    Intended Use

    The aap Radius Set 2.5 is indicated for the fixation of intra- and extra-articular fractures and osteotomies of the distal radius.

    Device Description

    Bone plates and screws are used for fixation of bone fragments, i.e., for treatment of bone fractures and other bone injuries. Bone plates are fixed by the use of bone screws. Bone plates and bone screws are implants. If the plates are used in conjunction with locking screws, a so called internal fixator will be realized (internal fixation).
    The aap Radius Set 2.5 consists of:
    Extra-art. Radius Plate WSP 2.5, (left and right version)
    Juxta-art. Radius Plate WSP 2.5, (left and right version)
    L-Plate WSP 2.5, (left and right version)
    Straight Plate WSP 2.5
    T-Plate WSP 2.5
    Cortical Screw WS 2.5, self-tapping (locking bone screw)
    Cortical Screw 2.5, self-tapping
    Set of Instruments, Radius Set 2.5

    AI/ML Overview

    The aap Radius Set 2.5 is a device indicated for the fixation of intra- and extra-articular fractures and osteotomies of the distal radius.

    Here's an analysis of the provided information regarding its acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission primarily focuses on mechanical performance to establish substantial equivalence.

    Acceptance CriteriaReported Device Performance
    Fatigue implant tests with progressive loadings: Represent worst-case clinical scenario.Substantial equivalence with respect to the mechanical performance of the aap system could be stated due to the test results gained. The subject device is safe and effective, and its performance meets the requirements of its pre-defined acceptance criteria and intended uses.
    Mechanical tests of Screws acc. to ASTM F543-07: Fulfil relevant requirements.The Screws fulfil the relevant requirements of ASTM F543-07 and pre-defined acceptance criteria and intended uses.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided summary does not explicitly state the specific sample sizes for the mechanical tests (e.g., number of plates, screws, or test repetitions). It only mentions "fatigue implant tests with progressive loadings" and "Mechanical tests of Screws acc. to ASTM F543-07."

    The data provenance is non-clinical testing, performed by the manufacturer, aap Implantate AG, in Germany. The study is prospective in the sense that these tests were conducted specifically for this submission to demonstrate substantial equivalence, rather than being a retrospective analysis of existing data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable as the evaluation relies on non-clinical mechanical testing based on established engineering standards (ASTM F543-07), not on expert consensus for clinical image interpretation or diagnosis. The "ground truth" here is adherence to specified mechanical properties.

    4. Adjudication Method for the Test Set

    This information is not applicable as the evaluation is based on objective mechanical testing results against established standards, not on human adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This type of study is typical for diagnostic or screening devices where human interpretation is involved. For a bone fixation device, the primary evaluation is mechanical performance.

    6. Standalone Performance Study

    Yes, a standalone performance study was implicitly done. The "Summary of Non-clinical tests" describes mechanical tests performed on the device components (implants and screws) to demonstrate their effectiveness and safety against pre-defined acceptance criteria and established standards (ASTM F543-07). This is a standalone evaluation of the device's physical properties and performance.

    7. Type of Ground Truth Used

    The type of ground truth used is engineering specifications and established mechanical testing standards. Specifically:

    • Pre-defined acceptance criteria for fatigue implant tests.
    • ASTM F543-07 for mechanical tests of screws.

    8. Sample Size for the Training Set

    This information is not applicable. The device is a physical medical implant, not an AI or machine learning algorithm. Therefore, there is no "training set" in the context of data-driven model development.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this type of device.