    K Number
    K123153

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-04-09

    (182 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Attain Command™ + SureValve™ Left Heart Delivery System and Attain Command™ + SureValve™ Guide Catheters for Left Heart Delivery are intended for introducing transvenous devices and leads into vessels of the left heart via the coronary sinus.

    The Attain Select™ II + SureValve™ delivery catheter system is indicated for the delivery of contrast medium and transvenous devices to the coronary sinus and left-heart venous anatomy. The delivery catheter system is indicated for use with outer quide catheters.

    Device Description

    The Attain Command™ + SureValve™ Left Heart Delivery System and Attain Command™ + SureValve™ Guide Catheters for Left Heart Delivery are intended for introducing transvenous devices and leads into vessels of the left heart via the coronary sinus.

    The Attain Command™ + SureValve™ Left-Heart Delivery System kits each contain two outer quide catheters with an integrated hemostasis valve (SureValve™), up to two valve tools, one dilator, one guide wire, and one slitter. The Attain Command™ + SureValve™ Left-Heart Delivery System is available in two models: Attain Command TM + SureValve ™ 6250VC Left Heart Delivery System and Attain Command™ + SureValve™ 6250VS Left Heart Delivery System. With the exception of the two guide catheters, all system components packaged in each kit are identical. Each guide catheter model is different with respect to the guide catheter shape and length.

    The Attain Command™ + SureValve™ Guide Catheters for Left Heart Delivery individual packs each contain one guide catheter with an integrated hemostasis valve (SureValve™), up to two valve tools, and one dilator. The Attain Command™ + SureValve™ Guide Catheters for Left Heart Delivery are available in 12 models. Each model is different with respect to the guide catheter shape and length and dilator length.

    The Attain Select™ II + SureValve™ delivery catheter system is designed to facilitate leftheart lead delivery to a desired cardiac vein. The Attain Select™ II + SureValve™ delivery catheter system is indicated for the delivery of contrast medium and transvenous devices to the coronary sinus and left-heart venous anatomy. The delivery catheter system is indicated for use with outer quide catheters. Together, the catheters function as a telescoping system that can provide additional subselecting capabilities.

    The delivery catheter system consists of a delivery catheter with an integrated hemostasis valve (SureValve™), an inner catheter, and up to two valve tools. The Attain Select™ II + SureValve™ delivery catheter system is available in 8 models. The Attain Select™ II + SureValve™ inner catheter is identical for all configurations. Each model is different with respect to delivery catheter shape and length.

    AI/ML Overview

    The provided text describes the 510(k) summary for several Medtronic devices: Attain Command™ + SureValve™ Left Heart Delivery System, Attain Command™ + SureValve™ Guide Catheters for Left Heart Delivery, and Attain Select™ II + SureValve™ Delivery Catheter System.

    Based on the document, these devices did not undergo a study involving AI or algorithm performance. The acceptance criteria and "study" that proves the device meets them are entirely based on non-clinical performance data, comparing the new devices to existing predicate devices.

    Here's the breakdown of the information requested, as it applies to this submission:


    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document does not explicitly list "acceptance criteria" for each test with numerical thresholds. Instead, it states that the devices "met all specified design and performance requirements" and that biocompatibility tests "indicate that the devices are biocompatible for their intended use."

    Acceptance Criteria CategoryReported Device Performance
    In Vitro Bench TestsDevices met all specified design and performance requirements.
    Dimensional TestingMet all specified design and performance requirements.
    Catheter Safety & Performance TestingMet all specified design and performance requirements.
    Component/Accessory Safety & Performance TestingMet all specified design and performance requirements.
    Interface Safety & Performance Testing (compatibility with transvenous devices)Met all specified design and performance requirements.
    Packaging QualificationsMet all specified design and performance requirements.
    Shelf Life TestingMet all specified design and performance requirements.
    BiocompatibilityDevices are biocompatible for their intended use.
    CytotoxicityPassed
    SensitizationPassed
    Intracutaneous ReactivityPassed
    Acute Systemic ToxicityPassed
    Material Mediated PyrogenPassed
    In-vitro HemolysisPassed
    Partial Thromboplastin Time (PTT) Coagulation TestingPassed
    C3a Complement Activation AssayPassed
    Sc5b-9 Complement Activation AssayPassed
    In Vivo ThromboresistancePassed
    Physicochemical Tests (Aqueous)Passed
    Physicochemical Tests (Non-Aqueous)Passed
    SterilizationAchieved a minimum Sterility Assurance Level (SAL) of 10-6.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Test Set Sample Size: Not specified for each individual non-clinical test. The document refers to "testing" and "validations" but does not provide specific numbers of units tested.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). Non-clinical tests were conducted by Medtronic Ireland.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. The "ground truth" for this device, in the context of the 510(k) submission, is established through the results of non-clinical, laboratory-based tests (bench tests, biocompatibility tests, sterilization validations) against pre-defined engineering and safety specifications, and compliance with relevant ISO standards. There were no human experts evaluating performance in this context, nor was there a "ground truth" established by experts in the way it would be for an AI diagnostic device (e.g., radiologists interpreting images).


    4. Adjudication Method for the Test Set

    Not applicable. This submission relies entirely on objective non-clinical test results rather than human judgments that would require adjudication.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This submission does not involve AI or algorithms, nor does it include a clinical study (MRMC or otherwise).


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This submission does not involve AI or algorithms.


    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by:

    • Engineering specifications and design requirements: For in vitro bench tests.
    • International Standards (e.g., ISO 10993-1:2009 for biocompatibility, ISO 11135-1 and EN556-1 for sterilization): These standards define the acceptable methodologies and criteria for evaluating device properties.
    • FDA Good Laboratory Practice (GLP) regulations (21 CFR, Part 58): For biocompatibility testing methodology.

    8. The Sample Size for the Training Set

    Not applicable. There is no training set as this is not an AI/ML device.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set.