Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K123980

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-03-05

    (69 days)

    Product Code
    Regulation Number
    878.4410
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Misonix SonicOne Ultrasonic Would Care System and Accessories are indicated for use in the debridement of wounds, such as, but not limited to, burn wounds, diabetic ulcers. bedsores and vaginal ulcers. soft tissue debridement and cleansing of the surgical site in applications in which, in the physician's judgment would require the use of an ultrasonic aspirator with sharp debridement.

    Device Description

    The SonicOne Plus Ultrasonic Wound Care System and Accessories is comprised of a generator, which feeds a 22.5 kHz electrical signal to a piezoelectric crystals mounted in a hand-held handpiece; the crystals then vibrate at the same frequency. The titanium tip attached to the handpiece amplifies the vibration. An irrigation unit is provided to introduce irrigation solution to the operative site. Accessories include Probe tips, Wrenches, Sterile and non sterile Tube sets and sterile Surgical Procedure bags and sterile Handpiece Sheaths.

    AI/ML Overview

    This document describes the Misonix SonicOne Plus Ultrasonic Wound Care System and Accessories, which is an ultrasonic surgical system used for wound debridement. The submission aims to demonstrate substantial equivalence to predicate devices, namely the SonicOne Ultrasonic Wound Care System (K112782) and the AUSS-7 Ultrasonic Surgical System (K070313).

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are based on compliance with a set of voluntary standards and successful completion of non-clinical tests. The "reported device performance" is essentially the statement that the device has "passed" these standards and tests. There are no specific quantitative performance metrics like sensitivity, specificity, or accuracy provided, as this is a medical device for physical intervention, not diagnostic imaging or AI.

    Acceptance Criteria (Voluntary Standards & Non-Clinical Tests)Reported Device Performance
    UL 60601-1 2nd Edition (Medical Electrical Equipment, Part 1: General Requirements for Safety)Passed
    IEC 60601-1* 2nd Edition (Medical Electrical Equipment, Part 1: General Requirements for Safety)Passed
    ISO 10993-1:2009 (Biological evaluation of medical devices —Part 1: Evaluation and testing)Passed
    ISO 10993-7:2008 (Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals)Passed
    ISO 14971:2007 (Medical devices -- Application of risk management to medical devices)Passed
    ISO 15223-1:2007/A1:2008 (Medical devices—Symbols to be used with medical device labels, labeling, and information to be supplied—Part 1: General requirements)Passed
    ISO 11607:2006 (Packaging for Terminally Sterilized Medical Devices)Passed
    ISO 11135:2007 (Sterilization of health care products -- Ethylene oxide -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices)Passed
    IEC 62304: 2006 (Medical Device Software-Software Life Cycle Processes)Passed
    IEC 60601-1-4 (General Requirements for safety-Programmable electrical medical systems)Passed
    FCC Part 18 (EMC Requirements)Passed
    Output Frequency MeasurementsPerformed
    Output Power Measurements (No Load to Maximum Load)Performed
    Tip Displacement MeasurementsPerformed
    Irrigation Flowrate MeasurementsPerformed
    Life TestsPerformed
    Input Power MeasurementsPerformed
    EMI TestsPerformed
    Dielectric Tests on Mains CircuitsPerformed
    Patient Current Leakage and Patient Sink Current MeasurementsPerformed
    Power Line Ground Leakage MeasurementsPerformed
    Dielectric Tests on Patient CircuitsPerformed
    Sterilization Validations (ANSI/AAMI/ISO 11135:1994 or latest and ISO 11135:2007)Validated
    Reprocessing (cleaning/disinfecting/sterilization) instructions validationValidated
    Software Validation (cleared under FDA 510k #K070313, meets FDA guideline "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and ISO 60601-1-4)Cleared and Meets Requirements

    2. Sample Size Used for the Test Set and the Data Provenance

    The document explicitly states: "As such, no additional clinical data was obtained in anticipation of this submission." This indicates that there was no dedicated test set involving human subjects or clinical data for this specific 510(k) submission. The safety and performance assessment relies on non-clinical tests, engineering validations, and substantial equivalence to previously cleared predicate devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. There was no clinical test set requiring ground truth establishment by experts for this submission. The device's safety and effectiveness are established through engineering and biological testing against established standards.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    Not applicable. There was no clinical test set requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If so, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This device is an ultrasonic surgical system, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC comparative effectiveness study regarding "human readers improve with AI" is irrelevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is established by engineering standards and biological safety standards (e.g., ISO, IEC, UL, FCC for electrical safety, biocompatibility, sterilization, and electromagnetic compatibility). For the software component, the ground truth is its compliance with FDA guidelines and ISO 60601-1-4, as validated by previous clearance (K070313).

    8. The Sample Size for the Training Set

    Not applicable. The device is not an AI algorithm that requires a training set in the typical sense. Its design and operational parameters are based on established engineering principles and its predicate devices.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. No training set for an AI algorithm was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1