Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K130153

    Validate with FDA (Live)

    Date Cleared
    2013-03-12

    (49 days)

    Product Code
    Regulation Number
    870.5310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation.

    The device is for use by qualified medical personnel trained in the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.

    AED Therapy: To be used in the presence of a suspected cardiac arrest on patients of at least 8 years of age that are unresponsive, not breathing and pulseless.

    Manual Defibrillation: Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients that are pulseless and unresponsive. Synchronous defibrillation is indicated for the termination of certain atrial and ventricular arrhythmias.

    Noninvasive External Pacing Therapy: The pacing option is intended for treating patients with symptomatic bradycardia. It can also be helpful in patients with asystole, if performed early.

    Pulse Oximetry: The SpO2 option is intended for use when it is beneficial to assess the patient's oxygen saturation level.

    Noninvasive Blood Pressure Monitoring: The NBP option is intended for noninvasive measurement of a patient's arterial blood pressure.

    End-Tidal CO2: The EtCO2 option is intended for noninvasive monitoring of a patient's exhaled carbon dioxide and to provide a respiration rate.

    12-Lead ECG: The 12-Lead ECG function is to provide a conventional diagnostic 12-Lead-ECG report, which may include measurement and interpretative statements.

    Q-CPR: The Q-CPR option provides feedback designed to encourage rescuers to perform resuscitation in accordance with AHA/ERC guidelines for chest compression rate depth, and duty cycle and ventilation rate, volume and flow rate (inflation time).

    Invasive Pressures: The Invasive Pressures option is indicated for measuring arterial, venous, intracranial, and other physiological pressures on patients.

    Temperature: The Temperature option is indicated for measure temperature in patients.

    Device Description

    The Philips HeartStart MRx with Airway Confirmation Assist (ACA) Defibrillator/Monitor is a modification of the FDA cleared HeartStart MRx Defibrillator/Monitor. This function of the MRx used to verify airway placement and aid in assessing ventilatory status when an adult or pediatric patient is being ventilated using an endotracheal or supraglottic airway and the airway is being monitored using capnography.

    Based on predetermined criteria, Airway Confirmation Assist continuously analyzes the CO2 waveform coming from the capnography monitor to decide if the waveform represents a valid breath. If the Airway Confirmation Assist is enabled in the MRx, and based on characteristics of the valid breaths, ACA will decide if the airway is established. It will then report a Pass/Fail decision to the MRx. The defibrillator/monitor will display the result to the user in the configured format (graphic and text). As an adjunct to clinical examination, ACA provides additional information along with capnography to help in verifying airway placement and aiding in assessment of ventilation status.

    AI/ML Overview

    This 510(k) summary (K130153) describes the Philips HeartStart MRx with Airway Confirmation Assist (ACA) Defibrillator/Monitor, a modification of the previously cleared HeartStart MRx. The ACA function is designed to verify airway placement and aid in assessing ventilatory status by continuously analyzing the CO2 waveform from a capnography monitor.

    It is important to note that the provided document does not contain a detailed study report with specific acceptance criteria, reported performance, or information regarding sample sizes, ground truth establishment, or expert involvement for the Airway Confirmation Assist feature. The document primarily focuses on demonstrating substantial equivalence to predicate devices and detailing the general intended uses of the overall HeartStart MRx device and its various options beyond the ACA.

    However, based on the available information, we can infer some aspects and highlight what is missing.


    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary does NOT explicitly state an acceptance criterion or specific reported performance metrics for the Airway Confirmation Assist (ACA) feature. It generally states:

    "Based on predetermined criteria, Airway Confirmation Assist continuously analyzes the CO2 waveform coming from the capnography monitor to decide if the waveform represents a valid breath. If the Airway Confirmation Assist is enabled in the MRx, and based on characteristics of the valid breaths, ACA will decide if the airway is established. It will then report a Pass/Fail decision to the MRx."

    And:

    "...it has been determined that the HeartStart MRx with Advanced Airway Confirmation Defibrillator/Monitor is suitable for its intended use."

    Without a specific study report for the ACA feature, a table of acceptance criteria and reported device performance cannot be accurately generated from this document. If this were a full study report, we would expect to see metrics like sensitivity, specificity, accuracy, or a pass/fail rate against a specified threshold for correct airway placement detection.


    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide information on the sample size used for testing the Airway Confirmation Assist (ACA) feature, nor does it specify the data provenance (e.g., country of origin, retrospective or prospective). The summary mentions "performance and usability testing" but gives no details about these tests for the ACA.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not specify the number or qualifications of experts used to establish ground truth for the Airway Confirmation Assist (ACA) feature's test set. For a device assessing airway placement, one would typically expect highly qualified medical professionals (e.g., intensivists, emergency physicians, anesthesi paramedics with specific training in intubation and capnography) to serve as the ground truth.


    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method (e.g., 2+1, 3+1, none) for the test set of the Airway Confirmation Assist (ACA) feature.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study for the Airway Confirmation Assist (ACA) feature. It does not provide an effect size of how much human readers (medical personnel) improve with AI vs. without AI assistance. The ACA is described as providing "additional information along with capnography to help in verifying airway placement and aiding in assessment of ventilation status," suggesting it's an adjunct rather than a complete replacement for human assessment, but no specific study comparing human performance with and without ACA is detailed.


    6. Standalone Performance Study

    While the device's ACA function "continuously analyzes the CO2 waveform... to decide if the waveform represents a valid breath" and "decide if the airway is established," the document does not explicitly describe a standalone (algorithm only without human-in-the-loop performance) study with specific metrics for the ACA feature. The description points to an algorithmic function ("Based on predetermined criteria," "continuously analyzes... to decide if the waveform represents a valid breath," "ACA will decide if the airway is established"), but performance data from such a study is absent.


    7. Type of Ground Truth Used

    The document does not specify the type of ground truth used for evaluating the Airway Confirmation Assist (ACA) feature. For airway placement, ground truth could involve direct visualization by an expert, fiberoptic bronchosopy, chest X-ray, or clinical assessment combined with continuous capnography waveform analysis by experienced clinicians.


    8. Sample Size for the Training Set

    The document does not provide information on the sample size used for the training set for the Airway Confirmation Assist (ACA) feature.


    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established for the Airway Confirmation Assist (ACA) feature.


    In conclusion, this 510(k) summary primarily focuses on regulatory approval based on substantial equivalence and general safety/performance. It lacks the detailed study information typically found in a clinical validation report for a specific algorithmic feature like the Airway Confirmation Assist. To answer most of these questions thoroughly, a more in-depth technical or clinical study report for the ACA feature would be required. The document indicates that "HeartStart MRx with Airway Confirmation Assist Defibrillator/Monitor has been subjected to performance and usability testing," but the specifics of this testing for the ACA component are not detailed here.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1