectiveness, and substantial equivalency claims.""Design verification/validation tests met established acceptance criteria.""deemed the modified product safe, effective, and comparable to the predicate device."
    Modifications developed under design controls"In addition, these changes were designed, developed and implemented under established design control and GMP processes..."
    Compliance with risk management for modifications"In accordance with ISO 14971, and internal risk management processes and procedures a defined risk analysis was used to identify, mitigate, or eliminate potential risks associated with the device modifications.""The risk evaluation for the device software and firmware modifications included the following tasks..."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a numerical sample size for a "test set" in the traditional sense of a clinical or analytical performance study. Given the nature of a Special 510(k) for software/firmware/hardware changes, the "test set" would refer to the data and scenarios used during verification and validation (V&V) testing.

    • Sample Size: Not explicitly stated as a number of patient samples. The V&V efforts would likely involve testing various functionalities, defect fixes using specific test cases, and potentially a range of instrument data (already available or specifically generated for V&V).
    • Data Provenance: Not explicitly stated. For "defect corrections," the data likely originated from "customer feedback" and scenarios that caused the identified defects. For general V&V, it would involve internal testing data. It's implied to be retrospective, as it addresses "customer feedback" and "defects" from prior versions, but specific details are absent.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable or not provided for this submission. The ground truth in this context is typically related to diagnostic accuracy, which is not being re-evaluated because performance claims are "transferred from predicate device."

    For defect fixes, the "ground truth" would be whether the defect is successfully resolved and the intended functionality works as designed. This is assessed by engineering and quality assurance teams during V&V. The document mentions "a trained risk assessment team" for FMEA, but not "experts" establishing a diagnostic ground truth for a test set.

    4. Adjudication Method for the Test Set

    Not applicable/not provided. No "adjudication method" for interpreting results from a test set is mentioned because the submission directly states that performance specifications and claims are unchanged and transferred from the predicate. The "ground truth" for V&V testing of software/firmware changes is based on successful execution of tests and resolution of identified bugs, not on expert consensus interpretation of diagnostic output.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an automated in vitro diagnostic (HPLC system) for measuring specific hemoglobin levels, not an AI-assisted diagnostic imaging or interpretation tool that involves human readers. Therefore, an MRMC study is not relevant to this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    The device itself (VARIANT II Hemoglobin Testing System with CDM Software) operates as a standalone automated system for measurement. The changes are to its software/firmware. The V&V testing would assess the functionality of this automated system in its modified state. So, the testing effectively evaluates the "standalone" performance of the modified system, but it's not a new standalone study; it's a re-validation of the existing standalone system after modifications. The performance claims are asserted to be the same as the predicate (which was a standalone device).

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the original predicate devices, the ground truth for establishing performance (e.g., accuracy of HbA1c or HbA2/F measurements) would have been based on comparison to reference methods or clinical outcomes.

    For this specific submission (device modification):

    • For defect corrections: The "ground truth" is the successful elimination of the defect and the proper functioning of the software features (e.g., CDM not crashing, calibrator reassignment working). This is validated through internal software testing.
    • For performance: The submission directly states "No change or impact, claims transferred from predicate device." This means the ground truth for performance measures (precision, accuracy, linearity, etc.) was established during the original predicate device's clearance and is implicitly inherited rather than re-established in detail for this modification. The V&V here confirmed that the modifications did not degrade the ability to achieve those previously established performance characteristics.

    8. The Sample Size for the Training Set

    Not applicable/not provided. This device is not described as involving a machine learning algorithm that requires a "training set." The software and firmware updates are for controlling the HPLC system and managing data, not for learning from data in the way an AI algorithm would.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As no training set for a machine learning algorithm is involved, this question is not relevant to the submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1