Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K123043

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-10-25

    (27 days)

    Product Code
    Regulation Number
    868.1400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IRMA mainstream gas analyzer is intended to be connected to other medical devices for monitoring of breath rate and the breathing gases CO2, N2O and the anesthetic agents Halothane, Enflurane, Isoflurane, Sevoflurane and Desflurane.

    It is intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit and patient room for adult, pediatric and infant patients. IRMA CO2 may also be used in the emergency medical services environment and road ambulances.

    Note: An IRMA mainstream gas analyzer shall only be connected to medical backboard devices approved by PHASEIN.

    Device Description

    IRMA is a mainstream respiratory gas analyzer based on infrared gas spectrometry. It is intended to be connected to another medical backboard device for display of respiratory parameters. IRMA is connected to the patient breathing circuit via the Airway Adapter. This premarket submission adds the M80 device as a host backboard display to IRMA. There are no changes to the previously cleared sensor technology.
    The following IRMA models are available:
    a) IRMA CO2 (model/catalogue number 200101), measurement of CO2.
    b) IRMA AX+ (model/catalogue number 200601), measurement of CO2, N2O and anesthetic agents with automatic agent identification
    The IRMA product family also includes the adult and pediatric/infant Airway Adapter. A modura holder and velcro holder are offered as optional accessories.

    AI/ML Overview

    The provided 510(k) summary for the IRMA - Infrared Mainstream Gas Analyzer (K123043) describes the device, its intended use, and comparison to predicate devices, but it does not explicitly list quantitative acceptance criteria or a detailed study proving the device meets specific performance metrics.

    However, based on the information provided, we can infer the acceptance criteria and the nature of the study:

    Inferred Acceptance Criteria and Reported Device Performance:

    The primary acceptance criterion is substantial equivalence to the predicate devices (ISA - Infrared Sidestream Gas Analyzer, K103604, and VEO Multigas Monitor for Pocket PC, K081601) in terms of performance, safety, and effectiveness. The reported device performance is that it met or exceeded these criteria.

    Acceptance Criteria (Inferred)Reported Device Performance
    Performance equivalent or superior to predicate devices (ISA, VEO Multigas Monitor) throughout the operating range for gas measurement capabilities (CO2, N2O, anesthetic agents)."The IRMA, models 200101 and 200601 demonstrated performance, safety and effectiveness equivalent or superior to its predicates in all characteristics."
    Safety equivalent or superior to predicate devices."The IRMA, models 200101 and 200601 demonstrated performance, safety and effectiveness equivalent or superior to its predicates in all characteristics."
    Effectiveness equivalent or superior to predicate devices."The IRMA, models 200101 and 200601 demonstrated performance, safety and effectiveness equivalent or superior to its predicates in all characteristics."
    Same basic technology concepts as predicate devices."The IRMA uses the same basic technology concepts used in the predicate devices."
    Same intended use as predicate devices."The intended use of the IRMA and its predicates are the same."
    Equivalent electric power consumption."All the devices consume equivalent amounts of electric power."
    Utilization of single-patient-use connections to interface with gases in the breathing circuit."All the devices...utilize single-patient-use connections to interface with gases in the breathing circuit."
    Equivalent labeling and materials."Labeling and materials used are equivalent."

    1. A table of acceptance criteria and the reported device performance:

    (See table above)


    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated. The document mentions "calibrated gas samples" and "legally marketed anesthesia and ventilation devices," but no specific number of samples or devices is provided.
    • Data Provenance: Not explicitly stated. The testing was described as "in direct comparison to predicates," suggesting laboratory or simulated use environments rather than clinical patient data. There is no mention of country of origin or whether it was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified. It's likely that the "ground truth" for gas concentrations was established by the calibrated gas samples themselves rather than human experts, and the performance with anesthesia/ventilation devices was assessed against their known specifications.

    4. Adjudication method for the test set:

    • Adjudication method: Not described. Given the nature of objective gas measurement, it's unlikely that human adjudication (e.g., 2+1, 3+1) was used for establishing ground truth regarding gas concentrations. The comparison would be against known reference values from calibrated samples.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. This device is a gas analyzer, not an AI-assisted diagnostic tool that requires human interpretation of images or other data.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance: Yes, implicitly. The device itself performs the gas analysis. The "testing in direct comparison to predicates throughout the operating range was conducted using calibrated gas samples and legally marketed anesthesia and ventilation devices." This implies evaluating the device's ability to accurately measure gases against a known standard (calibrated gases) and its functional compatibility with other medical equipment, without direct human intervention in the measurement process itself. The IRMA is a sensor that interfaces with a "medical backboard device for display of respiratory parameters," indicating its standalone measurement capability.

    7. The type of ground truth used:

    • Type of Ground Truth: The ground truth for gas concentration measurements was established using calibrated gas samples. For compatibility and functional testing, the "legally marketed anesthesia and ventilation devices" would serve as a benchmark for expected performance and integration.

    8. The sample size for the training set:

    • Sample Size for Training Set: Not applicable or not specified. This device is an infrared gas analyzer, a sensor technology, not a machine learning or AI-based system that typically uses a "training set" in the conventional sense. Its "training" would be inherent in its design, calibration, and manufacturing process based on physics and spectroscopy principles, rather than a data-driven model.

    9. How the ground truth for the training set was established:

    • Ground Truth for Training Set: Not applicable in the context of an AI/ML training set. For a physical device like this, the "ground truth" for its development would be based on established scientific principles of infrared spectroscopy and gas analysis, validated through rigorous engineering design, component selection, and manufacturing quality control.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1