Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K123343

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-07-11

    (253 days)

    Product Code
    Regulation Number
    862.1678
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QMS® Tacrolimus Immunoassay is intended for the quantitative determination of tacrolimus in human whole blood on automated clinical chemistry analyzers. The results obtained are used as an aid in the management of kidney, liver, and heart transplant patients receiving tacrolimus therapy. This in vitro diagnostic device is intended for clinical laboratory use only.

    The QMS® Tacrolimus Calibrator set is intended for use in calibration of the QMS® Tacrolimus Immunoassay.

    Device Description

    The QMS® Tacrolimus Immunoassay consists of separately packaged reagents (Reagent 1, Reagent 2 and Extraction Reagent) and calibrators (Calibrator A, B, C, D, E, and F).

    Reagent 1 (Antibody Reagent): <1.0% Anti-Tacrolimus monoclonal antibody (rabbit) in a buffer as stabilizer and <0.09% sodium azide as preservative. Configuration: 1 x 18 mL.
    Reagent 2 (Microparticle Reagent): <0.3% Tacrolimus-coated microparticles in buffer containing <0.09% sodium azide as preservative. Configuration: 1 x 12 mL.
    Extraction Reagent: 300 mM Zinc Sulfate and <0.09% sodium azide as preservative. Configuration: 1 x 50 mL.

    QMS® Tacrolimus Reagents are provided ready-to-use in liquid form and are to be stored at 2 to 8°C until the expiration date on the label.

    Calibrator Level, Target Concentration (ng/mL), Configuration:
    Calibrator A, 0.0, 1 x 4 mL
    Calibrator B, 2.0, 1 x 2 mL
    Calibrator C, 5.0, 1 x 2 mL
    Calibrator D, 10.0, 1 x 2 mL
    Calibrator E, 20.0, 1 x 2 mL
    Calibrator F, 30.0, 1 x 2 mL

    The QMS® Tacrolimus Calibrator kit is provided separately and packaged in amber glass bottles in a rectangular cardboard box with a 6-bottle divider. The calibrators are provided in liquid form and are to be stored at -20℃ ± 5℃ until the expiration date on the label. They are ready-to-use upon thawing.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the QMS® Tacrolimus Immunoassay, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Implied/Contextual)Reported Device Performance
    Functional Sensitivity (LOQ)Lowest concentration with ≤20% inter-assay CV over extended period0.9 ng/mL
    Precision (Total-Run)≤7.3% CV≤7.3% CV (at various levels across the assay range)
    Dilution Recovery (Linearity)Demonstrates linearity throughout the assay rangeLinear from 0.8 to 29.9 ng/mL
    Spike Recovery≤10% error≤10% error (for all samples)
    Method Comparison (vs. LC-MS/MS)High correlation (R value close to 1, slope close to 1, intercept close to 0)y = 1.111x + 0.53, R = 0.972
    Method Comparison (vs. Predicate Device)High correlation (R value close to 1, slope close to 1, intercept close to 0)y = 1.126x - 0.03, R = 0.937
    Specificity (Major Metabolites)Acceptable level of cross-reactivityPartial cross-reactivity
    Specificity (Co-administered/OTC Drugs)Minimal to no cross-reactivityMinimal to no cross-reactivity
    Interfering Substances≤10% error≤10% error (at concentrations tested)
    Reagent Stability (Accelerated)Stable for acceptable durationUp to 13 months at 2-8°C
    Reagent On-board StabilityStable for acceptable durationUp to 35 days on Beckman AU680® clinical analyzer
    Calibrator Stability (Accelerated)Stable for acceptable durationUp to 15 months at -20°C
    Calibrator Open Vial StabilityStable for acceptable duration while capped42 days at 2-8°C

    Note: The document implies acceptance criteria based on standard laboratory practices for in vitro diagnostics. Specific numerical thresholds for "high correlation" or "acceptable level" for parameters like comparison studies and specificity are not explicitly stated as acceptance criteria but are inferred from the reported results being presented as evidence of performance.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific number of samples used for the "test set" for each performance study. It mentions "samples were tested" implying internal validation.

    • Functional Sensitivity (LOQ): Not specified.
    • Precision: Not specified.
    • Dilution Recovery: Not specified.
    • Spike Recovery: Not specified.
    • Method Comparison: Not specified. It indicates "samples were tested" and compared to LC-MS/MS and the predicate device.
    • Specificity: Not specified. Indicated "studies were conducted for available major metabolites... medications routinely co-administered... and other over-the-counter drugs."
    • Interfering Substances: Not specified. Indicated "endogenous substances... were tested."
    • Reagent Stability: Not specified.
    • Calibrator Stability: Not specified.

    Data Provenance: The studies appear to be internal validation studies conducted by Thermo Fisher Scientific. The document does not specify country of origin for the data, nor does it explicitly state whether the data was retrospective or prospective, though it's typically prospective for performance validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this type of device (immunoassay). The "ground truth" for an immunoassay is established by reference methods or known concentrations, not by expert interpretation of images or other subjective data. For example:

    • Method Comparison: The "ground truth" is provided by the LC-MS/MS method (a gold standard analytical technique for tacrolimus quantification). The predicate device also serves as a comparison.
    • Dilution Recovery and Spike Recovery: The "ground truth" is the known concentration of tacrolimus added to the samples.
    • Specificity: The "ground truth" for cross-reactivity is the known chemical structure and concentration of the potential interfering substances.

    4. Adjudication Method for the Test Set

    This is not applicable to this type of device. Adjudication methods (like 2+1, 3+1) are typically used in studies where multiple human readers independently assess data (e.g., medical images) and a consensus is needed to establish ground truth or resolve discrepancies. For this immunoassay, results
    are quantitative measurements, not subjective interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study assesses how human readers' performance changes with an AI aid. The QMS® Tacrolimus Immunoassay is an in vitro diagnostic device that quantifies a substance in whole blood through an automated process, not an AI system that assists human interpretation of complex medical cases.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the performance studies detailed in the summary (Functional Sensitivity, Precision, Dilution Recovery, Spike Recovery, Method Comparison, Specificity, Interfering Substances, and Stability) represent the standalone performance of the QMS® Tacrolimus Immunoassay. This device operates as an automated system to quantify tacrolimus, without direct human intervention in the interpretative process during testing. The results are generated by the algorithm/device itself.

    7. The Type of Ground Truth Used

    The ground truth used for validating the QMS® Tacrolimus Immunoassay stems from:

    • Reference Method: For method comparison, the LC-MS/MS (Liquid Chromatography-Mass Spectrometry/Mass Spectrometry) method served as the primary reference standard. This is considered a highly sensitive and specific analytical technique for drug quantification.
    • Known Concentrations: For studies like Functional Sensitivity, Dilution Recovery, Spike Recovery, Specificity, and Interfering Substances, the ground truth was based on samples with known, precisely prepared concentrations of tacrolimus, its metabolites, co-administered drugs, or endogenous interfering substances.
    • Predicate Device Comparison: The previously cleared Abbott ARCHITECT® Tacrolimus Assay served as a comparative reference method for demonstrating substantial equivalence.

    8. The Sample Size for the Training Set

    The document does not specify a training set sample size. For an immunoassay, the concept of a "training set" in the context of machine learning (where algorithms learn from data) is not directly applicable in the same way. The immunoassays are developed and optimized through iterative processes using various reagents, controls, and calibration materials, rather than "training" an algorithm on a distinct dataset.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, a traditional "training set" with ground truth in the sense of machine learning is not described or applicable here. The development and optimization of an immunoassay rely on established chemical and biological principles, precise formulation of reagents, and validation against reference methods and known standards to ensure accuracy and reliability. The ground truth for such development is built upon fundamental analytical chemistry and knowledge of the analyte (tacrolimus).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1