Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K130266

    Validate with FDA (Live)

    Date Cleared
    2013-08-29

    (206 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EndoWrist® One Vessel Sealer is a bipolar electrosurgical instrument for use with the da Vinci Si Surgical System and the ERBE VIO 300 D electrosurgical generator. It is intended for grasping and blunt dissection of tissue and for bipolar coagulation and mechanical transection of vessels up to 7 mm in diameter and tissue bundles that fit in the jaws of the instrument. The EndoWrist One Vessel Sealer has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

    Device Description

    The EndoWrist One Vessel Sealer is an electrosurgical sealing and cutting instrument to be used in conjunction with the da Vinci Si Surgical System and the ERBE VIO 300 D electrosurgical generator. The outside diameter (O.D.) of the instrument shaft is 8.5 mm and the working length is 38 cm. The instrument is provided sterile and is a single-use, disposable device. The distal end has jaws with bipolar electrodes for sealing vessels and contains a cutting blade that extends through the jaws to transect sealed vessels and other tissues. The sealing and cutting functions are actuated by the daVinci Si Surgical System foot pedals. The ERBE VIO 300 D generator provides the high frequency (radio frequency) electrical current for tissue sealing.

    AI/ML Overview

    Here's an analysis based on the provided 510(k) summary:

    This 510(k) summary for the Intuitive Surgical EndoWrist® One Vessel Sealer is for clearance of a device that is stated to be identical in technological characteristics to a previously cleared predicate device (K110639). It also states that the proposed indications for use are a subset of another predicate device (K071728). In such cases, extensive new performance data may not be required beyond demonstrating equivalence and that the design outputs meet the design input requirements.

    Therefore, the information regarding in-depth clinical studies, ground truth establishment, expert consensus, and sample sizes for AI/algorithm performance is not present in this 510(k) summary, as it is not a submission for an AI/ML powered device. Instead, the focus is on a traditional surgical instrument.


    Acceptance Criteria and Device Performance

    The provided document does not explicitly state numerical acceptance criteria in a table format, nor does it provide specific reported device performance metrics in the way one would for an AI/ML algorithm. Instead, it relies on demonstrating substantial equivalence to predicate devices.

    The "acceptance criteria" can be inferred as the demonstration that the new device's design outputs meet its design input requirements and that its performance is equivalent to the predicate devices, not raising new questions of safety or effectiveness.

    Acceptance Criteria (Inferred from context)Reported Device Performance
    Design output meets design input requirements."Design validation... demonstrates that the subject device... meets the design input requirements."
    Substantial equivalence to predicate device (EndoWrist® One Vessel Sealer K110639)."Design validation and animal testing demonstrates that the subject device is substantially equivalent to the predicate devices."
    Not raising different questions of safety or effectiveness compared to predicates."The differences do not raise different questions of safety or effectiveness as compared to the predicate devices."
    Effective for bipolar coagulation and mechanical transection of vessels up to 7 mm.Not explicitly quantified in a performance table, but implied by the intended use and equivalence to predicates cleared for this purpose.
    Safe for use with da Vinci Si Surgical System and ERBE VIO 300 D generator.Implied by design validation and substantial equivalence.

    Study Details (Based on available information)

    1. Sample size used for the test set and the data provenance:

      • The document mentions "Design validation and animal testing." It does not specify the sample size for this testing or the provenance of the data in terms of country of origin or whether it was retrospective or prospective. Animal testing is typically prospective.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable/Not provided. The "ground truth" for a traditional surgical instrument would relate to its physical performance (e.g., ability to seal vessels, cut tissue, withstand stress), which is typically evaluated by engineers and veterinarians/surgeons in animal studies, not by expert interpretation of images or other data for establishing a medical diagnosis "ground truth."
    3. Adjudication method for the test set:

      • Not applicable/Not provided. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving human interpretation, not for evaluating the mechanical and electrosurgical performance of an instrument in design validation or animal testing.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is not an AI-powered device, and therefore, an MRMC comparative effectiveness study involving human readers and AI assistance was not conducted or reported.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No. This is not an AI-powered device.
    6. The type of ground truth used:

      • For this type of device, the "ground truth" would be established through a combination of engineering specifications, empirical measurements (e.g., burst pressure of sealed vessels, tissue cutting efficacy), and observations in animal models regarding the device's ability to perform its intended functions without adverse events. It would not be expert consensus, pathology, or outcomes data in the context of diagnostic accuracy.
    7. The sample size for the training set:

      • Not applicable/Not provided. This is not an AI/ML device that requires a "training set."
    8. How the ground truth for the training set was established:

      • Not applicable/Not provided. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1