510(k) Data Aggregation

    K Number
    K123656

    Validate with FDA (Live)

    Date Cleared
    2013-02-25

    (89 days)

    Product Code
    Regulation Number
    888.3050
    Age Range
    2 - 120
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When intended as an adjunct to fusion of the occipitocervical spine, cervical spine, and the thoracic spine (Occiput-T3), the VERTEX® Reconstruction System is indicated for skeletally mature patients using allograft and/or autograft for the following:

    DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion, and/or tumors.

    Occipitocervical Components: Plate Rod/Plates/Rods/Occipital Screws/Hooks

    The occipitocervical plate rods, plates, rods, occipital screws, and hooks are intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the occipitocervical junction and the cervical spine. When used to treat these occipitocervical and cervical conditions, these screws are limited to occipital fixation only. The screws are not intended to be placed in the cervical spine.

    Occipitocervical constructs require bilateral fixation to C2 and below. Note: Segmental fixation is recommended for these constructs.

    Hooks and Rods

    The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine.

    Multi-axial Screws

    The use of multi-axial screws is limited to placement in T1-T3. The screws are not intended to be placed in the cervical spine.

    Titanium ATLAS® Cable System to be used with the VERTEX® Reconstruction System allows for cable attachment to the posterior cervical or thoracic spine.

    Connectors

    In order to achieve additional levels of fixation, the VERTEX® Reconstruction System may be connected to the CD HORIZON® Spinal System rods with the VERTEX® rod connectors. Transition rods with differing diameters may also be used to connect the VERTEX® Reconstruction System to the CD HORIZON® Spinal System screws, hooks, and connectors. Refer to the CD HORIZON® Spinal System package insert for a list of the CD HORIZON® Spinal System indications of use.

    Device Description

    The VERTEX® Reconstruction System is a posterior svstem, which consists of a variety of shapes and sizes of plates, rods, hooks, screws, multi-axial screws, and connecting components, which can be rigidly locked to the rod in a variety of configurations, with each construct being tailor-made for the individual case. Titanium ATLAS® cable may be used with this system at the surgeon's discretion. See the package inserts of both of those systems for labeling limitations.

    The VERTEX® Reconstruction System is fabricated from medical grade titanium, medical grade titanium alloy, and medical grade cobalt chromium. Medical grade titanium, medical grade titanium alloy, and/or medical grade cobalt chromium may be used together. Never use titanium, titanium alloy, and/or cobalt chromium with stainless steel in the same construct. The VERTEX® Reconstruction System includes a retaining ring for the multi-axial screw made of Shape Memory Alloy (Nitinol - NiTi), Shape Memory Alloy is compatible with titanium, titanium alloy, and cobalt chromium implants only. Some components contain elastomeric stakes made of silicone adhesive commonly used in implantable medical devices. Do not use with stainless steel.

    To achieve best results, do not use any of the VERTEX® Reconstruction System implant components with components from any other system or manufacturer unless specifically labeled to do so in this or another MEDTRONIC document. As with all orthopaedic and neurosuraical implants, none of the VERTEX® Reconstruction System components should ever be reused under any circumstances.

    AI/ML Overview

    The provided document describes a 510(k) summary for the VERTEX® Reconstruction System, which is a spinal interlaminal fixation orthosis. The purpose of this 510(k) is to add additional components to the existing system. The document does not describe a study involving device performance metrics such as sensitivity, specificity, accuracy, or any form of clinical trial with human subjects where these metrics would be relevant.

    Instead, this submission focuses on demonstrating substantial equivalence to previously cleared predicate devices through non-clinical testing (mechanical testing) and by showing that the new components are identical in terms of material, indications for use, and intended use, and that performance specifications are met. Therefore, many of the requested elements about clinical study design (sample size for test set, data provenance, expert ground truth, adjudication, MRMC studies, standalone performance, training set details) are not applicable to this type of submission.

    Here's a breakdown of the available information:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the new components are based on demonstrating mechanical equivalence to existing predicate devices under specific ASTM standards.

    Acceptance Criteria CategorySpecific Criteria/TestReported Device Performance/Conclusion
    Mechanical PerformanceCompression Fatigue (ASTM 1717-12)Performed on subject tapered rods. The submission concludes that the subject devices demonstrated substantial equivalence to the previously listed predicate devices based on the results of the risk analysis and supporting documentation.
    Static Compression (ASTM 1717-12)Performed on subject tapered rods. Results contribute to the conclusion of substantial equivalence.
    Static Torsion (ASTM 1717-12)Performed on subject tapered rods. Results contribute to the conclusion of substantial equivalence.
    Design DifferencesNew components (lengths, diameters, overall design profile) do not present a worst-case scenario compared to predicates with respect to mechanical performance.Mechanical testing was not required for pre-cut/pre-bent rods, threaded rods, dominoes, crosslinks, or plates, because the differences were limited to additional lengths and diameters, which were determined not to present a worst-case scenario. This implies these design variations met the inherent safety/performance of the predicates.
    Material EquivalenceIdentical medical grade materials as predicate devices.The document states, "The subject and predicate parts are identical in terms of material..." The system is fabricated from medical grade titanium, medical grade titanium alloy, and medical grade cobalt chromium.
    Indications for UseIdentical indications for use as predicate devices.The document states, "...indications for use... are substantially equivalent to predicate VERTEX® Reconstruction System devices..." and "The subject and predicate parts are identical in terms of material, indications for use, intended use..."
    Intended UseIdentical intended use as predicate devices.The document states, "The subject and predicate parts are identical in terms of material, indications for use, intended use..."
    LabelingIdentical labeling as cleared predicate.The labeling is identical to that cleared in K121191 (S.E. 06/29/12).

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document does not specify the exact number of test samples (e.g., number of tapered rods) used for the mechanical tests. However, ASTM standards typically outline the minimum number of samples required for such testing.
    • Data Provenance: The tests were performed by Medtronic, likely in a laboratory setting, in accordance with ASTM (American Society for Testing and Materials) standards. This is a non-clinical, in-vitro study. Country of origin for data is not explicitly stated but Medtronic Sofamor Danek is based in Memphis, TN, USA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This submission is based on mechanical performance testing against established engineering standards (ASTM 1717-12) and comparison to predicate devices, not on clinical outcomes or expert interpretation of patient data. Therefore, no "experts" in the sense of clinical reviewers or diagnosticians were used to establish ground truth for a test set.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not Applicable. As this is non-clinical mechanical testing, there is no need for adjudication as typically understood in clinical or imaging studies. Mechanical test results are objective measurements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not a study of AI or a diagnostic device involving human readers or cases. It is a 510(k) for a spinal implant system based on substantial equivalence through mechanical testing and comparison to predicate devices.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an algorithm or AI device. It's a medical implant system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For the mechanical testing, the "ground truth" is defined by the performance specifications and criteria outlined in the ASTM 1717-12 standard for spinal implant assemblies. The device's performance is compared against these engineering benchmarks. For demonstrating substantial equivalence for other components, the "ground truth" is the established safety and performance of the predicate devices.

    8. The sample size for the training set

    • Not Applicable. There is no "training set" in the context of this 510(k) submission, as it does not involve machine learning or AI models.

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set, there is no ground truth established for it.